A comprehensive physical image quality evaluation of a selenium based digital x-ray imaging system for thorax radiography

被引:32
作者
Launders, JH [1 ]
Kengyelics, SM [1 ]
Cowen, AR [1 ]
机构
[1] Gen Infirm, Dept Med Phys, FAXiL, Leeds LS1 3EX, W Yorkshire, England
关键词
digital radiography; amorphous selenium; image quality; chest radiography;
D O I
10.1118/1.598276
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A selenium based digital x-ray system dedicated to chest radiography has been installed by the UK Department of Health's Medical Devices Agency at Leeds General Infirmary, UK, to undergo a comprehensive evaluation, including the physical image quality. The underlying characteristics which define the overall image quality of a system are the following: sensitometric response, modulation transfer function, and noise power spectrum. These have been measured objectively on preprocessed digital data acquired under relevant radiographic conditions. The image data is further processed prior to hard copy display. The displayed image quality may only be measured subjectively; threshold contrast detail detectability is such a measure which can be related to the objective measures of image quality. The objective imaging characteristics suggest that Thoravision has a significant advantage over conventional radiography imaging systems. However, subjective measures have demonstrated that the image processing can have a significant effect on the perceived image quality. Thoravision has the potential to deliver a significantly improved image quality to clinicians with no increase in radiation exposure to the patient, or image quality may be maintained with a reduction in radiation exposure. Digital image processing is central to the efficiency with which it achieves this. (C) 1998 American Association of Physicists in Medicine. [S0094-2405(98)02706-0].
引用
收藏
页码:986 / 997
页数:12
相关论文
共 28 条
[1]  
[Anonymous], 1986, 41 ICRU
[2]  
[Anonymous], 1979, PHYS MED IMAGING REC
[3]  
Bendat JS., 2011, RANDOM DATA ANAL MEA
[4]  
BIRCH R, 1979, CATALOGUE SPECTRAL D
[5]   XERORADIOGRAPHY [J].
BOAG, JW .
PHYSICS IN MEDICINE AND BIOLOGY, 1973, 18 (01) :3-37
[6]  
COWEN AR, 1995, IMAGING, V7, P204
[7]   IMAGE-INFORMATION TRANSFER PROPERTIES OF X-RAY FLUORESCENT SCREENS [J].
DICK, CE ;
MOTZ, JW .
MEDICAL PHYSICS, 1981, 8 (03) :337-346
[8]   X-RAY-IMAGING WITH AMORPHOUS SELENIUM - DETECTIVE QUANTUM EFFICIENCY OF PHOTOCONDUCTIVE RECEPTORS FOR DIGITAL MAMMOGRAPHY [J].
FAHRIG, R ;
ROWLANDS, JA ;
YAFFE, MJ .
MEDICAL PHYSICS, 1995, 22 (02) :153-160
[9]   A SIMPLE METHOD FOR DETERMINING THE MODULATION TRANSFER-FUNCTION IN DIGITAL RADIOGRAPHY [J].
FUJITA, H ;
TSAI, DY ;
ITOH, T ;
DOI, K ;
MORISHITA, J ;
UEDA, K ;
OHTSUKA, A .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1992, 11 (01) :34-39
[10]   INVESTIGATION OF BASIC IMAGING PROPERTIES IN DIGITAL RADIOGRAPHY .1. MODULATION TRANSFER-FUNCTION [J].
GIGER, ML ;
DOI, K .
MEDICAL PHYSICS, 1984, 11 (03) :287-295