A new automated method for analysis of gated-SPECT images based on a three-dimensional heart shaped model

被引:24
作者
Lomsky, M
Richter, J
Johansson, L
El-Ali, H
Åström, K
Ljungberg, M
Edenbrandt, L [1 ]
机构
[1] Malmo Univ Hosp, Dept Clin Physiol, SE-20502 Malmo, Sweden
[2] Sahlgrenska Univ Hosp, Dept Clin Physiol, Gothenburg, Sweden
[3] Lund Inst Technol, Dept Math, Lund, Sweden
[4] Lund Univ, Dept Radiat Phys, S-22101 Lund, Sweden
关键词
computer-assisted diagnosis; gated-SPECT; image processing; radionuclide imaging;
D O I
10.1111/j.1475-097X.2005.00619.x
中图分类号
Q4 [生理学];
学科分类号
071003 ;
摘要
A new automated method for quantification of left ventricular function from gated-single photon emission computed tomography ( SPECT) images has been developed. The method for quantification of cardiac function (CAFU) is based on a heart shaped model and the active shape algorithm. The model contains statistical information of the variability of left ventricular shape. CAFU was adjusted based on the results from the analysis of five simulated gated-SPECT studies with well defined volumes of the left ventricle. The digital phantom NURBS-based Cardiac-Torso (NCAT) and the Monte-Carlo method SIMIND were used to simulate the studies. Finally CAFU was validated on ten rest studies from patients referred for routine stress/rest myocardial perfusion scintigraphy and compared with Cedar-Sinai quantitative gated-SPECT (QGS), a commercially available program for quantification of gated-SPECT images. The maximal differences between the CAFU estimations and the true left ventricular volumes of the digital phantoms were 11 ml for the end-diastolic volume (EDV), 3 ml for the end-systolic volume (ESV) and 3% for the ejection fraction (EF). The largest differences were seen in the smallest heart. In the patient group the EDV calculated using QGS and CAFU showed good agreement for large hearts and higher CAFU values compared with QGS for the smaller hearts. In the larger hearts, ESV was much larger for QGS than for CAFU both in the phantom and patient studies. In the smallest hearts there was good agreement between QGS and CAFU. The findings of this study indicate that our new automated method for quantification of gated-SPECT images can accurately measure left ventricular volumes and EF.
引用
收藏
页码:234 / 240
页数:7
相关论文
共 17 条
[1]   Left ventricular ejection fraction and volumes from gated single photon emission tomographic myocardial perfusion images: Comparison between two algorithms working in three-dimensional space [J].
Everaert, H ;
Bossuyt, A ;
Franken, PR .
JOURNAL OF NUCLEAR CARDIOLOGY, 1997, 4 (06) :472-476
[2]  
Faber TL, 1999, J NUCL MED, V40, P650
[3]  
FICARO E, 2003, J NUCL CARDIOL, V10, pS23
[4]  
GERMANO G, 1995, J NUCL MED, V36, P2138
[5]  
Go V, 2004, J NUCL MED, V45, P912
[6]   Electrocardiogram-gated single-photon emission computed tomography versus cardiac magnetic resonance imaging for the assessment of left ventricular volumes and ejection fraction - A meta-analysis [J].
Ioannidis, JPA ;
Trikalinos, TA ;
Danias, PG .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 39 (12) :2059-2068
[7]   Validation of 4D-MSPECT and QGS for quantification of left ventricular volumes and ejection fraction from gated 99mTc-MIBI SPET:: comparison with cardiac magnetic resonance imaging [J].
Lipke, CSA ;
Kühl, HP ;
Nowak, B ;
Kaiser, HJ ;
Reinartz, P ;
Koch, KC ;
Buell, U ;
Schaefer, WM .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2004, 31 (04) :482-490
[8]   A MONTE-CARLO PROGRAM FOR THE SIMULATION OF SCINTILLATION CAMERA CHARACTERISTICS [J].
LJUNGBERG, M ;
STRAND, SE .
COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 1989, 29 (04) :257-272
[9]  
Nakajima K, 2001, J NUCL MED, V42, P1571
[10]   Comparison between ECTb and QGS for assessment of left ventricular function from gated myocardial perfusion SPECT [J].
Nichols, K ;
Santana, CA ;
Folks, R ;
Krawczynska, E ;
Cooke, CD ;
Faber, TL ;
Bergmann, SR ;
Garcia, EV .
JOURNAL OF NUCLEAR CARDIOLOGY, 2002, 9 (03) :285-293