A comparison of electrochemiluminescence and flow cytometry for the detection of natural latex-specific human immunoglobulin E

被引:13
作者
Kobrynski, L
Tanimune, L
Pawlowski, NA
Douglas, SD
Campbell, DE
机构
[1] CHILDRENS HOSP PHILADELPHIA,DIV ALLERGY IMMUNOL & INFECT DIS,JOSEPH STOKES JR RES INST,PHILADELPHIA,PA 19104
[2] UNIV PENN,SCH MED,DEPT PEDIAT,DIV ALLERGY IMMUNOL & INFECT DIS,PHILADELPHIA,PA 19104
[3] CHILDRENS HOSP PHILADELPHIA,DIV ALLERGY IMMUNOL & INFECT DIS,CLIN IMMUNOL LAB,PHILADELPHIA,PA 19104
关键词
D O I
10.1128/CDLI.3.1.42-46.1996
中图分类号
R392 [医学免疫学]; Q939.91 [免疫学];
学科分类号
100102 ;
摘要
In vitro correlates of type 1 hypersensitivity to natural latex (NL) proteins continue to be limited by both sensitivity and specificity. Methods which have detection limits in the picogram range, namely, radioallergosorbent assays (RAST) and enzyme-linked immunosorbent assays (ELISA), are inadequate for the identification of NL hypersensitivity: in certain-at-risk groups, such as health care workers, A How cytometry assay (FCA), previously shown to be comparable to RAST and ELISA in the identification of NL-sensitized pediatric patients with spina bifida, was compared with electrochemiluminescence (ECL) in the evaluation of pediatric patients with spina bifida and NL-sensitized adult health care workers, As with RAST and ELISA, ECL is capable of detecting picogram amounts of specific analyte, The ECL assay detected NL-specific immunoglobulin E (NL-IgE) in three of six health care workers with strong histories of NL hypersensitivity, All six patients were negative by FCA. Further, 2 of 11 spina bifida patients found to be NL-IgE negative by FCA were NL-IgE positive by ECL, These findings suggest that in sensitivity the ECL assay is an improvement over the FCA for the identification of NL-sensitive individuals.
引用
收藏
页码:42 / 46
页数:5
相关论文
共 15 条
[1]   IGE REACTIVITY TO 14-KD AND 27-KD NATURAL-RUBBER PROTEINS IN LATEX-ALLERGIC CHILDREN WITH SPINA-BIFIDA AND OTHER CONGENITAL-ANOMALIES [J].
ALENIUS, H ;
PALOSUO, T ;
KELLY, K ;
KURUP, V ;
REUNALA, T ;
MAKINENKILJUNEN, S ;
TURJANMAA, K ;
FINK, J .
INTERNATIONAL ARCHIVES OF ALLERGY AND IMMUNOLOGY, 1993, 102 (01) :61-66
[2]  
BLACKBURN GF, 1991, CLIN CHEM, V37, P1534
[3]   THE RUBBER ELONGATION-FACTOR OF RUBBER TREES (HEVEA-BRASILIENSIS) IS THE MAJOR ALLERGEN IN LATEX [J].
CZUPPON, AB ;
CHEN, Z ;
RENNERT, S ;
ENGELKE, T ;
MEYER, HE ;
HEBER, M ;
BAUR, X .
JOURNAL OF ALLERGY AND CLINICAL IMMUNOLOGY, 1993, 92 (05) :690-697
[4]   SKIN AND SEROLOGIC TESTING IN THE DIAGNOSIS OF LATEX ALLERGY [J].
KELLY, KJ ;
KURUP, V ;
ZACHARISEN, M ;
RESNICK, A ;
FINK, JN .
JOURNAL OF ALLERGY AND CLINICAL IMMUNOLOGY, 1993, 91 (06) :1140-1145
[5]   IMMUNOGLOBULIN-E REACTIVITY TO LATEX ANTIGENS IN THE SERA OF PATIENTS FROM FINLAND AND THE UNITED-STATES [J].
KURUP, VP ;
KELLY, KJ ;
TURJANMAA, K ;
ALENIUS, H ;
REUNALA, T ;
PALOSUO, T ;
FINK, JN .
JOURNAL OF ALLERGY AND CLINICAL IMMUNOLOGY, 1993, 91 (06) :1128-1134
[6]   CHARACTERIZATION OF LATEX ANTIGEN AND DEMONSTRATION OF LATEX-SPECIFIC ANTIBODIES BY ENZYME-LINKED-IMMUNOSORBENT-ASSAY IN PATIENTS WITH LATEX HYPERSENSITIVITY [J].
KURUP, VP ;
KELLY, KJ ;
RESNICK, A ;
BANSAL, NK ;
FINK, JN .
ALLERGY PROCEEDINGS, 1992, 13 (06) :329-334
[7]   CHARACTERIZATION OF A MONOCLONAL-ANTIBODY AGAINST LATEX PROTEIN ASSOCIATED WITH LATEX ALLERGY [J].
KURUP, VP ;
KUMAR, A ;
KELLY, KJ ;
FINK, JN .
JOURNAL OF ALLERGY AND CLINICAL IMMUNOLOGY, 1993, 92 (05) :638-643
[8]  
KWITTKEN PL, 1995, PEDIATRICS, V95, P693
[9]   FLOW CYTOMETRIC MEASUREMENT OF IMMUNOGLOBULIN-E TO NATURAL LATEX PROTEINS [J].
KWITTKEN, PL ;
PAWLOWSKI, NA ;
SWEINBERG, SK ;
DOUGLAS, SD ;
CAMPBELL, DE .
CLINICAL AND DIAGNOSTIC LABORATORY IMMUNOLOGY, 1994, 1 (02) :197-201
[10]   SYNTHETIC PEPTIDE HOMOLOGOUS TO THE ENVELOPE PROTEINS OF RETROVIRUSES SHARES A CROSS-REACTING EPITOPE WITH THE CD4 RECEPTOR [J].
ROTHMANN, J ;
HASSAN, NF ;
CAMPBELL, DE ;
KAMANI, N ;
DOUGLAS, SD .
JOURNAL OF CLINICAL MICROBIOLOGY, 1990, 28 (01) :112-115