Truly hybrid x-ray/MR imaging: Toward a streamlined clinical system

被引:29
作者
Ganguly, A
Wen, ZF
Daniel, BL
Butts, K
Kee, ST
Rieke, V
Do, HM
Pelc, NJ
Alley, MT
Fahrig, R
机构
[1] Stanford Univ, Lucas MRS Ctr, Dept Radiol, Stanford, CA 94305 USA
[2] Stanford Univ, Lucas MRS Ctr, Dept Phys, Stanford, CA 94305 USA
关键词
X-ray fluoroscopy; interventional MRI; truly hybrid imaging; signal to noise ratio; system compatibility;
D O I
10.1016/j.acra.2005.03.076
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Rationale and Objective. We have installed an improved x-ray/MR (XMR) truly hybrid system with higher imaging signal-to-noise ratio (SNR) and versatility than our first prototype. In our XMR design, a fixed anode x-ray fluoroscopy system is positioned between the two donut-shaped magnetic poles of a 0.5T GE Signa-SP magnet (SP-XMR). This paper describes the methods for increased compatibility between the upgraded x-ray and MR systems that have helped improve patient management. Materials and Methods. A GE OEC 9800 system (GE OEC Salt Lake, City" UT) was specially reconfigured for permitting x-ray fluoroscopy inside the interventional magnet. A higher power x-ray tube, a new permanent tube mounting system, automatic exposure control (AEC), remote controlled collimators, choice of multiple frame rates, DICOM image compatibility, magnetically shimmed x-ray detector, x-ray compatible MR coil, and better RF shielding are the highlights of the new system. A total of 23 clinical procedures have been conducted with SP-XMR guidance of which five were performed using the new system. Results. The 70% increased power for fluoroscopy, and a new 6 times higher power single frame imaging mode, has improved imaging capability. The choice of multiple imaging frame rates, AEC, and collimator control allow reduction in x-ray exposure to the patient. The DICOM formatting has permitted easy transfer of clinical images over the hospital PACS network. The increased MR compatibility of the detector and the x-ray transparent MR coil has enabled faster switching between x-ray and MR imaging modes. Conclusion. The improvements introduced in our SP-XMR system have further streamlined x-ray/MR hybrid imaging. Additional clinical procedures could benefit from the new SP-XMR imaging.
引用
收藏
页码:1167 / 1177
页数:11
相关论文
共 23 条
[1]  
Chu RM, 2001, NEUROIMAG CLIN N AM, V11, P715
[2]   Performance of a static-anode/flat-panel x-ray fluoroscopy system in a diagnostic strength magnetic field: A truly hybrid x-ray/MR imaging system [J].
Fahrig, R ;
Wen, Z ;
Ganguly, A ;
DeCrescenzo, G ;
Rowlands, JA ;
Stevens, GM ;
Saunders, RF ;
Pelc, NJ .
MEDICAL PHYSICS, 2005, 32 (06) :1775-1784
[3]   First use of a truly-hybrid X-ray/MR imaging system for guidance of brain biopsy [J].
Fahrig, R ;
Heit, G ;
Wen, Z ;
Daniel, BL ;
Butts, K ;
Pelc, NJ .
ACTA NEUROCHIRURGICA, 2003, 145 (11) :995-997
[4]   Truly hybrid interventional MR/x-ray system: Investigation of in vivo applications [J].
Fahrig, R ;
Butts, K ;
Wen, ZF ;
Saunders, R ;
Kee, ST ;
Sze, DY ;
Daniel, BL ;
Laerum, F ;
Pelc, NJ .
ACADEMIC RADIOLOGY, 2001, 8 (12) :1200-1207
[5]  
Fahrig R, 2001, J MAGN RESON IMAGING, V13, P294, DOI 10.1002/1522-2586(200102)13:2<294::AID-JMRI1042>3.0.CO
[6]  
2-X
[7]  
FAHRIG R, 2003, INT SOC MAGN RES MED
[8]  
FAHRIG R, UNPUB MED PHYS
[9]  
FAHRIG R, 2004, US BOARD TRUSTEES LE
[10]  
Hall WA, 2001, NEUROIMAG CLIN N AM, V11, P705