Assessing Blood Pressure Accuracy of an Aneroid Sphygmomanometer in a National Survey Environment

被引:16
作者
Ostchega, Yechiam [1 ]
Prineas, Ronald J. [2 ]
Nwankwo, Tatiana [1 ]
Zipf, George [1 ]
机构
[1] Ctr Dis Control & Prevent, Div Hlth & Nutr Examinat Stat, Natl Ctr Hlth Stat, US Dept HHS, Hyattsville, MD 20782 USA
[2] Wake Forest Univ, Bowman Gray Sch Med, Dept Epidemiol & Prevent, Winston Salem, NC USA
关键词
aneroid; blood pressure; hypertension; mercury sphygmomanometer; NHANES; US DEMOGRAPHIC-TRENDS; MIDARM CIRCUMFERENCE; VALIDATION; SOCIETY; DEVICES; HEALTH; CUFFS;
D O I
10.1038/ajh.2010.232
中图分类号
R6 [外科学];
学科分类号
1002 ; 100210 ;
摘要
BACKGROUND The "gold standard" employed for obtaining blood pressure (BP) for all the National Health and Nutrition Examination Surveys (NHANES) has been the mercury sphygmomanometer (HgS). Because of environmental concerns, there is a need to explore an alternative to HgS. METHODS We compared the accuracy of the Welch Allyn 767 wall aneroid sphygmomanometer (AnS) to the HgS in children and adults and by BP cuff sizes. Each participant had three BP measurements per device recorded sequentially. The order of the devices and observer were random. A total of 727 individuals participating in the NHANES participated in the study. RESULTS The mean AnS readings were not statistically significantly different from those of the HgS with the exception of systolic BP (SBP) in aged 8-17 years (mean difference 1.10, s.d. 4.87). There were no statistically significantly different by BP cuff sizes. Agreement for the prevalence of hypertension (BP >= 140 systolic or diastolic >= 90 mm Hg) was above chance (K = 0.81; sensitivity = 81%; specificity = 98%) with AnS readings underestimating by 1.66% (18.33 vs. 20%, P > 0.05) compared to the HgS reading. CONCLUSIONS With the exception of SBP in ages 8-17 years, the AnS device readings were not significantly different from HgS readings by age or BP cuff sizes selection. Agreement for hypertension classification is good. An accurate and well-calibrated AnS could therefore provide an acceptable alternative to the use of a HgS in surveys, although with appropriate caution given the 81% sensitivity with regard to hypertension thresholds that was observed.
引用
收藏
页码:322 / 327
页数:6
相关论文
共 20 条
  • [1] *ASS ADV MED INSTR, 1992, AM NAT STAND EL AUT
  • [2] Bland JM, 1999, STAT METHODS MED RES, V8, P135, DOI 10.1177/096228029900800204
  • [3] Are aneroid sphygmomanometers accurate in hospital and clinic settings?
    Canzanello, VJ
    Jensen, PL
    Schwartz, GL
    [J]. ARCHIVES OF INTERNAL MEDICINE, 2001, 161 (05) : 729 - 731
  • [4] Centers for Disease Control and Prevention, About the National Health and Nutrition Examination Survey
  • [5] Seventh Report of the Joint National Committee on Prevention, Detection, Evaluation, and Treatment of High Blood Pressure
    Chobanian, AV
    Bakris, GL
    Black, HR
    Cushman, WC
    Green, LA
    Izzo, JL
    Jones, DW
    Materson, BJ
    Oparil, S
    Wright, JT
    Roccella, EJ
    [J]. HYPERTENSION, 2003, 42 (06) : 1206 - 1252
  • [6] Validation of Heine Gamma G7 (G5) and XXL-LF aneroid devices for blood pressure measurement
    Dorigatti, Francesca
    Bonso, Elisa
    Zanier, Ada
    Palatini, Paolo
    [J]. BLOOD PRESSURE MONITORING, 2007, 12 (01) : 29 - 33
  • [7] Evaluating the Accuracy of an Aneroid Sphygmomanometer in a Clinical Trial Setting
    Ma, Yong
    Temprosa, Marinella
    Fowler, Sarah
    Prineas, Ronald J.
    Montez, Maria G.
    Brown-Friday, Janet
    Carrion-Petersen, Mary L.
    Whittington, Tracy
    [J]. AMERICAN JOURNAL OF HYPERTENSION, 2009, 22 (03) : 263 - 266
  • [8] The mercury sphygmomanometer should be abandoned before it is proscribed
    Markandu, ND
    Whitcher, F
    Arnold, A
    Carney, C
    [J]. JOURNAL OF HUMAN HYPERTENSION, 2000, 14 (01) : 31 - 36
  • [9] Applying environmental product design to biomedical products research
    Messelbeck, J
    Sutherland, L
    [J]. ENVIRONMENTAL HEALTH PERSPECTIVES, 2000, 108 : 997 - 1002
  • [10] How accurate are sphygmomanometers?
    Mion, D
    Pierin, AMG
    [J]. JOURNAL OF HUMAN HYPERTENSION, 1998, 12 (04) : 245 - 248