Precise isotope ratio determination of common Pb using quadrupole LA-ICP-MS with optimized laser sampling conditions and a robust mixed-gas plasma

被引:41
作者
Crowe, SA [1 ]
Fryer, BJ
Samson, IM
Gagnon, JE
机构
[1] Univ Windsor, Dept Earth Sci, Windsor, ON N9B 3P4, Canada
[2] Univ Windsor, Great Lakes Inst Environm Res, Windsor, ON N9B 3P4, Canada
关键词
D O I
10.1039/b303225f
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This work presents a method for the precise (0.2% RSE) determination of common (i.e. non-radiogenic) Pb isotope ratios using quadrupole Nd:YAG (266 nm) LA-ICP-MS at low (similar to2 ppm) Pb concentrations. Laser sampling conditions significantly influence the precision of Pb isotope ratio measurements. This paper presents a set of optimum sampling conditions for the described system. Setting the laser focus above the sample surface significantly improves the precision of ratio measurements due to increased count rates and a reduction in the heterogeneity of particulate matter produced by fracturing at the site of ablation. In addition, using a more robust, mixed-gas (Ar-N-2) plasma significantly increases sensitivity and reduces mass bias. With a mixed Ar-N-2 plasma and optimized laser sampling conditions, single collector quadrupole LA-ICP-MS can be superior, for some applications (e.g., where micrometre-scale spatial resolution is important), to TIMS, and at low Pb concentrations is a cost-effective alternative to LA-MC-ICP-MS. The application of single detector, quadrupole LA-ICP-MS to the precise determination of common Pb isotope ratios in minerals has not been previously documented. This method is a powerful tool for use in isotope tracer studies of ore deposits and has potential applications to a range of environmental problems.
引用
收藏
页码:1331 / 1338
页数:8
相关论文
共 38 条
[1]   Direct determination of lead isotope ratios by laser ablation-inductively coupled plasma-quadrupole mass spectrometry in lake sediment samples [J].
Aries, S ;
Motelica-Heino, M ;
Freydier, R ;
Grezes, T ;
Polvé, M .
GEOSTANDARDS NEWSLETTER-THE JOURNAL OF GEOSTANDARDS AND GEOANALYSIS, 2001, 25 (2-3) :387-398
[2]  
BARRIE CT, 2000, GEOLOGY GEOCHEMISTRY
[3]   HIGH-RESOLUTION SIMS ANALYSIS OF COMMON LEAD [J].
BELSHAW, NS ;
ONIONS, RK ;
MARTEL, DJ ;
BURTON, KW .
CHEMICAL GEOLOGY, 1994, 112 (1-2) :57-70
[4]   A new variable dispersion double-focusing plasma mass spectrometer with performance illustrated for Pb isotopes [J].
Belshaw, NS ;
Freedman, PA ;
O'Nions, RK ;
Frank, M ;
Guo, Y .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 1998, 181 :51-58
[5]   Effects of crater development on fractionation and signal intensity during laser ablation inductively coupled plasma mass spectrometry [J].
Borisov, OV ;
Mao, XL ;
Russo, RE .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2000, 55 (11) :1693-1704
[6]   Climate and ocean dynamics and the lead isotopic records in Pacific ferromanganese crusts [J].
Christensen, JN ;
Halliday, AN ;
Godfrey, LV ;
Hein, JR ;
Rea, DK .
SCIENCE, 1997, 277 (5328) :913-918
[7]  
COMPSTON W, 1991, GEOCHEM SOC SPEC PUB, V3
[8]  
CROWE SA, 2002, GAC MAC, V27, P24
[9]   High precision lead isotope ratio measurements by multicollector-ICP-MS in variable matrices [J].
Ehrlich, S ;
Karpas, Z ;
Ben-Dor, L ;
Halicz, L .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2001, 16 (09) :975-977
[10]   A comparison between quadrupole, double focusing and multicollector ICP-MS instruments. Part I. Evaluation of total combined uncertainty for lead isotope ratio measurements [J].
Encinar, JR ;
Alonso, JIG ;
Sanz-Medel, A ;
Main, S ;
Turner, PJ .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2001, 16 (04) :315-321