A positive displacement micropump for microdialysis

被引:14
作者
Cunneen, J
Lin, YC
Caraffini, S
Boyd, JG
Hesketh, PJ [1 ]
Lunte, SM
Wilson, GS
机构
[1] Univ Illinois, Dept Elect Engn & Comp Sci, Microfabricat Applicat Lab, Chicago, IL 60680 USA
[2] Univ Illinois, Dept Mech Engn, Chicago, IL 60607 USA
[3] Natl Cheng Kung Univ, Dept Engn Sci, Tainan, Taiwan
[4] Univ Kansas, Dept Chem & Pharmaceut Chem, Lawrence, KS 66047 USA
关键词
D O I
10.1016/S0957-4158(98)00008-7
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
Miniature fluid pumps, measuring 15 x 4 x 1 mm, have been microfabricated with silicon, glass, and polyimide. Pumps have been tested with deionized water and 10% glycerol solutions as the working solutions. The pumps have been operated with a pneumatic drive or a piezoelectric drive. Flow rates from 0.1 to 110 mu l min(-1) have been achieved. The maximum pressure generated by the pumps was 56 cm of water. Pumps have been operated with a dialysis probe and with a microchannel load. The pump lifetime is limited by the degradation in the performance of the polyimide components in the pump. The power consumption was less than 1 mW at a drive frequency of 10 Hz. (C) 1998 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:561 / 583
页数:23
相关论文
共 37 条
[11]  
HESKETH PJ, 1997, P 21 INT C MICR MIEL, P63
[12]  
HESKETH PJ, 1997, P ECS
[13]   ONLINE COUPLING OF IN-VIVO MICRODIALYSIS SAMPLING WITH CAPILLARY ELECTROPHORESIS [J].
HOGAN, BL ;
LUNTE, SM ;
STOBAUGH, JF ;
LUNTE, CE .
ANALYTICAL CHEMISTRY, 1994, 66 (05) :596-602
[14]  
LIN YC, 1996, THESIS U ILLINOIS CH
[15]  
LIN YC, 1996, SENS ACT WORKSH HILT
[16]  
LIN YC, 1995, MICROSTRUCTURES MICR, V2, P67
[17]  
MANZ A, 1995, MESA MONOGRAPH, P2
[18]  
OLSSON A, 1996, P IEEE 10 ANN INT WO, P305
[19]   SILICON AS A MECHANICAL MATERIAL [J].
PETERSEN, KE .
PROCEEDINGS OF THE IEEE, 1982, 70 (05) :420-457
[20]   MICROFABRICATED CHEMICAL MEASUREMENT SYSTEMS [J].
RAMSEY, JM ;
JACOBSON, SC ;
KNAPP, MR .
NATURE MEDICINE, 1995, 1 (10) :1093-1096