Sensitive determination of Cd and Pb by differential pulse stripping voltammetry with in situ bismuth-modified zeolite doped carbon paste electrodes

被引:111
作者
Cao, Linyuan [2 ]
Jia, Jianbo [1 ]
Wang, Zhenhui [2 ]
机构
[1] Chinese Acad Sci, Changchun Inst Appl Chem, State Key Lab Electroanalyt Chem, Changchun 130022, Peoples R China
[2] Henan Normal Univ, Coll Chem & Environm Sci, Xinxiang 453007, Peoples R China
基金
中国国家自然科学基金;
关键词
bismuth film electrode; zeolite doped carbon paste electrode; differential pulse stripping voltammetry; cadmium; lead;
D O I
10.1016/j.electacta.2007.09.024
中图分类号
O646 [电化学、电解、磁化学];
学科分类号
081704 ;
摘要
Here we investigated the analytical performances of the bismuth-modified zeolite doped carbon paste electrode (BiF-ZDCPE) for trace Cd and Pb analysis. The characteristics of bismuth-modified electrodes were improved greatly via addition of synthetic zeolite into carbon paste. To obtain high reproducibility and sensitivity, optimum experimental conditions for bismuth deposition Were Studied. Voltammetric responses of the BiF-ZDCPEs prepared with different ratios of zeolite, carbon powder, and silicone, were examined under same conditions. The in Situ plated (zeolite/graphite powder/silicone, 10/190/80 w/w) BiF-ZDCPEs exhibited the most sensitive response to Cd and Pb in 0.10 M acetate buffer(pH 4.5). The detection limits of the modified electrode were 0.08 mu g L-1 for Cd(II) and 0.10 mu gL(-1) for Pb(II) based on three times the standard deviation of the baseline with a preconcentration time of 120s under optimal conditions, respectively. The modified electrode showed well linear response to both Cd(II) and Pb(II) over the concentration range from 1.0 to 20.0 mu g L-1. The BiF-ZDCPEs were successfully applied to the determination of Cd(11) and Pb(11) in real samples, and the results were in agreement with those of atomic absorption spectroscopy (AAs) (C) 2007 Elsevier Ltd. All rights reserved.
引用
收藏
页码:2177 / 2182
页数:6
相关论文
共 30 条
[1]   Development of a novel carbon paste electrode containing a natural zeolite for the voltammetric determination of copper [J].
Alpat, SK ;
Yuksel, U ;
Akcay, H .
ELECTROCHEMISTRY COMMUNICATIONS, 2005, 7 (02) :130-134
[2]   Bismuth film microelectrodes for heavy metals monitoring by anodic stripping voltammetry [J].
Baldo, MA ;
Daniele, S ;
Bragato, C .
JOURNAL DE PHYSIQUE IV, 2003, 107 :103-106
[3]   Effect of Bi(III) concentration on the stripping voltammetric response of in situ bismuth-coated carbon paste and gold electrodes [J].
Baldrianova, L. ;
Svancara, I. ;
Vlcek, M. ;
Economou, A. ;
Sotiropoulos, S. .
ELECTROCHIMICA ACTA, 2006, 52 (02) :481-490
[4]   Anodic stripping voltammetry at in situ bismuth-plated carbon and gold microdisc electrodes in variable electrolyte content unstirred solutions [J].
Baldrianova, L. ;
Svancara, I. ;
Economou, A. ;
Sotiropoulos, S. .
ANALYTICA CHIMICA ACTA, 2006, 580 (01) :24-31
[5]   Accumulation and voltammetric determination of complexed metal ions at zeolite-modified sensor electrodes [J].
Bing, C ;
Kryger, L .
TALANTA, 1996, 43 (01) :153-160
[6]   Mercury film electrodes: developments, trends and potentialities for electroanalysis [J].
Economou, A ;
Fielden, PR .
ANALYST, 2003, 128 (03) :205-212
[7]   Bismuth-film electrodes: recent developments and potentialities for electroanalysis [J].
Economou, A .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 2005, 24 (04) :334-340
[8]   Quantitative determination of zinc in milkvetch by anodic stripping voltammetry with bismuth film electrodes [J].
Guo, ZY ;
Feng, F ;
Hou, YX ;
Jaffrezic-Renault, N .
TALANTA, 2005, 65 (04) :1052-1055
[9]   Novel electrode for electrochemical stripping analysis based on carbon paste modified with bismuth powder [J].
Hocevar, SB ;
Svancara, I ;
Vytras, K ;
Ogorevc, B .
ELECTROCHIMICA ACTA, 2005, 51 (04) :706-710
[10]   A study on operational parameters for advanced use of bismuth film electrode in anodic stripping voltammetry [J].
Hocevar, SB ;
Ogorevc, B ;
Wang, J ;
Pihlar, B .
ELECTROANALYSIS, 2002, 14 (24) :1707-1712