Characterisation of wear debris from UHMWPE on zirconia ceramic, metal-on-metal and alumina ceramic-on-ceramic hip prostheses generated in a physiological anatomical hip joint simulator

被引:94
作者
Tipper, JL [1 ]
Firkins, PJ
Besong, AA
Barbour, PSM
Nevelos, J
Stone, MH
Ingham, E
Fisher, J
机构
[1] Univ Leeds, Div Microbiol, Leeds LS2 9JT, W Yorkshire, England
[2] Univ Leeds, Sch Mech Engn, Leeds LS2 9JT, W Yorkshire, England
[3] Leeds Gen Infirm, Dept Orthopaed Surg, Leeds LS1 3EX, W Yorkshire, England
基金
英国工程与自然科学研究理事会;
关键词
wear; wear debris; hip prostheses; simulation;
D O I
10.1016/S0043-1648(01)00653-6
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
There is currently much interest in the characterisation of wear debris from different types of artificial hip joints. There have been numerous studies on the wear of UHMWPE in hip joint simulators, but relatively few studies on the wear of alternative materials such as metal-on-metal (MOM) and ceramic-on-ceramic (COC). The aim of this study was to compare the wear volumes and wear debris generated from zirconia ceramic-on-UHMWPE, MOM and COC hip joints under identical conditions in the same hip joint simulator. All prostheses showed an initial higher 'bedding in' wear rate, which was followed by a lower steady state wear rate. The zirconia ceramic-on UHMWPE prostheses showed the highest wear rates (31 +/- 4.0 mm(3)/million cycles), followed by the MOM (1.23 +/- 0.5 mm/million cycles), with the COC prostheses showing significantly (P < 0.01) lower wear rates at 0.05 +/- 0.02 mm(3)/million cycles. The mode (+/- 95% confidence Emits) of the size distribution of the UHMWPE wear debris was 300 +/- 200, 30 +/- 2.25 nm for the metal particles, and 9 +/- 0.5 nm for the ceramic wear particles. The UHMWPE particles were significantly larger (P < 0.05) than the metal and ceramic wear particles, and the metal particles were significantly larger (P < 0.05) than the ceramic wear particles. A variety of morphologies and sizes were observed for the UHMWPE wear particles, including submicrometer granules and large flakes in excess of 50 mum. However, the wear particles generated in both the MOM and COC articulations were very uniform in size and oval or round in shape. This investigation has demonstrated substantial differences in volumetric wear. The in vitro wear rates for the zirconia-on-UHMWPE and MOM are comparable with clinical studies and the UHMWPE and metal wear particles were similar to the wear debris isolated from retrieved tissues. However, the alumina/alumina wear rate was lower than some clinical retrieval studies, and the severe wear patterns and micrometer-sized particles described in vivo were not reproduced here. This study revealed significant differences in the wear volumes and particle sizes from the three different prostheses. In addition, this study has shown that the alternative bearing materials such as MOM and COC may offer a considerable advantage over the more traditional articulations which utilise UHMWPE as a bearing material, both in terms of wear volume and osteolytic potential. (C) 2001 Elsevier Science BN. All rights reserved.
引用
收藏
页码:120 / 128
页数:9
相关论文
共 61 条
[1]   The effects of particulate cobalt, chromium and cobalt-chromium alloy on human osteoblast-like cells in vitro [J].
Allen, MJ ;
Myer, BJ ;
Millett, PJ ;
Rushton, N .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1997, 79B (03) :475-482
[2]  
Amstutz HC, 1996, CLIN ORTHOP RELAT R, pS11
[3]   A hip joint simulator study using simplified loading and motion cycles generating physiological wear paths and rates [J].
Barbour, PSM ;
Stone, MH ;
Fisher, J .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1999, 213 (H6) :455-467
[4]  
BENDALL SP, 1996, P 5 WORLD BIOM C TOR
[5]   Quantitative comparison of wear debris from UHMWPE that has and has not been sterilised by gamma irradiation [J].
Besong, AA ;
Tipper, JL ;
Ingham, E ;
Stone, MH ;
Wroblewski, BM ;
Fisher, J .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1998, 80B (02) :340-344
[6]  
BESONG AA, 1997, P WORLD TRIB C
[7]   Wear of ultra-high molecular weight polyethylene acetabular cups in a physiological hip joint simulator in the anatomical position using bovine serum as a lubricant [J].
Bigsby, RJA ;
Hardaker, CS ;
Fisher, J .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1997, 211 (03) :265-269
[8]  
Bragdon C R, 1996, Proc Inst Mech Eng H, V210, P157, DOI 10.1243/PIME_PROC_1996_210_408_02
[9]  
Campbell P, 1996, Proc Inst Mech Eng H, V210, P167, DOI 10.1243/PIME_PROC_1996_210_409_02
[10]   WIDESPREAD DISSEMINATION OF METAL DEBRIS FROM IMPLANTS [J].
CASE, CP ;
LANGKAMER, VG ;
JAMES, C ;
PALMER, MR ;
KEMP, AJ ;
HEAP, PF ;
SOLOMON, L .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (05) :701-712