Area measurement of knife-edge and cylindrical apertures using ultra-low force contact fibre probe on a CMM

被引:5
作者
Muralikrishnan, B. [1 ]
Stone, J. A. [1 ]
Stoup, J. R. [1 ]
机构
[1] Natl Inst Stand & Technol, Gaithersburg, MD 20899 USA
关键词
D O I
10.1088/0026-1394/45/3/004
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Several radiometric and photometric measurements depend on high accuracy area measurement of precision apertures. Some apertures have sharp edges and are generally measured optically. At the Precision Engineering Division of the National Institute of Standards and Technology (NIST), we have developed a contact fibre (fiber) probe for diameter and form measurement of micro-holes (holes of size 100 mu m or larger). This probe exerts extremely small forces, under 5 mu N, and can therefore be used on knife-edge apertures without causing edge damage. We have measured the diameter and roundness of three knife-edge and one cylindrical apertures with this probe. The uncertainty in diameter ranges from 0.06 mu m (k = 1) to 0.17 mu m (k = 1). The uncertainty contributions from the probing system and machine positioning are together only 35 nm (k = 1). The largest contributors to the diameter uncertainty are the overall form (sampling uncertainty) and surface finish (mechanical filtering due to finite probe size) of the aperture.
引用
收藏
页码:281 / 289
页数:9
相关论文
共 18 条
[1]  
*ANSI NCSL, Z54021997 ANSI NCSL
[2]   Measurement of small apertures [J].
Fedchak, JA ;
Carter, AC ;
Raju, D .
METROLOGIA, 2006, 43 (02) :S41-S45
[3]   A NONCONTACT MEASUREMENT OF RADIOMETRIC APERTURES WITH AN OPTICAL MICROTOPOGRAPHY SENSOR [J].
FISCHER, J ;
STOCK, M .
MEASUREMENT SCIENCE AND TECHNOLOGY, 1992, 3 (08) :693-698
[4]   Geometric area measurements of circular apertures for radiometry at NIST [J].
Fowler, J ;
Litorja, M .
METROLOGIA, 2003, 40 (01) :S9-S12
[5]   Summary of high-accuracy aperture-area measurement capabilities at the NIST [J].
Fowler, JB ;
Saunders, RD ;
Parr, AC .
METROLOGIA, 2000, 37 (05) :621-623
[6]   HIGH-ACCURACY MEASUREMENT OF APERTURE AREA RELATIVE TO A STANDARD KNOWN APERTURE [J].
FOWLER, JB ;
DEZSI, G .
JOURNAL OF RESEARCH OF THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY, 1995, 100 (03) :277-283
[7]   High-accuracy aperture-area measurement facilities at the National Institute of Standards and Technology [J].
Fowler, JB ;
Durvasula, RS ;
Parr, AC .
METROLOGIA, 1998, 35 (04) :497-500
[8]   A non-contact technique providing improved accuracy in area measurements of radiometric apertures [J].
Hartmann, J ;
Fischer, J ;
Seidel, J .
METROLOGIA, 2000, 37 (05) :637-640
[9]   Advanced comparator method for measuring ultra-small aperture areas [J].
Hartmann, J .
MEASUREMENT SCIENCE AND TECHNOLOGY, 2001, 12 (10) :1678-1682
[10]   A new optical method for high-accuracy determination of aperture area [J].
Ikonen, E ;
Toivanen, P ;
Lassila, A .
METROLOGIA, 1998, 35 (04) :369-372