We report on the performance of a measurement system for the recognition of individual analytes and their binary mixtures which is based on a multiarray of four micromachined silicon cantivelers actuated at their resonance frequency. The cantilevers have been functionalized by organic polymers [polydimethylsiloxane (PDMS) and polyvinylpyridine (PVP)] and amorphous nitrogen-rich carbon nitride films. We found that the sensitivity and selectivity of the cantilevers coated with CNx films change according to the layer thickness. Our results show that the selected combination of sensitive layers ensures a wide range of specific, reversible and reproducible sensor responses upon exposure to methanol, 2-propanol, water and their binary mixtures. Further, it was found that the differences in recovery times of PDMS and CNx films after exposure to the two alcohols and their mixtures could be used especially for low analyte concentrations as a second characteristic in addition to the resonance frequency shift for the identification of individual components in the mixtures. (C) 2003 American Vacuum Society.
引用
收藏
页码:2931 / 2936
页数:6
相关论文
共 15 条
[1]   A chemical sensor based on a microfabricated cantilever array with simultaneous resonance-frequency and bending readout [J].
Battiston, FM ;
Ramseyer, JP ;
Lang, HP ;
Baller, MK ;
Gerber, C ;
Gimzewski, JK ;
Meyer, E ;
Güntherodt, HJ .
SENSORS AND ACTUATORS B-CHEMICAL, 2001, 77 (1-2) :122-131
[2]   A chemical sensor array based system for protecting wastewater treatment plants [J].
Bourgeois, W ;
Gardey, G ;
Servieres, M ;
Stuetz, RM .
SENSORS AND ACTUATORS B-CHEMICAL, 2003, 91 (1-3) :109-116
[3]   Classification performance of carbon black-polymer composite vapor detector arrays as a function of array size and detector composition [J].
Burl, MC ;
Sisk, BC ;
Vaid, TP ;
Lewis, NS .
SENSORS AND ACTUATORS B-CHEMICAL, 2002, 87 (01) :130-149
[4]   Surface acoustic wave humidity sensors: a comparison between different types of sensitive membrane [J].
Caliendo, C ;
Verona, E ;
Anisimkin, VI .
SMART MATERIALS & STRUCTURES, 1997, 6 (06) :707-715
[5]  
DAECHE F, 2001, ACTUATORS B, V80, P169
[6]   Low-power gas sensors based on work-function measurement in low-cost hybrid flip-chip technology [J].
Fleischer, M ;
Ostrick, B ;
Pohle, R ;
Simon, E ;
Meixner, H ;
Bilger, C ;
Daeche, F .
SENSORS AND ACTUATORS B-CHEMICAL, 2001, 80 (03) :169-173
[7]   A semiconducting metal-oxide array for monitoring fish freshness [J].
Hammond, J ;
Marquis, B ;
Michaels, R ;
Oickle, B ;
Segee, B ;
Vetelino, J ;
Bushway, A ;
Camire, ME ;
Davis-Dentici, K .
SENSORS AND ACTUATORS B-CHEMICAL, 2002, 84 (2-3) :113-122
[8]  
KHARAZ A, 1995, SENSOR ACTUAT A-PHYS, V47, P1
[9]   Atomic force microscopy and lateral force microscopy using piezoresistive cantilevers [J].
Linnemann, R ;
Gotszalk, T ;
Rangelow, IW ;
Dumania, P ;
Oesterschulze, E .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY B, 1996, 14 (02) :856-860
[10]   A semiconducting metal oxide sensor array for the detection of NOx and NH3 [J].
Marquis, BT ;
Vetelino, JF .
SENSORS AND ACTUATORS B-CHEMICAL, 2001, 77 (1-2) :100-110