Fast-scanning high-flux microprobe for biological X-ray fluorescence microscopy and microXAS

被引:16
作者
Barrea, R. A. [1 ]
Gore, D. [1 ]
Kujala, N. [1 ]
Karanfil, C. [2 ]
Kozyrenko, S. [1 ]
Heurich, R. [1 ]
Vukonich, M. [1 ]
Huang, R. [3 ]
Paunesku, T. [4 ,5 ]
Woloschak, G. [4 ,5 ]
Irving, T. C. [1 ]
机构
[1] IIT, Dept Biol Chem & Phys Sci, BioCAT, Chicago, IL 60616 USA
[2] Mugla Univ, Fac Arts & Sci, Dept Phys, TR-48187 Kotekli Mugla, Turkey
[3] Hauptman Woodward Inst, IMCA CAT, Buffalo, NY 14203 USA
[4] Northwestern Univ, Feinberg Sch Med, Dept Radiat Oncol, Chicago, IL 60611 USA
[5] Northwestern Univ, Feinberg Sch Med, Dept Radiol, Chicago, IL 60611 USA
关键词
synchrotron X-ray imaging; microprobe; microarrays; microXAS; microspectroscopy; biological samples; tissues; BENT SILICON CRYSTAL; ABSORPTION SPECTROSCOPY; LAUE GEOMETRY; SYNCHROTRON; BEAMLINE; COPPER; FACILITY; REVEALS; CANCER; APS;
D O I
10.1107/S0909049510016869
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
There is a growing interest in the biomedical community in obtaining information concerning the distribution and local chemical environment of metals in tissues and cells. Recently, biological X-ray fluorescence microscopy (XFM) has emerged as the tool of choice to address these questions. A fast-scanning high-flux X-ray microprobe, built around a recently commissioned pair of 200 mm-long Rh-coated silicon Kirkpatrick-Baez mirrors, has been constructed at BioCAT beamline 18ID at the Advanced Photon Source. The new optical system delivers a flux of 1.3 x 10(12) photons s(-1) into a minimum focal spot size of similar to 3-5 mu m FWHM. A set of Si drift detectors and bent Laue crystal analyzers may be used in combination with standard ionization chambers for X-ray fluorescence measurements. BioCAT's scanning software allows fast continuous scans to be performed while acquiring and storing full multichannel analyzer spectra per pixel on-the-fly with minimal overhead time (< 20 ms per pixel). Together, the high-flux X-ray microbeam and the rapid-scanning capabilities of the BioCAT beamline allow the collection of XFM and micro X-ray absorption spectroscopy (microXAS) measurements from as many as 48 tissue sections per day. This paper reports the commissioning results of the new instrument with representative XFM and microXAS results from tissue samples.
引用
收藏
页码:522 / 529
页数:8
相关论文
共 28 条
[1]  
Barrea RA, 2005, PHYS SCRIPTA, VT115, P867
[2]  
BARREA RA, 2010, UNPUB
[3]  
BARREA RA, 2006, IPAP C SERIES, V7, P230
[4]   Synchrotron X-Ray Imaging Reveals a Correlation of Tumor Copper Speciation With Clioquinol's Anticancer Activity [J].
Barrea, Raul A. ;
Chen, Di ;
Irving, Thomas C. ;
Dou, Q. Ping .
JOURNAL OF CELLULAR BIOCHEMISTRY, 2009, 108 (01) :96-105
[5]   High-flux hard X-ray microbeam using a single-bounce capillary with doubly focused undulator beam [J].
Barrea, Raul A. ;
Huang, Rong ;
Cornaby, Sterling ;
Bilderback, Donald H. ;
Irving, Thomas C. .
JOURNAL OF SYNCHROTRON RADIATION, 2009, 16 :76-82
[6]   Disulfiram, a clinically used anti-alcoholism drug and copper-binding agent, induces apoptotic cell death in breast cancer cultures and xenografts via inhibition of the proteasome activity [J].
Chen, Di ;
Cui, Qiuzhi Cindy ;
Yang, Huanjie ;
Dou, Q. Ping .
CANCER RESEARCH, 2006, 66 (21) :10425-10433
[7]   Dynamically figured Kirkpatrick Baez x-ray micro-focusing optics [J].
Eng, PJ ;
Newville, M ;
Rivers, ML ;
Sutton, SR .
X-RAY MICROFOCUSING: APPLICATIONS AND TECHNIQUES, 1998, 3449 :145-156
[8]   The localisation and micro-mapping of copper and other trace elements in breast tumours using a synchrotron micro-XRF system [J].
Farquharson, M. J. ;
Geraki, K. ;
Falkenberg, G. ;
Leek, R. ;
Harris, A. .
APPLIED RADIATION AND ISOTOPES, 2007, 65 (02) :183-188
[9]   X-ray fluorescence microscopy reveals large-scale relocalization and extracellular translocation of cellular copper during angiogenesis [J].
Finney, Lydia ;
Mandava, Suneeta ;
Ursos, Lyann ;
Zhang, Wen ;
Rodi, Diane ;
Vogt, Stefan ;
Legnini, Daniel ;
Maser, Jorg ;
Ikpatt, Francis ;
Olopade, Olufunmilayo I. ;
Glesne, David .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 2007, 104 (07) :2247-2252
[10]   The BioCAT undulator beamline 18ID: a facility for biological non-crystalline diffraction and X-ray absorption spectroscopy at the Advanced Photon Source [J].
Fischetti, R ;
Stepanov, S ;
Rosenbaum, G ;
Barrea, R ;
Black, E ;
Gore, D ;
Heurich, R ;
Kondrashkina, E ;
Kropf, AJ ;
Wang, S ;
Zhang, K ;
Irving, TC ;
Bunker, GB .
JOURNAL OF SYNCHROTRON RADIATION, 2004, 11 :399-405