Inappropriate shocks from implanted cardioverter defibrillators caused by sensing of diaphragmatic myopotentials

被引:14
作者
Peters, RW
Cooklin, M
Brockman, R
Shorofsky, SR
Gold, MR
机构
[1] Univ Maryland, Sch Med, Dept Med, Div Cardiol, Baltimore, MD 21201 USA
[2] Dept Vet Affairs Med Ctr, Baltimore, MD USA
关键词
implantable cardioverter defibrillator; oversensing; diaphragmatic potentials;
D O I
10.1023/A:1009768822780
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Inappropriate shocks remain one of the major problems associated with the use of implantable cardioverter defibrillators (ICD). We describe 3 patients who had inappropriate shocks due to oversensing of diaphragmatic myopotentials. In all 3, oversensing could be demonstrated during deep inspiration or Valsalva maneuver. The problem was remedied in one by insertion of a separate rate sensing lead in the right ventricular outflow tract, in the second by decreasing the rate of antibradycardia pacing of the ICD, and in the third by decreasing the sensitivity of the ICD. We conclude that the possibility of sensed diaphragmatic myopotentials should be considered at the time of ICD implant so that appropriate preventive measures can be taken.
引用
收藏
页码:367 / 370
页数:4
相关论文
共 7 条
[1]   INHIBITION OF BIPOLAR DEMAND PACEMAKER BY DIAPHRAGMATIC MYOPOTENTIALS [J].
BAROLD, SS ;
ONG, LS ;
FALKOFF, MD ;
HEINLE, RA .
CIRCULATION, 1977, 56 (04) :679-683
[2]  
BAROLD SS, 1993, CARDIAC PACING, P537
[3]  
DAMLE RS, 1997, CIRCULATION, P721
[4]   EVALUATION OF OUTPATIENTS EXPERIENCING IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR SHOCKS ASSOCIATED WITH MINIMAL SYMPTOMS [J].
HAMER, ME ;
CLAIR, WK ;
WILKINSON, WE ;
GREENFIELD, RA ;
PRITCHETT, ELC ;
PAGE, RL .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (05) :938-943
[5]   SPURIOUS DISCHARGES DUE TO LATE INSULATION BREAK IN ENDOCARDIAL SENSING LEADS FOR CARDIOVERTER-DEFIBRILLATORS [J].
PETERS, RW ;
FOSTER, AH ;
SHOROFSKY, SR ;
NOLAN, DA ;
GOLD, MR .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (03) :478-481
[6]   Sensing and tachyarrhythmia detection problems in implantable cardioverter defibrillators [J].
Reiter, MJ ;
Mann, DE .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1996, 7 (06) :542-558
[7]   An unusual cause of ICD shock [J].
Sticherling, C ;
Klingenheben, T ;
Hohnloser, SH .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (09) :2265-2267