Comparison between YAP:Ce and CsI(Tl) multipillar matrices

被引:8
作者
Vittori, F
Malatesta, T
de Notaristefani, F
机构
[1] Univ Rome La Sapienza, Ist Nazl Fis Nucl, Sect Rome 1, Dept Phys, I-00100 Rome, Italy
[2] Ist Nazl Fis Nucl, Sect Rome 1, Regina Elena Inst Phys, Med Phys Lab, Rome, Italy
[3] E Amaldi Univ Rome 3, Ist Nazl Fis Nucl, Sect Rome 1, Dept Phys, Rome, Italy
关键词
YAP : Ce; CsI(T1); multipillar structure; position-sensitive photomultipliers; segmented crystal scintillators; gamma-ray imager;
D O I
10.1016/S0168-9002(98)00763-3
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The characteristics of segmented CsI(TI) and YAP:Ce scintillators coupled to a position-sensitive photomultiplier tube (PSPMT Hamamatsu R2486) have been measured and compared to evaluate the potential use of such scintillators as precise X- and gamma-ray imagers. For such configurations, we obtained good spatial resolution and linearity over their entire 40 x 40 mm(2) Fields of View (FOVs). The scintillating matrices consist of orthogonal arrays of crystal pillars with 0.6 x 0.6 mm(2) cross sections and thicknesses of 3 and 10 mm. The 3 mm thick. CsI(TI) matrix shows a light yield of 44% compared to NaI(TI), an energy resolution of 22% FWHM at 122 keV, a single pillar spatial response of 0.5 mm FWHM, and a spatial resolution of 0.83 mm. (C) 1998 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:497 / 506
页数:10
相关论文
共 15 条
[1]   SCINTILLATION PROPERTIES OF YAP-CE [J].
BACCARO, S ;
BLAZEK, K ;
DENOTARISTEFANI, F ;
MALY, P ;
MARES, JA ;
PANI, R ;
PELLEGRINI, R ;
SOLURI, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1995, 361 (1-2) :209-215
[2]   YAP MULTICRYSTAL GAMMA-CAMERA PROTOTYPE [J].
BLAZEK, K ;
DENOTARISTEFANI, F ;
MALY, P ;
PANI, R ;
PELLEGRINI, R ;
PERGOLA, A ;
SCOPINARO, F ;
SOLURI, A .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (05) :1474-1482
[3]   First results from a YAP:Ce gamma camera for small animal studies [J].
deNotaristefani, F ;
Pani, R ;
Scopinaro, F ;
Barone, LM ;
Blazek, K ;
DeVincentis, G ;
Malatesta, T ;
Maly, P ;
Pellegrini, R ;
Pergola, A ;
Soluri, A ;
Vittori, F .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (06) :3264-3271
[4]  
DENOTARISTEFANI F, 1996, SCINT95, P559
[5]  
DOREMBOS P, 1994, 9403 DELFT U TECHN R
[6]   Thin scintillators and position sensitive photomultiplier tubes for hard X-ray imaging in space [J].
Kroeger, RA ;
Grove, JE ;
Inderhees, SE ;
Johnson, WN ;
Kinzer, RL ;
Kurfess, JD ;
Phlips, BP .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) :881-884
[7]  
MALATESTA T, 1998, IN PRESS NUCL PHYS B
[8]  
MALY P, COMMUNICATION
[9]  
MOSES WW, 1996, SCINT 95, P9
[10]   An ISPA-camera for gamma rays [J].
Puertolas, D ;
Piedigrossi, D ;
Pani, R ;
Leutz, H ;
Gys, T ;
deNotaristefani, F ;
DAmbrosio, C .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (06) :2221-2228