摘要
Background - Continuous-flow rotary pumps with axial design are increasingly used for left ventricular assist support. The efficacy of this design compared with pulsatile, volume displacement pumps, with respect to characteristics of left ventricular unloading, and exercise performance remains largely unstudied. Methods and Results - Thirty-four patients undergoing implantation with a pulsatile, volume displacement pump operating in a full-to-empty cycle (HeartMate XVE; Thoratec Inc, Pleasanton, Calif; n = 16) or continuous-flow rotary pump with an axial design operating at a fixed rotor speed (HeartMate II; Thoratec Inc; n = 18) were evaluated with right heart catheterization and echocardiography preoperatively and at 3 months postoperatively and cardiopulmonary exercise testing 3 months postoperatively. Support with either the XVE or II resulted in significant (P < 0.05) increases in cardiac output and reduction in mean pulmonary artery and pulmonary wedge pressures. Exercise capacity at 3 months was similar between groups (% predicted peak VO2 - XVE: 46.8 +/- 10.2 versus II: 49.1 +/- 13.6). Echocardiography at 3 months demonstrated a significantly (P < 0.05) greater reduction in left ventricular end-diastolic volume (-49 +/- 16% versus -35 +/- 20%), left ventricular end-systolic volume (-59 +/- 20 versus -37 +/- 21%), and percent mitral valve regurgitant volume (-99 +/- 2% versus -52 +/- 56%) for the XVE compared with II, respectively. Conclusions - The HeartMate XVE or II provided equivalent degrees of hemodynamic support and exercise capacity. The XVE was associated with greater left ventricular volume unloading. Characteristics of left ventricular pressure and volume unloading between these pump designs and mode of operation do not influence early exercise performance.
引用
收藏
页码:I8 / I15
页数:8
相关论文
共 28 条
[1]   Rotary blood pump flow spontaneously increases during exercise under constant pump speed: Results of a chronic study [J].
Akimoto, T ;
Yamazaki, K ;
Litwak, P ;
Litwak, KN ;
Tagusari, O ;
Mori, T ;
Antaki, JF ;
Kameneva, MV ;
Watach, MJ ;
Umezu, M ;
Tomioka, J ;
Kormos, RL ;
Koyanagi, H ;
Griffith, BP .
ARTIFICIAL ORGANS, 1999, 23 (08) :797-801
[2]  
Bjornstad H, 2001, EUR HEART J, V22, P37
[3]   Development of the Nimbus/Pittsburgh axial flow left ventricular assist system [J].
Butler, K ;
Thomas, D ;
Antaki, J ;
Borovetz, H ;
Griffith, B ;
Kameneva, M ;
Kormos, R ;
Litwak, P .
ARTIFICIAL ORGANS, 1997, 21 (07) :602-610
[4]   Development of the Nimbus/University of Pittsburgh innovative ventricular assist system [J].
Butler, KC ;
Dow, JJ ;
Litwak, P ;
Kormos, RL ;
Borovetz, HS .
ANNALS OF THORACIC SURGERY, 1999, 68 (02) :790-794
[5]   Exercise performance in patients with end-stage heart failure after implantation of a left ventricular assist device and after heart transplantation -: An outlook for permanent assisting? [J].
de Jonge, N ;
Kirkels, H ;
Lahpor, JR ;
Klöpping, C ;
Hulzebos, EJ ;
de la Rivière, AB ;
de Medina, EOR .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 37 (07) :1794-1799
[6]   A miniature implantable axial flow ventricular assist device [J].
DeBakey, ME .
ANNALS OF THORACIC SURGERY, 1999, 68 (02) :637-640
[7]   Implantable left ventricular assist devices provide an excellent outpatient bridge to transplantation and recovery [J].
DeRose, JJ ;
Umana, JP ;
Argenziano, M ;
Catanese, KA ;
Gardocki, MT ;
Flannery, M ;
Levin, HR ;
Sun, BC ;
Rose, EA ;
Oz, MC .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1997, 30 (07) :1773-1777
[8]   QUANTITATIVE DOPPLER ASSESSMENT OF VALVULAR REGURGITATION [J].
ENRIQUEZSARANO, M ;
BAILEY, KR ;
SEWARD, JB ;
TAJIK, AJ ;
KROHN, MJ ;
MAYS, JM .
CIRCULATION, 1993, 87 (03) :841-848
[9]   Long-term follow-up of thoratec ventricular assist device bridge-to-recovery patients successfully removed from support after recovery of ventricular function [J].
Farrar, DJ ;
Holman, WR ;
McBride, LR ;
Kormos, RL ;
Icenogle, TB ;
Hendry, PJ ;
Moore, CH ;
Loisance, DY ;
El-Banayosy, A ;
Frazier, H .
JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2002, 21 (05) :516-521
[10]   Multicenter clinical evaluation of the HeartMate vented electric left ventricular assist system in patients awaiting heart transplantation [J].
Frazier, OH ;
Rose, EA ;
Oz, MC ;
Dembitsky, W ;
McCarthy, P ;
Radovancevic, B ;
Poirier, VL ;
Dasse, KA .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2001, 122 (06) :1186-1195