Calibration of tracking systems in a surgical environment

被引:112
作者
Birkfellner, W
Watzinger, F
Wanschitz, F
Ewers, R
Bergmann, H
机构
[1] Univ Vienna, Gen Hosp Vienna, Dept Biomed Engn & Phys, A-1090 Vienna, Austria
[2] Univ Vienna, Sch Med, Clin Oral & Maxillofacial Surg, A-1090 Vienna, Austria
[3] Ludwig Boltzmann Inst Nucl Med & Endocrinol, Vienna, Austria
基金
奥地利科学基金会;
关键词
computer-aided surgery (CAS); electromagnetic tracking system; frameless stereotaxy; hybrid tracking system;
D O I
10.1109/42.736028
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
The purpose of this paper was to assess to what extent an optical tracking system (OTS) used for position determination in computer-aided surgery (CAS) can be enhanced by combining it with a direct current (dc) driven electromagnetic tracking system (EMTS), The main advantage of the EMTS is the fact that it is not dependent on a free line-of-sight. Unfortunately, the accuracy of the EMTS is highly affected by nearby ferromagnetic materials. We have explored, to what extent the influence of the metallic equipment in the operating room (OP) Can be compensated by collecting precise information on the nonlinear local error in the EMTS by using the OTS for setting up a calibration look-up table. After calibration of the EMTS and registration of the sensor systems in the OR we have found the average euclidean deviation in position readings between the de tracker and the OTS reduced from 2.9 +/- 1.0 mm to 2.1 +/- 0.8 mm within a half-sphere of 530-mm radius around the magnetic field emitter. Furthermore we have found the calibration to be stable after re-registration of the sensors under varying conditions such as different heights of the OR table and varying positions of the OR equipment over a longer time interval. These results encourage the further development of a hybrid magnetooptical tracker for computer-aided surgery where the electromagnetic tracker acts as an auxiliary source of position information for the optical system. Strategies for enhancing the reliability of the proposed hybrid magnetooptic tracker by detecting artifacts induced by mobile ferromagnetic objects such as surgical tools are discussed.
引用
收藏
页码:737 / 742
页数:6
相关论文
共 12 条
[1]   LEAST-SQUARES FITTING OF 2 3-D POINT SETS [J].
ARUN, KS ;
HUANG, TS ;
BLOSTEIN, SD .
IEEE TRANSACTIONS ON PATTERN ANALYSIS AND MACHINE INTELLIGENCE, 1987, 9 (05) :699-700
[2]   Systematic distortions in magnetic position digitizers [J].
Birkfellner, W ;
Watzinger, F ;
Wanschitz, F ;
Enislidis, G ;
Kollmann, C ;
Rafolt, D ;
Nowotny, R .
MEDICAL PHYSICS, 1998, 25 (11) :2242-2248
[3]  
Birkfellner W, 1998, INT CONGR SER, V1165, P927
[4]   VARIABLES AFFECTING THE ACCURACY OF STEREOTAXIC LOCALIZATION USING COMPUTERIZED-TOMOGRAPHY [J].
BUCHOLZ, RD ;
HO, HW ;
RUBIN, JP .
JOURNAL OF NEUROSURGERY, 1993, 79 (05) :667-673
[5]  
GEDROVICS A, 1997, THESIS TU VIENNA
[6]  
GOERSS SJ, 1994, STEREOT FUNCT NEUROS, V63, P89, DOI 10.1159/000100299
[8]   Magnetic tracker calibration for improved augmented reality registration [J].
Livingston, MA ;
State, A .
PRESENCE-TELEOPERATORS AND VIRTUAL ENVIRONMENTS, 1997, 6 (05) :532-546
[9]   Registration of head volume images using implantable fiducial markers [J].
Maurer, CR ;
Fitzpatrick, JM ;
Wang, MY ;
Galloway, RL ;
Maciunas, RJ ;
Allen, GS .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1997, 16 (04) :447-462
[10]  
State A., 1996, Computer Graphics Proceedings. SIGGRAPH '96, P429, DOI 10.1145/237170.237282