A prototype small CdTe gamma camera for radioguided surgery and other imaging applications

被引:52
作者
Tsuchimochi, M
Sakahara, H
Hayama, K
Funaki, M
Ohno, R
Shirahata, T
Orskaug, T
Maehlum, G
Yoshioka, K
Nygard, E
机构
[1] Nippon Dent Univ, Sch Dent Niigata, Dept Oral & Maxillofacial Radiol, Niigata 9518580, Japan
[2] Hamamatsu Univ Sch Med, Dept Radiol, Hamamatsu, Shizuoka 43131, Japan
[3] Acrorad Co Ltd, Tokyo, Japan
[4] Integrated Detector & Elect AS, Hovik, North Ireland
关键词
CdTe; gamma camera; semiconductor; sentinel lymph node biopsy; radioguided surgery;
D O I
10.1007/s00259-003-1301-3
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 [临床医学]; 100207 [影像医学与核医学]; 1009 [特种医学];
摘要
Gamma probes have been used for sentinel lymph node biopsy in melanoma and breast cancer. However, these probes can provide only radioactivity counts and variable pitch audio output based on the intensity of the detected radioactivity. We have developed a small semiconductor gamma camera (SSGC) that allows visualisation of the size, shape and location of the target tissues. This study is designed to characterise the performance of the SSGC for radioguided surgery of metastatic lesions and for other imaging applications amenable to the smaller format of this prototype imaging system. The detector head had 32 cadmium telluride semiconductor arrays with a total of 1,024 pixels, and with application-specific integrated circuits (ASICs) and a tungsten collimator. The entire assembly was encased in a lead housing measuring 152 mmx166 mmx65 rum. The effective visual field was 44.8 mmx44.8 mm. The energy resolution and imaging aspects were tested. Two spherical 5-mm- and 15-mm-diameter technetium-99m radioactive sources that had activities of 0.15 MBq and 100 MBq, respectively, were used to simulate a sentinel lymph node and an injection site. The relative detectability of these foci by the new detector and a conventional scintillation camera was studied. The prototype was also examined in a variety of clinical applications. Energy resolution [full-width at half-maximum (FWHM)] for a single element at the centre of the field of view was 4.2% at 140 keV (Tc-99m), and the mean energy resolution of the CdTe detector arrays was approximately 7.8%. The spatial resolution, represented by FWHM had a mean value of 1.56+/-0.05 mm. Simulated node foci could be visualised clearly by the SSGC using a 15-s acquisition time. In preliminary clinical tests, the SSGC successfully imaged diseases in a variety of tissues, including salivary and thyroid glands, temporomandibular joints and sentinel lymph nodes. The SSGC has significant potential for diagnosing diseases and facilitating subsequent radioguided surgery.
引用
收藏
页码:1605 / 1614
页数:10
相关论文
共 43 条
[1]
ALAVI A, 1978, J NUCL MED, V19, P422
[2]
CADMIUM TELLURIDE MATRIX GAMMA-CAMERA [J].
ALLISON, JD .
MEDICAL PHYSICS, 1980, 7 (03) :202-206
[3]
Progress in developing focal-plane-multiplexer readout for large CdZnTe arrays for nuclear medicine applications [J].
Barber, HB ;
Marks, DG ;
Apotovsky, BA ;
Augustine, FL ;
Barrett, HH ;
Butler, JF ;
Dereniak, EL ;
Doty, FP ;
Eskin, JD ;
Hamilton, WJ ;
Matherson, KJ ;
Venzon, JE ;
Woolfenden, JM ;
Young, ET .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 380 (1-2) :262-265
[4]
TIME-DEPENDENT POLARIZATION OF CDTE GAMMA-RAY DETECTORS [J].
BELL, RO ;
ENTINE, G ;
SERREZE, HB .
NUCLEAR INSTRUMENTS & METHODS, 1974, 117 (01) :267-271
[5]
A method to evaluate intra-operative gamma probes for sentinel lymph node localisation [J].
Britten, AJ .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1999, 26 (02) :76-83
[6]
Burak W E Jr, 1996, Surg Technol Int, V5, P259
[8]
De Cicco C, 1998, J NUCL MED, V39, P2080
[9]
DUSI W, 1993, PHYS MEDICA, V9, P195
[10]
ASSESSMENT OF BONE REMODELING IN THE TEMPOROMANDIBULAR-JOINT BY SERIAL UPTAKE MEASUREMENT OF TECHNETIUM 99M LABELED METHYLENE DIPHOSPHONATE WITH A CADMIUM TELLURIDE PROBE [J].
ENGELKE, W ;
TSUCHIMOCHI, M ;
RUTTIMANN, UE ;
HOSAIN, F .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTICS, 1991, 71 (03) :357-363