Flat panel computed tomography of human ex vivo heart and bone specimens: initial experience

被引:29
作者
Nikolaou, K [1 ]
Flohr, T
Stierstorfer, K
Becker, CR
Reiser, MF
机构
[1] Univ Munich, Dept Clin Radiol, Munich, Germany
[2] Siemens Med Solut, CT Div, Forchheim, Germany
关键词
computed tomography; flat panel detector;
D O I
10.1007/s00330-004-2537-4
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this technical investigation was the detailed description of a prototype flat panel detector computed tomography system (FPCT) and its initial evaluation in an ex vivo setting. The prototype FPCT scanner consists of a conventional radiographic flat panel detector, mounted on a multi-slice CT scanner gantry. Explanted human ex vivo heart and foot specimens were examined. Images were reformatted with various reconstruction algorithms and were evaluated for high-resolution anatomic information. For comparison purposes, the ex vivo specimens were also scanned with a conventional 16-detector-row CT scanner (Sensation 16, Siemens Medical Solutions, Forchheim, Germany). With the FPCT prototype used, a 1,024x768 resolution matrix can be obtained, resulting in an isotropic voxel size of 0.254.254.25 mm at the iso-center. Due to the high spatial resolution, very small structures such as trabecular bone or third-degree, distal branches of coronary arteries could be visualized. This first evaluation showed that flat panel detector systems can be used in a cone-beam computed tomography scanner and that very high spatial resolutions can be achieved. However, there are limitations for in vivo use due to constraints in low contrast resolution and slow scan speed.
引用
收藏
页码:329 / 333
页数:5
相关论文
共 22 条
[1]  
Anxionnat R, 1998, J NEURORADIOLOGY, V25, P251
[2]  
Bruder H, 2000, IEEE T MED IMAGING, V19, P873, DOI 10.1109/42.887836
[3]   Coronary artery disease: new insights and their implications for radiology [J].
Dewey, M ;
Borges, AC ;
Kivelitz, D ;
Taupitz, M ;
Wagner, S ;
Baumann, G ;
Hamm, B .
EUROPEAN RADIOLOGY, 2004, 14 (06) :1048-1054
[4]   Multiplanar reconstructions and three-dimensional imaging (computed rotational osteography) of complex fractures by using a C-arm system: Initial results [J].
El-Sheik, M ;
Heverhagen, JT ;
Alfke, H ;
Froelich, JJ ;
Hornegger, J ;
Brunner, T ;
Klose, KJ ;
Wagner, HJ .
RADIOLOGY, 2001, 221 (03) :843-849
[5]   Computed tomography and magnetic resonance imaging for noninvasive coronary angiography and plaque imaging - Current and potential future concepts [J].
Fayad, ZA ;
Fuster, V ;
Nikolaou, K ;
Becker, C .
CIRCULATION, 2002, 106 (15) :2026-2034
[6]   New technical developments in multislice CT - Part 1: Approaching isotropic resolution with sub-millimeter 16-slice scanning [J].
Flohr, T ;
Stierstorfer, K ;
Bruder, H ;
Simon, J ;
Schaller, S .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2002, 174 (07) :839-845
[7]   New technical developments in multislice CT, part 2: Sub-millimeter 16-slice scanning and increased gantry rotation speed for cardiac Imaging [J].
Flohr, T ;
Bruder, H ;
Stierstorfer, K ;
Simon, J ;
Schaller, S ;
Ohnesorge, B .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2002, 174 (08) :1022-1027
[8]   A performance comparison of flat-panel imager-based MV and kV cone-beam CT [J].
Groh, BA ;
Siewerdsen, JH ;
Drake, DG ;
Wong, JW ;
Jaffray, DA .
MEDICAL PHYSICS, 2002, 29 (06) :967-975
[9]   Cone-beam computed tomography with a flat-panel imager: Initial performance characterization [J].
Jaffray, DA ;
Siewerdsen, JH .
MEDICAL PHYSICS, 2000, 27 (06) :1311-1323
[10]  
Kalender WA, 2003, RADIOLOGE, V43, P379, DOI 10.1007/s00117-003-0897-4