Electrocardiogram-gated single-photon emission computed tomography versus cardiac magnetic resonance imaging for the assessment of left ventricular volumes and ejection fraction - A meta-analysis

被引:90
作者
Ioannidis, JPA
Trikalinos, TA
Danias, PG
机构
[1] Beth Israel Deaconess Med Ctr, Div Cardiol, Dept Med, Boston, MA 02215 USA
[2] Harvard Univ, Sch Med, Boston, MA USA
[3] Univ Ioannina, Sch Med, Dept Hyg & Epidemiol, Clin Trials & Evidence Based Med Unit, GR-45110 Ioannina, Greece
[4] Univ Ioannina, Sch Med, Dept Hyg & Epidemiol, Clin & Mol Epidemiol Unit, GR-45110 Ioannina, Greece
[5] Tufts Univ, Sch Med, Dept Med, Boston, MA 02111 USA
关键词
D O I
10.1016/S0735-1097(02)01882-X
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVES The purpose of this study was to evaluate the accuracy of electrocardiogram (ECG)-gated single-photon emission computed tomography (SPECT) for assessment of left ventricular (LV) end-diastolic volume (EDV), end-systolic volume (ESV) and ejection fraction (EF) compared with the gold standard of cardiac magnetic resonance imaging (MRI). BACKGROUND Several comparisons of ECG-gated SPECT with cardiac MRI have been performed for evaluation of LV volumes and EF, but each has considered few subjects, thus leaving uncertainty about the frequency of discrepancies between the two methods. METHODS We performed a meta-analysis of data on 164 subjects from nine studies comparing ECG-gated SPECT versus cardiac MRI. Data were pooled in correlation and regression analyses relating ECG-gated SPECT and cardiac MRI measurements. The frequency of discrepancies of at least 30 ml in EDV, 20 ml in ESV and 5% or 10% in EF and concordance for EF less than or equal to40% versus >40% were determined. RESULTS There was an overall excellent correlation between ECG-gated SPECT and cardiac MRI for EDV (r = 0.89), ESV (r = 0.92) and EF (r = 0.87). However, rates of discrepancies for individual subjects were considerable (37% [95% confidence interval [CI], 26% to 50%] for at least 30 ml in EDV; 35% [95% CI, 23% to 49%] for at least 20 ml in ESV; 52% [95% CI, 37% to 63%] for at least 5% in EF; and 23% [95% CI, 11% to 42%] for at least 10% in EF). The misclassification rate for the 40% EF cutoff was 11%. CONCLUSIONS Electrocardiogram-gated SPECT measurements of EDV, ESV and EF show high correlation with cardiac MRI measurements, but substantial errors may occur in individual patients. Electrocardiogram-gated SPECT offers useful functional information, but cardiac MRI should be used when accurate measurement is required.
引用
收藏
页码:2059 / 2068
页数:10
相关论文
共 35 条
[1]   Left ventricular function: Correlation of quantitative gated SPECT and MR imaging over a wide range of values [J].
Bavelaar-Croon, CDL ;
Kayser, HWM ;
van der Wall, EE ;
de Roos, A ;
Dibbets-Schneider, P ;
Pauwels, EKJ ;
Germano, G ;
Atsma, DE .
RADIOLOGY, 2000, 217 (02) :572-575
[2]   Comparison of gated single-photon emission computed tomography with magnetic resonance imaging for evaluation of left ventricular function in ischemic cardiomyopathy [J].
Bax, JJ ;
Lamb, H ;
Dibbets, P ;
Pelikan, H ;
Boersma, E ;
Viergever, EP ;
Germano, G ;
Vliegen, HW ;
de Roos, A ;
Pauwels, EKJ ;
Van der Wall, EE .
AMERICAN JOURNAL OF CARDIOLOGY, 2000, 86 (12) :1299-1305
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]   LEFT-VENTRICULAR VOLUME MEASUREMENT USING CARDIAC AXIS NUCLEAR-MAGNETIC-RESONANCE IMAGING - VALIDATION BY CALIBRATED VENTRICULAR ANGIOGRAPHY [J].
CRANNEY, GB ;
LOTAN, CS ;
DEAN, L ;
BAXLEY, W ;
BOUCHARD, A ;
POHOST, GM .
CIRCULATION, 1990, 82 (01) :154-163
[5]   Combined assessment of myocardial perfusion and left ventricular function with exercise technetium-99m sestamibi gated single-photon emission computed tomography can differentiate between ischemic and nonischemic dilated cardiomyopathy [J].
Danias, PG ;
Ahlberg, AW ;
Clark, BA ;
Messineo, F ;
Levine, MG ;
McGill, CC ;
Mann, A ;
Clive, J ;
Dougherty, JE ;
Waters, DD ;
Heller, GV .
AMERICAN JOURNAL OF CARDIOLOGY, 1998, 82 (10) :1253-1258
[6]  
DEPUEY E, 1995, J NUCL MED, V36, P952
[7]   Comparison of Tc-99m sestamibi and Tl-201 gated perfusion SPECT [J].
DePuey, EG ;
Parmett, S ;
Ghesani, M ;
Rozanski, A ;
Nichols, K ;
Salensky, H .
JOURNAL OF NUCLEAR CARDIOLOGY, 1999, 6 (03) :278-285
[8]  
DEPUEY EG, 1993, J NUCL MED, V34, P1871
[9]  
Faber TL, 1999, J NUCL MED, V40, P650
[10]   Evaluation of left ventricular endocardial volumes and ejection fractions computed from gated perfusion SPECT with magnetic resonance imaging: Comparison of two methods [J].
Faber, TL ;
Vansant, JP ;
Pettigrew, RI ;
Galt, JR ;
Blais, M ;
Chatzimavroudis, G ;
Cooke, CD ;
Folks, RD ;
Waldrop, SM ;
Gurtler-Krawczynska, E ;
Wittry, MD ;
Garcia, EV .
JOURNAL OF NUCLEAR CARDIOLOGY, 2001, 8 (06) :645-651