Radiostereometric analysis comparison of wear of highly cross-linked polyethylene against 36-vs 28-mm femoral heads

被引:71
作者
Bragdon, Charles R.
Greene, Meridith E.
Freiberg, Andrew A.
Harris, William H.
Malchau, Henrik
机构
[1] Massachusetts Gen Hosp, Boston, MA 02114 USA
[2] Massachusetts Gen Hosp, Harvard Med Sch, Orthopaed Biomech & Biomat Lab, Boston, MA USA
[3] Massachusetts Gen Hosp, Harvard Med Sch, Dept Orthopaed Surg, Adult Reconstruct Unit, Boston, MA USA
[4] Massachusetts Gen Hosp, Harvard Med Sch, Dept Orthopaed Surg, Boston, MA USA
关键词
RSA; total hip arthroplasty; wear; hip; outcomes; polyethylene; MOLECULAR-WEIGHT POLYETHYLENE; TOTAL HIP-ARTHROPLASTY; RADIO STEREOMETRIC ANALYSIS; 3RD-BODY WEAR; PENETRATION; LARGER; SIZES;
D O I
10.1016/j.arth.2007.03.009
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
100224 [整形外科学];
摘要
This study used radiostereometric analysis (RSA) to compare the femoral head penetration of 28- vs 36-mm-diameter femoral heads into highly cross-linked polyethylene in 2 groups of total hip arthroplasty patients. Thirty patients were enrolled in this RSA study using highly cross-linked polyethylene (Longevity, Zimmer Inc, Warsaw, Idaho) against either 28- or 36-mm-diameter cobalt chrome femoral heads. At 3-year follow-up, there was no significant difference in the total average femoral head penetration, including both creep and wear, using 3 methods of RSA measurement between the 2 groups. Importantly, after bedding-in, there was no further significant increase in the amount of femoral head penetration (ie, wear) with either head size between years 1 and 3. There were no radiographic signs of lysis or radiolucencies at a minimum 3-year follow-up.
引用
收藏
页码:125 / 129
页数:5
相关论文
共 23 条
[1]
RSA wear measurements with or without markers in total hip arthroplasty [J].
Börlin, N ;
Röhrl, SM ;
Bragdon, CR .
JOURNAL OF BIOMECHANICS, 2006, 39 (09) :1641-1650
[2]
Steady-state penetration rates of electron beam-irradiated, highly cross-linked polyethylene at an average 45-month follow-up [J].
Bragdon, Charles R. ;
Barrett, Susan ;
Martell, John M. ;
Greene, Meridith E. ;
Malchau, Henrik ;
Harris, William H. .
JOURNAL OF ARTHROPLASTY, 2006, 21 (07) :935-943
[3]
Comparison of two digital radio stereometric analysis methods in the determination of femoral head penetration in a total hip replacement phantom [J].
Bragdon, CR ;
Estok, DM ;
Malchau, H ;
Kärrholm, J ;
Yuan, XH ;
Bourne, R ;
Veldhoven, J ;
Harris, WH .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2004, 22 (03) :659-664
[4]
Experimental assessment of precision and accuracy of radio stereometric analysis for the determination of polyethylene wear in a total hip replacement model [J].
Bragdon, CR ;
Malchau, H ;
Yuan, XH ;
Perinchief, R ;
Kärrholm, J ;
Börlin, N ;
Estok, DM ;
Harris, WH .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2002, 20 (04) :688-695
[5]
Third-body wear testing of a highly cross-linked acetabular liner - The effect of large femoral head size in the presence of particulate poly (methyl-methacrylate) debris [J].
Bragdon, CR ;
Jasty, M ;
Muratoglu, OK ;
Harris, WH .
JOURNAL OF ARTHROPLASTY, 2005, 20 (03) :379-385
[6]
Third-body wear of highly cross-linked polyethylene in a hip simulator [J].
Bragdon, CR ;
Jasty, M ;
Muratoglu, OK ;
O'Connor, DO ;
Harris, WH .
JOURNAL OF ARTHROPLASTY, 2003, 18 (05) :553-561
[7]
BRAGDON CR, 2004, THESIS GOTEBORG U
[8]
Range of motion and stability in total hip arthroplasty with 28-, 32-, 38-, and 44-mm femoral head sizes - An in vitro study [J].
Burroughs, BR ;
Hallstrom, B ;
Golladay, GJ ;
Hoeffel, D ;
Harris, WH .
JOURNAL OF ARTHROPLASTY, 2005, 20 (01) :11-19
[9]
Femoral head sizes larger than 32 mm against highly cross-linked polyethylene [J].
Burroughs, BR ;
Rubash, HE ;
Harris, WH .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2002, (405) :150-157
[10]
Larger femoral heads: A triumph of hope over reason! In the affirmative [J].
Callaghan, JJ ;
Brown, TD ;
Pedersen, DR ;
Johnston, RC .
JOURNAL OF ARTHROPLASTY, 2003, 18 (03) :82-84