Improving the detection limit of anion-selective electrodes: An iodide-selective membrane with a nanomolar detection limit

被引:110
作者
Malon, A
Radu, A
Qin, W
Qin, Y
Ceresa, A
Maj-Zurawska, M
Bakker, E [1 ]
Pretsch, E
机构
[1] Auburn Univ, Dept Chem, Auburn, AL 36849 USA
[2] ETH Honggerberg, Organ Chem Lab, CH-8093 Zurich, Switzerland
[3] Univ Warsaw, Dept Chem, PL-02093 Warsaw, Poland
关键词
D O I
10.1021/ac026454r
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The lower detection limit and the selectivity behavior of anion-selective electrodes (ISEs) are improved by using optimized inner solutions and membrane compositions. With a membrane based on the recently described ionophore [9]mercuracarborand-3, a detection limit of 2 x 10(-9) M has been achieved for iodide. Nevertheless, the improvements are less pronounced than in the case of cation ISEs. This is mainly due to the fact that so far no anion ISE is known with the extremely high selectivities of cation ISEs. If the membrane does not contain an ionophore, leaching of the ion exchanger from the membrane into the sample is also a relevant limiting factor except for ion exchangers of very high lipophilicity.
引用
收藏
页码:3865 / 3871
页数:7
相关论文
共 36 条
[1]  
[Anonymous], 1976, PURE APPL CHEM, V48, P127, DOI DOI 10.1351/PAC197648010127
[2]   Mercuracarborand "anti-crown ether"-based chloride sensitive liquid/polymeric membrane electrodes [J].
Badr, IHA ;
Diaz, M ;
Hawthorne, MF ;
Bachas, LG .
ANALYTICAL CHEMISTRY, 1999, 71 (07) :1371-1377
[3]   A selective optical sensor based on [9]mercuracarborand-3, a new type of ionophore with a chloride complexing cavity [J].
Badr, IHA ;
Johnson, RD ;
Diaz, M ;
Hawthorne, MF ;
Bachas, LG .
ANALYTICAL CHEMISTRY, 2000, 72 (18) :4249-4254
[4]   Determination of unbiased selectivity coefficients of neutral carrier-based cation-selective electrodes [J].
Bakker, E .
ANALYTICAL CHEMISTRY, 1997, 69 (06) :1061-1069
[5]   The new wave of ion-selective electrodes. [J].
Bakker, E ;
Pretsch, E .
ANALYTICAL CHEMISTRY, 2002, 74 (15) :420A-426A
[6]  
Bakker E, 1999, ELECTROANAL, V11, P915, DOI 10.1002/(SICI)1521-4109(199909)11:13<915::AID-ELAN915>3.0.CO
[7]  
2-J
[8]   Potentiometry at trace levels [J].
Bakker, E ;
Pretsch, E .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2001, 20 (01) :11-19
[9]   Selectivity of potentiometric ion sensors [J].
Bakker, E ;
Pretsch, E ;
Bühlmann, P .
ANALYTICAL CHEMISTRY, 2000, 72 (06) :1127-1133
[10]   Carrier-based ion-selective electrodes and bulk optodes. 1. General characteristics [J].
Bakker, E ;
Buhlmann, P ;
Pretsch, E .
CHEMICAL REVIEWS, 1997, 97 (08) :3083-3132