Intravascular ultrasonic assessment of Stent diameters derived from manufacturer's compliance charts

被引:35
作者
Costa, JD
Mintz, GS
Carlier, SG [1 ]
Costa, RA
Fujii, K
Sano, K
Kimura, M
Lui, J
Weisz, G
Moussa, I
Dangas, G
Mehran, R
Lansky, AJ
Kreps, EM
Collins, M
Stone, GW
Moses, JW
Leon, MB
机构
[1] Cardiovasc Res Fdn, New York, NY USA
[2] Columbia Univ, Med Ctr, New York, NY USA
关键词
D O I
10.1016/j.amjcard.2005.02.049
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We used intravascular ultrasound (IVUS) to assess the accuracy of manufacturers' stent balloon compliance charts. Many interventional cardiologists rely on manufacturers' compliance charts to select stent size and optimize stent diameters according to inflation pressures during percutaneous procedures. We randomly selected 212 patients who had de novo coronary lesions that had been treated with a single, bare metal, 3.0-mm stent (Bx velocity, NIR, TETRA/PENTA, S660/S670/S7) under IVUS guidance. Cases of stent overlap and postdilatation with another balloon were excluded. Predicted stent diameters were derived from each manufacturer's compliance charts, and stent size and final maximal deployment pressures were derived from each physician's report. IVUS-measured minimum stent diameters (range 1.4 to 4.0 mm, mean 2.79 +/- 0.48) were smaller than predicted diameters (range 3.1 to 4.57 mm, mean 3.79 +/- 0.44). The ratio of IVUS to predicted diameters ranged from 44% to 97% (mean 74 +/- 10%). This finding was common to all 3 stent sizes: 74 +/- 12% for 3.0 mm, 73 +/- 9% for 3.5 mm, and 74 9% for 4.0-mm stents (p = 0.9). This finding was also common to all 4 stent manufacturers, 72 +/- 8% for Boston Scientific, 76 +/- 11% for Guidant, 73 +/- 9% for Cordis, and 74 11% for Medtronic (p = 0.13), and to different stent lengths. Only 3.8% of the stents achieved 90% of the predicted minimum stent diameters, and only 24.6% achieved 80% of the predicted minimum stent diameters. In conclusion, in human coronary arteries, minimal stent diameter measured by IVUS is significantly smaller than that predicted by in vitro compliance charts. These differences are independent of stent manufacturer, length, diameter, and deployment pressure. (c) 2005 Elsevier Inc. All rights reserved.
引用
收藏
页码:74 / 78
页数:5
相关论文
共 18 条
[1]  
Barragan P, 2000, CATHETER CARDIO INTE, V50, P112, DOI 10.1002/(SICI)1522-726X(200005)50:1<112::AID-CCD25>3.0.CO
[2]  
2-X
[3]   Mechanisms of residual lumen stenosis after high-pressure stent implantation -: A quantitative coronary angiography and intravascular ultrasound study [J].
Bermejo, J ;
Botas, J ;
García, E ;
Elízaga, J ;
Osende, J ;
Soriano, J ;
Abeytua, M ;
Delcán, JL .
CIRCULATION, 1998, 98 (02) :112-118
[4]  
Blasini R, 1997, CATHETER CARDIO DIAG, V42, P113, DOI 10.1002/(SICI)1097-0304(199710)42:2<113::AID-CCD2>3.0.CO
[5]  
2-G
[6]   In vivo assessment of stent expansion and recoil in normal porcine coronary arteries - Differential outcome by stent design [J].
Carrozza, JP ;
Hosley, SE ;
Cohen, DJ ;
Baim, DS .
CIRCULATION, 1999, 100 (07) :756-760
[7]   Reference chart derived from post-stent-implantation intravascular ultrasound predictors of 6-month expected restenosis on quantitative coronary angiography [J].
de Feyter, PJ ;
Kay, P ;
Disco, C ;
Serruys, PW .
CIRCULATION, 1999, 100 (17) :1777-1783
[8]   Comparison of quantitative coronary angiography, intravascular ultrasound, and coronary pressure measurement to assess optimum stent deployment [J].
Hanekamp, CEE ;
Koolen, JJ ;
Pijls, NHJ ;
Michels, HR ;
Bonnier, HJRM .
CIRCULATION, 1999, 99 (08) :1015-1021
[9]   Overestimation of stent delivery balloon diameters by manufacturers' compliance tables: A quantitative coronary analysis of duet and NIR stent implantation [J].
Hehrlein, C ;
DeVries, JJ ;
Wood, TA ;
Haller, SD ;
Arab, A ;
Kloostra, AC ;
Lauer, MA ;
Fischell, TA .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2001, 53 (04) :474-478
[10]   Intravascular ultrasound predictors of angiographic restenosis in lesions treated with Palmaz-Schatz stents [J].
Hoffmann, R ;
Mintz, GS ;
Mehran, R ;
Pichard, AD ;
Kent, KM ;
Satler, LF ;
Popma, JJ ;
Wu, HS ;
Leon, MB .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 31 (01) :43-49