Quantitative evaluation of the performance of different electrical tomography devices

被引:26
作者
Hahn, G
Thiel, F
Dudykevych, T
Frerichs, I
Gersing, E
Schröder, T
Hartung, C
Hellige, G
机构
[1] Univ Gottingen, Dept Anaesthesiol Res, Ctr Anaesthesiol Emergency & Intens Care Med TL 1, D-37075 Gottingen, Germany
[2] Leibniz Univ Hannover, Inst Biomed Engn & Hosp Engn, D-30167 Hannover, Germany
来源
BIOMEDIZINISCHE TECHNIK | 2001年 / 46卷 / 04期
关键词
electrical impedance tomography; EIT; impedance; phantom;
D O I
10.1515/bmte.2001.46.4.91
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Two widely used electrical tomography systems, the Sheffield Mark I and the DAS-01P, were quantitatively evaluated and compared to the newly developed Goe-MF system. The performance was quantified using a hardware phantom which closely matches the real input and transfer impedances of the human thorax and allows measurements equivalent to different states of lung inflation. Our results demonstrate that adequate averaging is necessary for noise reduction for the Mark I and especially for the DAS-01P system to get meaningful results even in visualizing maximal respiratory manoeuvres. The Goe-MF system showed a notably improved signal-to-noise ratio which allows also dynamic measurements at low levels of lung volume changes, e.g., in intensive care lung injury patients.
引用
收藏
页码:91 / 95
页数:5
相关论文
共 20 条
[1]  
Barber D C, 1987, Clin Phys Physiol Meas, V8 Suppl A, P47, DOI 10.1088/0143-0815/8/4A/006
[2]   APPLIED POTENTIAL TOMOGRAPHY [J].
BARBER, DC ;
BROWN, BH .
JOURNAL OF PHYSICS E-SCIENTIFIC INSTRUMENTS, 1984, 17 (09) :723-733
[3]  
Brown B H, 1987, Clin Phys Physiol Meas, V8 Suppl A, P91, DOI 10.1088/0143-0815/8/4A/012
[4]   APPLIED POTENTIAL TOMOGRAPHY - POSSIBLE CLINICAL-APPLICATIONS [J].
BROWN, BH ;
BARBER, DC ;
SEAGAR, AD .
CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT, 1985, 6 (02) :109-121
[5]  
BROWN BH, 1990, P COP M EL IMP TOM C, P162
[6]   Heterogeneous distribution of pulmonary ventilation in intensive care patients detected by functional electrical impedance tomography [J].
Frerichs, I ;
Golisch, W ;
Hahn, G ;
Kurpitz, M ;
Burchardi, H ;
Hellige, G .
JOURNAL OF INTENSIVE CARE MEDICINE, 1998, 13 (04) :168-173
[7]   Monitoring perioperative changes in distribution of pulmonary ventilation by functional electrical impedance tomography [J].
Frerichs, I ;
Hahn, G ;
Golisch, W ;
Kurpitz, M ;
Burchardi, H ;
Hellige, G .
ACTA ANAESTHESIOLOGICA SCANDINAVICA, 1998, 42 (06) :721-726
[8]   Monitoring regional lung ventilation by functional electrical impedance tomography during assisted ventilation [J].
Frerichs, I ;
Hahn, G ;
Schiffmann, H ;
Berger, C ;
Hellige, G .
ELECTRICAL BIOIMPEDANCE METHODS: APPLICATIONS TO MEDICINE AND BIOTECHNOLOGY, 1999, 873 :493-505
[9]   A phantom for electrical impedance tomography [J].
Griffiths, H. .
Clinical Physics and Physiological Measurement, 1988, 9 (SUPPL. A) :15-20
[10]   A COLE PHANTOM FOR EIT [J].
GRIFFITHS, H .
PHYSIOLOGICAL MEASUREMENT, 1995, 16 :A29-A38