Sample introduction systems for reversed phase LC-ICP-MS of selenium using large amounts of methanol - comparison of systems based on membrane desolvation, a spray chamber and direct injection

被引:30
作者
Bendahl, L [1 ]
Gammelgaard, B [1 ]
机构
[1] Danish Univ Pharmaceut Sci, Dept Analyt Chem, DK-2100 Copenhagen, Denmark
关键词
D O I
10.1039/b415717f
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Three sample introduction systems were compared for their efficiency in introducing a variety of biologically relevant selenium species in organic solutions with the purpose of mimicking the effluent from a chromatographic system. The sample introduction systems consisted of a microconcentric nebuliser (MCN) in combination with a membrane desolvator ( MD), a MCN in combination with a cyclonic spray chamber ( CS) and a direct injection nebuliser ( DIN), respectively. Thirteen selenium standards diluted in 50% methanol were introduced in each system at a flow rate of 50 mu l min(-1). Except for the selenite and methaneseleninic acid standards that were lost in the MCN-MD system, similar normalized response factors were achieved for the standards regardless of the sample introduction system. When trimethylselenonium and selenate standards were continuously infused in a linear methanol gradient from 10 to 90%, the signal was almost unaffected when the MCN-MD was used as interface. The signals from the MCN-CS introduction system were almost unaffected in methanol concentrations between 10 and 40%, after which they slowly declined. The signal from the DIN showed a plateau in signal intensity at methanol concentrations between 35 and 70%. Detection limits in flow injection analysis, expressed as the concentration that would result in a signal equivalent to three times the peak to peak noise of the baseline, were estimated to be 200, 150 and 100 ng Se L-1 for the MCN-CS, MCN-MD and DIN, respectively. The independence of the signal from the methanol concentration in the eluent allowed quanti. cation of human selenoprotein P isoforms by gradient microbore reversed phase chromatography (mu RPC) on the basis of an internal trimethylselenonium standard. The MCN-MD system was applied for mu RPC analysis of Se-77 enriched yeast samples and more than 30 selenium containing compounds were separated using a linear gradient from 5% to 95% methanol, some of the selenium species eluting in 85% methanol.
引用
收藏
页码:410 / 416
页数:7
相关论文
共 32 条
[1]   A comparison of nebulizers for microbore LC-ICP-MS with mobile phases containing methanol [J].
Ackley, KL ;
Sutton, KL ;
Caruso, JA .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2000, 15 (09) :1069-1073
[2]   A direct injection high efficiency nebulizer interface for microbore high-performance liquid chromatography-inductively coupled plasma mass spectrometry [J].
Acon, BW ;
McLean, JA ;
Montaser, A .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2001, 16 (08) :852-857
[3]   CHARACTERIZATION AND OPTIMIZATION OF A DIRECT INJECTION NEBULIZER FOR INTRODUCTION OF ORGANIC-SOLVENTS AND VOLATILE ANALYTE SPECIES INTO AN INDUCTIVELY COUPLED PLASMA [J].
AVERY, TW ;
CHAKRABARTY, C ;
THOMPSON, JJ .
APPLIED SPECTROSCOPY, 1990, 44 (10) :1690-1698
[4]   Separation and identification of Se-methylselenogalactosamine -: a new metabolite in basal human urine -: by HPLC-ICP-MS and CE-nano-ESI-(MS)2 [J].
Bendahl, L ;
Gammelgaard, B .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2004, 19 (08) :950-957
[5]   Interfacing capillary electrophoresis with inductively coupled plasma mass spectrometry by direct injection nebulization for selenium speciation [J].
Bendahl, L ;
Gammelgaard, B ;
Jons, O ;
Farver, O ;
Hansen, SH .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2001, 16 (01) :38-42
[6]   Non-spectral interference effects in inductively coupled plasma mass spectrometry using direct injection high efficiency and microconcentric nebulisation [J].
Björn, E ;
Frech, W .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2001, 16 (01) :4-11
[7]   EXCITATION TEMPERATURE AND ELECTRON-DENSITY IN THE INDUCTIVELY COUPLED PLASMA AQUEOUS VS ORGANIC-SOLVENT INTRODUCTION [J].
BLADES, MW ;
CAUGHLIN, BL .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1985, 40 (04) :579-591
[8]   EFFECTS OF ORGANIC-SOLVENTS IN INDUCTIVELY COUPLED PLASMA ATOMIC EMISSION-SPECTROMETRY [J].
BOORN, AW ;
BROWNER, RF .
ANALYTICAL CHEMISTRY, 1982, 54 (08) :1402-1410
[9]   2-Dimensional gel electrophoresis technique for yeast selenium-containing proteins -: sample preparation and MS approaches for processing 2-D gel protein spots [J].
Chassaigne, H ;
Chéry, CC ;
Bordin, G ;
Vanhaecke, F ;
Rodriguez, AR .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2004, 19 (01) :85-95
[10]   Detection and quantification of selenium in proteins by means of gel electrophoresis and electrothermal vaporization ICP-MS [J].
Chéry, CC ;
Chassaigne, H ;
Verbeeck, L ;
Cornelis, R ;
Vanhaecke, F ;
Moens, L .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2002, 17 (06) :576-580