A cryogenic radiometer for absolute neutron rate measurement

被引:11
作者
Chowdhuri, Z [1 ]
Hansen, GL
Jane, V
Keith, CD
Lozowski, WM
Snow, WM
Dewey, MS
Gilliam, DM
Greene, GL
Nico, JS
Thompson, AK
Wietfeldtd, FE
机构
[1] NIST, Ctr Neutron Res, Gaithersburg, MD 20899 USA
[2] Indiana Univ, Cyclotron Facil, Bloomington, IN 47408 USA
关键词
D O I
10.1063/1.1605492
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We detail the design and operation of an electrical substitution radiometer, which measures the absolute fluence of thermal neutron beams based on neutron absorption in a Li-6-rich metallic alloy. We demonstrate that the device possesses a linear response to electrical power in the 50-500 nW range with better than 0.1% absolute accuracy and negligible zero offset. We show the response of the radiometer to a 3 meV neutron beam. We compare the noise of the radiometer to the noise observed upstream in a thin rate monitor and show that the radiometer noise is understood quantitatively. We establish that the radiometer measures neutron rates above 10(5) s(-1) to an absolute accuracy of 0.1%. (C) 2003 American Institute of Physics.
引用
收藏
页码:4280 / 4293
页数:14
相关论文
共 58 条
[41]   A RADIOMETRIC DETERMINATION OF THE STEFAN-BOLTZMANN CONSTANT AND THERMODYNAMIC TEMPERATURES BETWEEN -40-DEGREES-C AND + 100-DEGREES-C [J].
QUINN, TJ ;
MARTIN, JE .
PHILOSOPHICAL TRANSACTIONS OF THE ROYAL SOCIETY A-MATHEMATICAL PHYSICAL AND ENGINEERING SCIENCES, 1985, 316 (1536) :85-+
[42]   Synchrotron-radiation-operated cryogenic electrical-substitution radiometer as the high-accuracy primary detector standard in the ultraviolet, vacuum-ultraviolet, and soft-x-ray spectral ranges [J].
Rabus, H ;
Persch, V ;
Ulm, G .
APPLIED OPTICS, 1997, 36 (22) :5421-5440
[43]   High precision electrical substitution radiometer based on superconducting-resistive transition edge thermometry [J].
Reintsema, CD ;
Koch, JA ;
Grossman, EN .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1998, 69 (01) :152-163
[44]   A CALORIMETER FOR NEUTRON-FLUX MEASUREMENT [J].
RICHARDSON, JM ;
CHUPP, TE ;
ROBERTSON, RGH ;
WILKERSON, JF .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 306 (1-2) :291-299
[45]   Accurate determination of thermal neutron flux via cryogenic calorimetry [J].
Richardson, JM ;
Snow, WM ;
Chowdhuri, Z ;
Greene, GL .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (03) :550-555
[46]  
RICHARDSON JM, 1993, THESIS HARVARD U
[47]   CALORIMETRIC MEASUREMENT OF THERMAL-NEUTRON FLUX [J].
ROBERTSON, RGH ;
KOEHLER, PE .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1986, 251 (02) :307-312
[48]   BOLOMETER FOR MEASUREMENTS ON HIGH-TEMPERATURE PLASMAS [J].
SCHIVELL, J ;
RENDA, G ;
LOWRANCE, J ;
HSUAN, H .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1982, 53 (10) :1527-1534
[49]   ANNEALING EXPERIMENTS ON PURE LEAD AFTER ELECTRON-IRRADIATION AT 4.7 K AND BELOW 3 K [J].
SCHROEDER, H ;
SCHILLING, W .
RADIATION EFFECTS AND DEFECTS IN SOLIDS, 1976, 30 (04) :243-253
[50]  
SCHWARZ W, 1987, PHYS REV LETT, V58, P1228