Aerosol delivery to young children by pMDI-spacer: Is facemask design important?

被引:17
作者
Esposito-Festen, J
Ates, B
van Vliet, F
Hop, W
Tiddens, H
机构
[1] Erasmus Univ, Med Ctr, Sophia Childrens Hosp, Dept Pediat,Div Resp Med, NL-3000 CB Rotterdam, Netherlands
[2] Erasmus Univ, Med Ctr, Dept Hosp Pharm, Rotterdam, Netherlands
[3] Erasmus Univ, Med Ctr, Dept Biostat, Rotterdam, Netherlands
关键词
spacer; facemask; young children; wheeze; aerosol;
D O I
10.1111/j.1399-3038.2005.00285.x
中图分类号
R392 [医学免疫学];
学科分类号
100102 ;
摘要
This study aimed at identifying in a daily-life setting the influence of facemask design on drug delivery via a spacer to young children. In a 4-week randomized crossover study, 24 children (7-23-months old) with recurrent wheeze tested the AstraZeneca(R), Galemed(R), and Hans Rudolph(R) facemask combined with the NebuChamber(R) at home. Each mask was tested twice daily for seven consecutive days. Filters positioned between the NebuChamber and facemask trapped the budesonide aerosol (200 mu g, Pulmicort(R)). Parents were asked to score the child's degree of cooperation during administration on diary cards. The administration procedure was evaluated through video recordings. Mean filter dose (standard deviation (s.d.)), expressed as % of nominal dose, was 39% (14), 47% (12), and 42% (11) for the AstraZeneca, the Galemed and the Hans Rudolph mask, respectively. Irrespective of the degree of cooperation, the Galemed mask gave significantly higher mean filter doses than the other masks (level of significance) (p < 0.045). Median (range) within-subject dose variability, expressed, as coefficient of variation (CV), was 37% (19-255), 32% (9-114), and 30% (9-115) for the AstraZeneca mask, the Galemed mask and the Hans Rudolph mask, respectively, not significant. Dose variability increased with decreasing cooperation for all three masks (p = 0.007). Drug delivery to young children with recurrent wheeze by means of the NebuChamber can be enhanced using the Galemed facemask. Dose variability seems to be independent of facemask design but mainly depends on cooperation.
引用
收藏
页码:348 / 353
页数:6
相关论文
共 21 条
[1]   INFLUENCE OF SPACER DEVICE ON DRUG-DELIVERY TO YOUNG-CHILDREN WITH ASTHMA [J].
AGERTOFT, L ;
PEDERSEN, S .
ARCHIVES OF DISEASE IN CHILDHOOD, 1994, 71 (03) :217-219
[2]   Comparison of efficiency and preference of metal and plastic spacers in preschool children [J].
Amirav, I ;
Tiosano, T ;
Chamny, S ;
Chirurg, S ;
Oren, S ;
Grossman, Z ;
Kahan, E ;
Newhouse, MT ;
Mansour, Y .
ANNALS OF ALLERGY ASTHMA & IMMUNOLOGY, 2004, 93 (03) :249-252
[3]   Redesigned face mask improves "Real life" aerosol delivery for nebuchamber [J].
Amirav, I ;
Mansour, Y ;
Mandelberg, A ;
Bar-Ilan, I ;
Newhouse, MT .
PEDIATRIC PULMONOLOGY, 2004, 37 (02) :172-177
[4]   Aerosol therapy with valved holding chambers in young children: Importance of the facemask seal [J].
Amirav, I ;
Newhouse, MT .
PEDIATRICS, 2001, 108 (02) :389-394
[5]   THE EFFECT OF DELAY, MULTIPLE ACTUATIONS AND SPACER STATIC CHARGE ON THE IN-VITRO DELIVERY OF BUDESONIDE FROM THE NEBUHALER [J].
BARRY, PW ;
OCALLAGHAN, C .
BRITISH JOURNAL OF CLINICAL PHARMACOLOGY, 1995, 40 (01) :76-78
[6]   INHALED BUDESONIDE FOR TREATMENT OF RECURRENT WHEEZING IN EARLY-CHILDHOOD [J].
BISGAARD, H ;
MUNCK, SL ;
NIELSEN, JP ;
PETERSEN, W ;
OHLSSON, SV .
LANCET, 1990, 336 (8716) :649-651
[7]  
BISGAARD H, 1995, EUR RESPIR J, V8, P856
[8]   A NONELECTROSTATIC SPACER FOR AEROSOL DELIVERY [J].
BISGAARD, H ;
ANHOJ, J ;
KLUG, B ;
BERG, E .
ARCHIVES OF DISEASE IN CHILDHOOD, 1995, 73 (03) :226-230
[9]  
Bisgaard H, 1994, EUR RESPIR REV, V4, P15
[10]   Development of a scale to measure children's metered-dose inhaler and spacer technique [J].
Boccuti, L ;
Celano, M ;
Geller, RJ ;
Phillips, KM .
ANNALS OF ALLERGY ASTHMA & IMMUNOLOGY, 1996, 77 (03) :217-221