Verification of absolute dose rates for intravascular brachytherapy beta sources

被引:12
作者
Piessens, M
Reynaert, N
机构
[1] Onze Lieve Vrouw Hosp, Dept Radiat Therapy & Oncol, B-9300 Aalst, Belgium
[2] State Univ Ghent, Standard Dosimetry Lab, B-9000 Ghent, Belgium
关键词
D O I
10.1088/0031-9155/45/8/311
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The absolute dose rate around beta sources for intravascular brachytherapy was measured with thin thermoluminescent (TL) dosimeters and radiochromic film. The calibration of these dosimeters was performed at the dose maximum in a 6 MeV electron beam of a linear accelerator. Compared to the values given by the manufacturers, higher dose rates of about 20% for Sr-90/Y-90 source trains and 20 to 40% for Y-90 line sources were found. We also measured the dose rate for a planar Sr-90/Y-90 source, which was calibrated at NIST (National Institute of Standards and Technology). This source serves as a reference for the determination of the dose rates by the supplier of the Y-90 sources. For this source we measured a 13% higher dose rate compared to the value given in the calibration report by NIST. Our observation can have important consequences for the interpretation and comparison of the dose-related clinical results of intravascular brachytherapy with these and other beta sources. Also the comparison of the clinical results obtained with beta sources and gamma sources, i.e. Ir-192, depends largely on the accuracy of the absolute dosimetry of the beta sources.
引用
收藏
页码:2219 / 2231
页数:13
相关论文
共 9 条
[1]  
BRIESMEISTER JF, 1997, LA12625M LOS ALM NAT
[2]   Radiochromic film dosimetry of a high dose rate beta source for intravascular brachytherapy [J].
Duggan, DM ;
Coffey, CW ;
Lobdell, JL ;
Schell, MC .
MEDICAL PHYSICS, 1999, 26 (11) :2461-2464
[3]  
ICRU, 1997, 56 ICRU
[4]   Intravascular brachytherapy physics: Report of the AAPM Radiation Therapy Committee Task Group No. 60 [J].
Nath, R ;
Amols, H ;
Coffey, C ;
Duggan, D ;
Jani, S ;
Li, ZF ;
Schell, M ;
Soares, C ;
Whiting, J ;
Cole, PE ;
Crocker, I ;
Schwartz, R .
MEDICAL PHYSICS, 1999, 26 (02) :119-152
[5]  
*NCS, 1989, 5 NCS
[6]   Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55 [J].
Niroomand-Rad, A ;
Blackwell, CR ;
Coursey, BM ;
Gall, KP ;
Galvin, JM ;
McLaughlin, WL ;
Meigooni, AS ;
Nath, R ;
Rodgers, JE ;
Soares, CG .
MEDICAL PHYSICS, 1998, 25 (11) :2093-2115
[7]   A novel system for intracoronary beta-irradiation: Description and dosimetric results [J].
Popowski, Y ;
Verin, V ;
Schwager, M ;
Nouet, P ;
Papirov, I ;
Rouzaud, M ;
Urban, P ;
Landis, JR ;
Rutishauser, W ;
Kurtz, JM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 36 (04) :923-931
[8]   HIGH-DOSE-RATE BRACHYTHERAPY FOR PREVENTION OF RESTENOSIS AFTER PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY - PRELIMINARY DOSIMETRIC TESTS OF A NEW SOURCE PRESENTATION [J].
POPOWSKI, Y ;
VERIN, V ;
PAPIROV, I ;
NOUET, P ;
ROUZAUD, M ;
GROB, E ;
SCHWAGER, M ;
URBAN, P ;
RUTISHAUSER, W ;
KURTZ, JM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 33 (01) :211-215
[9]   Calibration and characterization of beta-particle sources for intravascular brachytherapy [J].
Soares, CG ;
Halpern, DG ;
Wang, CK .
MEDICAL PHYSICS, 1998, 25 (03) :339-346