A wide dynamic range, high-spatial-resolution scanning system for radiochromic dye films

被引:48
作者
Aydarous, AS [1 ]
Darley, PJ [1 ]
Charles, MW [1 ]
机构
[1] Univ Birmingham, Sch Phys & Astron, Birmingham B15 2TT, W Midlands, England
关键词
D O I
10.1088/0031-9155/46/5/304
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper describes the evaluation of an inexpensive, commercially available 35 mm transparency slide scanner as a potential alternative scanning device for GafChromic (TM) HD-810 radiochromic dye film. Besides its low cost, the principal advantages of this type of scanner are high spatial resolution and high speed (a typical scan taking less than 1 min). With broad-band illumination the useful dose range using grey-scale imaging of GafChromic (TM) HD-810 is limited to about 50-800 Gy. By using the colour-scale imaging capability of the scanner we have been able to achieve a significant extension covering a similar range(15-2000 Gy) to that attainable using monochromatic illumination. The short-term reproducibility of the system is good, with a coefficient of variation of doses estimated from repeat scanning of uniformly exposed calibration films of less than 2%. Long-term stability is ensured by the scanning of a manufacturer-supplied test slide. The slide scanner system has been used in the determination of depth dose distributions from a model 'hot particle' source containing (RU)-R-106/Rh. GafChromic (TM) dye film stacks irradiated by the source were read out on both the slide scanner and a conventional Joyce Loebl MDM6 scanning stage microdensitometer. The overall agreement between the dose estimates provided by the two systems was within 10%.
引用
收藏
页码:1379 / 1389
页数:11
相关论文
共 12 条
[1]  
CHARLES MW, 1991, RADIAT PROT DOSIM, V39, P39
[2]  
CHU RDH, 1990, RADIAT PHYS CHEM, V35, P767
[3]  
DARLEY PJ, 1991, RADIAT PROT DOSIM, V39, P61
[4]   Origins and dosimetry of 'hot particles' from nuclear plant operation [J].
Darley, PJ ;
Charles, MW ;
Othman, IE ;
Al-Aydarous, AS ;
Mill, AJ .
RADIATION PROTECTION DOSIMETRY, 2000, 92 (1-3) :131-137
[5]  
GANN RG, 1999, DESKTOP SCANNERS IMA
[6]  
KAURIN DG, 1996, P 9 INT C INT RAD PR, V3, P28
[7]   Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55 [J].
Niroomand-Rad, A ;
Blackwell, CR ;
Coursey, BM ;
Gall, KP ;
Galvin, JM ;
McLaughlin, WL ;
Meigooni, AS ;
Nath, R ;
Rodgers, JE ;
Soares, CG .
MEDICAL PHYSICS, 1998, 25 (11) :2093-2115
[8]   Verification of absolute dose rates for intravascular brachytherapy beta sources [J].
Piessens, M ;
Reynaert, N .
PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (08) :2219-2231
[9]  
SOARES CG, 1991, RADIAT PROT DOSIM, V39, P55
[10]   High-resolution dosimetry using radiochromic film and a document scanner [J].
Stevens, MA ;
Turner, JR ;
Hugtenburg, RP ;
Butler, PH .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (11) :2357-2365