Evaluation and extraction of the Accufix atrial J lead

被引:12
作者
Daoud, EG [1 ]
Kou, W [1 ]
Davidson, T [1 ]
Niebauer, M [1 ]
Bogun, F [1 ]
Castellani, M [1 ]
Chan, KK [1 ]
Goyal, R [1 ]
Harvey, M [1 ]
Strickberger, SA [1 ]
Man, KC [1 ]
Morady, F [1 ]
机构
[1] VET AFFAIRS MED CTR,ANN ARBOR,MI
关键词
D O I
10.1016/S0002-8703(96)90352-X
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The Accufix (Telectronics Pacing Systems, Englewood, Colo.) atrial active fixation pacemaker lead has a preformed J stiff retention wire that can fracture and erode through the atrium. We screened 85 patients with digital fluoroscopy and detected 18 (21%) lead fractures. Fourteen (16%) leads were successfully extracted by percutaneous techniques. The mean time for extraction and placement of a new atrial lead was 109 +/- 53 min, and a lead >1 year old but not in the presence of a fracture was associated with a significantly longer extraction time. There were two extraction complications. One patient had a fever following the extraction, and one patient had a focal sensory deficit 1 week after the extraction probably caused by a paradoxical embolus. Standard percutaneous extraction tools can be used for successful and safe removal of a fractured Accufix lead.
引用
收藏
页码:266 / 269
页数:4
相关论文
共 9 条
  • [1] ENDOCARDIAL PACING LEADS - THE GOOD, THE BAD, AND THE UGLY
    BRINKER, JA
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (05): : 953 - 954
  • [2] CHRONIC TRANSVENOUS PACEMAKER LEAD REMOVAL USING A UNIQUE, SEQUENTIAL TRANSVENOUS SYSTEM
    BRODELL, GK
    CASTLE, LW
    MALONEY, JD
    WILKOFF, BL
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1990, 66 (12) : 964 - 966
  • [3] INTRAVASCULAR LEAD EXTRACTION USING LOCKING STYLETS AND SHEATHS
    BYRD, CL
    SCHWARTZ, SJ
    HEDIN, NB
    GOODE, LB
    FEARNOT, NE
    SMITH, HJ
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (12): : 1871 - 1875
  • [4] COLLINS KA, 1995, COMMUNICATION 0123
  • [5] COLLINS KA, 1995, COMMUNICATION 0427
  • [6] THE DOTTER RETRIEVER AND PIGTAIL CATHETER - EFFICACY IN EXTRACTION OF CHRONIC TRANSVENOUS PACEMAKER LEADS
    ESPINOSA, RE
    HAYES, DL
    VLIETSTRA, RE
    OSBORN, MJ
    MCGOON, MD
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (12): : 2337 - 2342
  • [7] INTRAVASCULAR LEAD EXTRACTION USING LOCKING STYLETS, SHEATHS, AND OTHER TECHNIQUES
    FEARNOT, NE
    SMITH, HJ
    GOODE, LB
    BYRD, CL
    WILKOFF, BL
    SELLERS, TD
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (12): : 1864 - 1870
  • [8] ATRIAL J PACING LEAD RETENTION WIRE FRACTURE - RADIOGRAPHIC ASSESSMENT, INCIDENCE OF FRACTURE, AND CLINICAL MANAGEMENT
    LLOYD, MA
    HAYES, DL
    HOLMES, DR
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1995, 18 (05): : 958 - 964
  • [9] 5-YEARS EXPERIENCE WITH INTRAVASCULAR LEAD EXTRACTION
    SMITH, HJ
    FEARNOT, NE
    BYRD, CL
    WILKOFF, BL
    LOVE, CJ
    SELLERS, TD
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (11): : 2016 - 2020