In-vitro assessment of a registration protocol for image guided implant dentistry

被引:86
作者
Birkfellner, W
Solar, P
Gahleitner, A
Huber, K
Kainberger, F
Kettenbach, J
Homolka, P
Diemling, M
Watzek, G
Bergmann, H
机构
[1] Univ Vienna, Gen Hosp, Dept Biomed Engn & Phys, A-1090 Vienna, Austria
[2] Univ Vienna, Sch Dent, Dept Oral Surg, A-1090 Vienna, Austria
[3] Univ Vienna, Gen Hosp, Dept Osteoradiol, A-1090 Vienna, Austria
[4] Univ Vienna, Gen Hosp, Dept Angiog & Intervent Radiol, A-1090 Vienna, Austria
[5] Ludwig Boltzmann Inst Nucl Med & Endocrinol, Vienna, Austria
关键词
image guided therapy; computer aided surgery; implant dentistry; registration; accuracy;
D O I
10.1034/j.1600-0501.2001.012001069.x
中图分类号
R78 [口腔科学];
学科分类号
1003 [口腔医学];
摘要
In this study a computer aided navigation In this study a computer aided navigation technique for accurate positioning of oral implants was assessed. An optical tracking system with specially designed tools for monitoring the position of surgical instruments relative to the patient was used to register 5 partially or completely edentulous jaw models. Besides the accuracy of the tracking system, the precision of localizing a specific position on 3-dimensional preoperative imagery is governed by the registration algorithm which conveys the coordinate system of the preoperative computed tomography (CT) scan to the actual patient position. Two different point-to-point registration algorithms were compared their suitability for this application. The accuracy was determined separately for the localization error of the position measurement hardware (fiducial localization error FLE) and the error as reported by the registration algorithm (fiducial registration error FRE). The overall error of the navigation procedure was determined as the localization error of additional landmarks (steel spheres, 0.5 mm diameter) after registration (target registration error - TRE). Images of the jaw models were obtained using a high resolution CT scan (1.5 mm slice thickness, 1 mm table feed, incremental scanning, 120 kV, 150 mAs, 512 . 512 matrix, FOV 120 mm). The accuracy of the position measurement probes was 0.69+/-0.15 mm (FLE). Using 3 implanted fiducial markers, FRE was 0.71+/-0.12 mm on average ano 1.00+/-0.13 mm maximum. TRE was found to be 1.23+/-0.28 mm average and 1.87+/-0.47 mm maximum. increasing the number of fiducial markers to a total of 5 did not significantly improve precision. Furthermore it was found that a registration algorithm based on solving an eigenvalue problem is the superior approach for point-to-point matching in terms of mathematical stability. The was 0.71+/-0.12 mm on average and 1.00+/-0.13 mm maximum. TRE was found to be experimental results indicate that positioning accuracy of oral implants may benefit computer aided intraoperative navigation. The accuracy achieved compares well to the resolution of the CT scan used. Further development of point-to-point/point-to-surFace registration methods and tracking hardware has the potential to improve the precision of the method even further. Our system has potential to reduce the intraoperative risk of causing damage to critical anatomic structures, to minimize the efforts in prosthetic modelling, and to simplify the task of transferring preoperative planning data precisely to the operating room in general.
引用
收藏
页码:69 / 78
页数:10
相关论文
共 51 条
[1]
[Anonymous], 3 DIMENSIONAL BIOMED
[2]
LEAST-SQUARES FITTING OF 2 3-D POINT SETS [J].
ARUN, KS ;
HUANG, TS ;
BLOSTEIN, SD .
IEEE TRANSACTIONS ON PATTERN ANALYSIS AND MACHINE INTELLIGENCE, 1987, 9 (05) :699-700
[3]
DENTAL IMPLANT TREATMENT PLANNING WITH REFORMATTED COMPUTED-TOMOGRAPHY [J].
BESIMO, C ;
LAMBRECHT, JT ;
NIDECKER, A .
DENTOMAXILLOFACIAL RADIOLOGY, 1995, 24 (04) :264-267
[4]
A METHOD FOR REGISTRATION OF 3-D SHAPES [J].
BESL, PJ ;
MCKAY, ND .
IEEE TRANSACTIONS ON PATTERN ANALYSIS AND MACHINE INTELLIGENCE, 1992, 14 (02) :239-256
[5]
Systematic distortions in magnetic position digitizers [J].
Birkfellner, W ;
Watzinger, F ;
Wanschitz, F ;
Enislidis, G ;
Kollmann, C ;
Rafolt, D ;
Nowotny, R .
MEDICAL PHYSICS, 1998, 25 (11) :2242-2248
[6]
Calibration of tracking systems in a surgical environment [J].
Birkfellner, W ;
Watzinger, F ;
Wanschitz, F ;
Ewers, R ;
Bergmann, H .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (05) :737-742
[7]
BRANEMARK P I, 1969, Scandinavian Journal of Plastic and Reconstructive Surgery, V3, P81
[8]
Chassat F, 1998, LECT NOTES COMPUT SC, V1496, P277, DOI 10.1007/BFb0056211
[9]
3-D ultrasound imaging: A review [J].
Fenster, A ;
Downey, DB .
IEEE ENGINEERING IN MEDICINE AND BIOLOGY MAGAZINE, 1996, 15 (06) :41-51
[10]
Predicting error in rigid-body point-based registration [J].
Fitzpatrick, JM ;
West, JB ;
Maurer, CR .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (05) :694-702