Effects of waveform and polarity on defibrillation thresholds in humans using a transvenous lead system

被引:36
作者
Shorofsky, SR
Gold, MR
机构
[1] Division of Cardiology, Department of Medicine, Univ. of Maryland Medical Center, Baltimore, MD
[2] Division of Cardiology, Univ. of Maryland Medical System, Baltimore, MD 21201
关键词
D O I
10.1016/S0002-9149(96)00284-6
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Minimizing defibrillation thresholds is important to allow for implantation of downsized pulse generators with reduced outputs while maintaining an adequate defibrillation safety margin, Recent studies have demonstrated a significant reduction in monophasic defibrillation thresholds with a transvenous lead when the polarity was reversed (proximal coil = cathode), However, conflicting data exist concerning the effect of polarity reversal on biphasic defibrillation thresholds, The present study was designed to evaluate prospectively the effect of waveform shape and polarity on defibrillation thresholds in humans, The group studied consisted of 26 patients undergoing cardioverter-defibrillator implantation for standard indications, All data were obtained with a transvenous lead alone configuration, Defibrillation thresholds were determined using a step-down protocol with the initial waveform and polarity randomized, Reversing polarity significantly decreased the delivered energy at defibrillation threshold with monophasic waveforms (14.8 +/- 7.1 vs 20.4 +/- 8.9 J; p <0.001), but had no effect on the overall efficacy of biphasic waveforms (11.1 +/- 5.5 vs 12.2 +/- 6.5 J), In the subgroup of patients with high biphasic defibrillation thresholds (greater than or equal to 15 J), reversing polarity decreased the defibrillation threshold from 18.2 +/- 5.1 to 13.3 +/- 5.8 J (p <0.001), Similarly, the improvement in defibrillation thresholds with reversing polarity of monophasic waveforms was confined to the subgroup of patients with higher defibrillation thresholds. Therefore, the lack of group effect of polarity on biphasic defibrillation thresholds may be simply due to the overall lowering of defibrillation thresholds by this waveform.
引用
收藏
页码:313 / 316
页数:4
相关论文
共 25 条
[1]  
BABBS CF, 1983, MED INSTRUM, V17, P18
[2]   ELEVATION OF VENTRICULAR DEFIBRILLATION THRESHOLD IN DOGS BY ANTI-ARRHYTHMIC DRUGS [J].
BABBS, CF ;
YIM, GKW ;
WHISTLER, SJ ;
TACKER, WA ;
GEDDES, LA .
AMERICAN HEART JOURNAL, 1979, 98 (03) :345-350
[3]   EVALUATION OF ELECTRODE POLARITY ON DEFIBRILLATION EFFICACY [J].
BARDY, GH ;
IVEY, TD ;
ALLEN, MD ;
JOHNSON, G ;
GREENE, HL .
AMERICAN JOURNAL OF CARDIOLOGY, 1989, 63 (07) :433-437
[4]   A PROSPECTIVE RANDOMIZED EVALUATION OF BIPHASIC VERSUS MONOPHASIC WAVEFORM PULSES ON DEFIBRILLATION EFFICACY IN HUMANS [J].
BARDY, GH ;
IVEY, TD ;
ALLEN, MD ;
JOHNSON, G ;
MEHRA, R ;
GREENE, HL .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1989, 14 (03) :728-733
[5]   ANODAL EXCITATION OF CARDIAC MUSCLE [J].
CRANEFIELD, PF ;
HOFFMAN, BF ;
SIEBENS, AA .
AMERICAN JOURNAL OF PHYSIOLOGY, 1957, 190 (02) :383-390
[6]  
DAVY JM, 1984, CIRCULATION, V70, P406
[7]   THE EFFECTS OF CARDIOVASCULAR DRUGS ON THE DEFIBRILLATION THRESHOLD AND THE PATHOLOGICAL EFFECTS ON THE HEART USING AN AUTOMATIC IMPLANTABLE DEFIBRILLATOR [J].
DEEB, GM ;
HARDESTY, RL ;
GRIFFITH, BP ;
THOMPSON, ME ;
HEILMAN, MS ;
MYEROWITZ, RL .
ANNALS OF THORACIC SURGERY, 1983, 35 (04) :361-366
[8]   INTERNAL CARDIAC DEFIBRILLATION THRESHOLD - EFFECTS OF ACUTE-ISCHEMIA [J].
JONES, DL ;
SOHLA, A ;
KLEIN, GJ .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1986, 9 (03) :322-331
[9]   THE AUTOMATIC IMPLANTABLE CARDIOVERTER DEFIBRILLATOR AS ANTIARRHYTHMIC TREATMENT MODALITY OF CHOICE FOR SURVIVORS OF CARDIAC-ARREST UNRELATED TO ACUTE MYOCARDIAL-INFARCTION [J].
LEHMANN, MH ;
STEINMAN, RT ;
SCHUGER, CD ;
JACKSON, K .
AMERICAN JOURNAL OF CARDIOLOGY, 1988, 62 (10) :803-805
[10]  
MCDANIEL WC, 1987, MED INSTRUM, V21, P170