Coronary arteries: assessment of image quality and optimal reconstruction window in retrospective ECG-gated multislice CT at 375-ms gantry rotation time

被引:52
作者
Hamoir, XL
Flohr, T
Hamoir, V
Labaki, L
Tricquet, JY
Duhamel, A
Kirsch, J
机构
[1] Clin Notre Dame, Dept Med Imaging, B-7500 Tournai, Belgium
[2] Siemens Med Syst Inc, Forchheim, Germany
[3] Clin Notre Dame, Dept Cardiol, B-7500 Tournai, Belgium
[4] Univ Lille, Dept Biostat, Fac Med, Lille, France
关键词
multislice helical CT; coronary artery; ECG gating;
D O I
10.1007/s00330-004-2541-8
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Our objective was to evaluate the image quality of a 16-slice CT system with a rotation time of 375 ms in the assessment of coronary arteries. One hundred patients underwent iodine-enhanced CT coronary angiography within a single breath hold. Images were reconstructed in diastole, 300, 350 400, 450, 500 and 550 ms prior to the onset of the next R-wave using absolute reverse retrospective ECG gating. The 15 coronary seaments of the AHA classification were consensually reviewed by two radiologists. On the whole, best quality imaging was obtained with reconstruction intervals of -350 ins and 400 ms in hi h percentages of each segment (P<0.0001). Only 6.2% of the arteries with a diameter greater than or equal to 1.5 rum were not assessable because of extensive calcifications (3.9%), cardiac motion artifacts (1.9%), lack of enhancement (0.2%) and stent artifacts (0.3%). In patients with a heart rate above 70 beats per minute, the percentage of assessable segments decreased to 88%, while at a lower heart rate it increased to 95%. In 61% of the patients, all segments were assessable. In conclusion, this generation of CT technique may allow visualization of coronary arteries with a low percentage of non-assessable segments.
引用
收藏
页码:296 / 304
页数:9
相关论文
共 18 条
[1]  
*AD HOC COMM GRAD, 1975, CIRCULATION, V51, P5
[2]   Multislice CT coronary angiography: Evaluation of an automatic vessel detection tool [J].
Dewey, M ;
Schnapauff, D ;
Laule, M ;
Lembcke, A ;
Borges, AC ;
Rutsch, W ;
Hamm, B ;
Rogalla, P .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2004, 176 (04) :478-483
[3]   Heart rate adaptive optimization of spatial and temporal resolution for electrocardiogram-gated multislice spiral CT of the heart [J].
Flohr, T ;
Ohnesorge, B .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2001, 25 (06) :907-923
[4]   Advances in cardiac imaging with 16-section CT systems [J].
Flohr, TG ;
Schoepf, UJ ;
Kuettner, A ;
Halliburton, S ;
Bruder, H ;
Suess, C ;
Schmidt, B ;
Hofmann, L ;
Yucel, EK ;
Schaller, S ;
Ohnesorge, BM .
ACADEMIC RADIOLOGY, 2003, 10 (04) :386-401
[5]   Noninvasive visualization of coronary arteries using contrast-enhanced multidetector CT: Influence of heart rate on image quality and stenosis detection [J].
Giesler, T ;
Baum, U ;
Ropers, D ;
Ulzheimer, S ;
Wenkel, E ;
Mennicke, M ;
Bautz, W ;
Kalender, WA ;
Daniel, WG ;
Achenbach, S .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2002, 179 (04) :911-916
[6]   Heart-rate-adapted image reconstruction in multidetector-row cardiac CT: influence of physiological and technical prerequisite on image quality [J].
Herzog, C ;
Abolmaali, N ;
Balzer, JO ;
Baunach, S ;
Ackermann, H ;
Dogan, S ;
Britten, MB ;
Vogl, TJ .
EUROPEAN RADIOLOGY, 2002, 12 (11) :2670-2678
[7]   Multislice helical CT of the heart with retrospective ECG gating: reduction of radiation exposure by ECG-controlled tube current modulation [J].
Jakobs, TF ;
Becker, CR ;
Ohnesorge, B ;
Flohr, T ;
Suess, C ;
Schoepf, UJ ;
Reiser, MF .
EUROPEAN RADIOLOGY, 2002, 12 (05) :1081-1086
[8]   Individually weight-adapted examination protocol in retrospectively ECG-gated MSCT of the heart [J].
Jung, B ;
Mahnken, AH ;
Stargardt, A ;
Simon, J ;
Flohr, TG ;
Schaller, S ;
Koos, R ;
Günther, RW ;
Wildberger, JE .
EUROPEAN RADIOLOGY, 2003, 13 (12) :2560-2566
[9]   MDCT:: cardiology indications [J].
Kopp, AF ;
Küttner, A ;
Trabold, T ;
Heuschmid, M ;
Schröder, S ;
Claussen, CD .
EUROPEAN RADIOLOGY, 2003, 13 :M102-M115
[10]   Multidetector-row CT cardiac imaging with 4 and 16 slices for coronary CTA and imaging of atherosclerotic plaques [J].
Kopp, AF ;
Küttner, A ;
Heuschmid, M ;
Schröder, S ;
Ohnesorge, B ;
Claussen, CD .
EUROPEAN RADIOLOGY, 2002, 12 (Suppl 2) :S17-S24