Radiochromic film dosimetry of a high dose rate beta source for intravascular brachytherapy

被引:27
作者
Duggan, DM [1 ]
Coffey, CW
Lobdell, JL
Schell, MC
机构
[1] Vanderbilt Univ, Med Ctr, Ctr Radiat Oncol, Nashville, TN 37232 USA
[2] Novoste Corp, Norcross, GA 30093 USA
[3] Univ Rochester, Med Ctr, Dept Radiat Oncol, Rochester, NY 14642 USA
关键词
intravascular brachytherapy; Sr-90; radiochromic film; brachytherapy dosimetry; quality assurance;
D O I
10.1118/1.598814
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Good clinical physics practice requires that dose rates of brachytherapy sources be checked by the institution using them, as recommended by American Association of Physicists in Medicine Task Group 56 and The American College of Radiology. For intravascular brachytherapy with catheter-based systems, AAPM Task Group 60 recommends that the dose rate be measured at a reference point located at a radial distance of 2 mm from the center of the catheter axis. AAPM Task Group 60 also recommends that the dose rate along the catheter axis at a radial distance of 2 mm should be uniform to within +/- 10% in the center two-thirds of the treated length, and the relative dose rate in the plane perpendicular to the catheter axis through the center of the source should be measured at distances from 0.5 mm to R-90 (the distance from a point source within which 90% of the energy is deposited) at intervals of 0.5 mm. Radiochromic him dosimetry has been used to measure the dose distribution in a plane parallel to and at a radial distance of 2 mm from the axis of a novel, catheter-based, beta source for intravascular brachytherapy. The dose rate was averaged along a line parallel to the catheter axis at a radial distance of 2 mm, in the centered 24.5 mm of the treated length. This average dose rate agreed with the dose rate measured with a well ionization chamber by the replacement method using source trains calibrated with an extrapolation chamber at the National institute of Standards and Technology. Ail of the dose rates in the centered 24.5 mm of a line parallel to the axis at a distance of 2 mm were within +/- 10% of the average. (C) 1999 American Association of Physicists in Medicine. [S0094-2405(99)01711-3].
引用
收藏
页码:2461 / 2464
页数:4
相关论文
共 17 条
[1]  
*AM COLL RAD, 1997, 1997 ACR STAND
[2]   Dosimetric considerations for catheter-based beta and gamma emitters in the therapy of neointimal hyperplasia in human coronary arteries [J].
Amols, HI ;
Zaider, M ;
Weinberger, J ;
Ennis, R ;
Schiff, PB ;
Reinstein, LE .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 36 (04) :913-921
[3]  
ATTIX F. H., 1986, Introduction to radiological physics and radiation dosimetry
[4]   Dose distribution for a 32P-impregnated coronary stent:: Comparison of theoretical calculations and measurements with radiochromic film [J].
Duggan, DM ;
Coffey, CW ;
Levit, S .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 40 (03) :713-720
[5]   Radiation dose from a phosphorous-32 impregnated wire mesh vascular stent [J].
Janicki, C ;
Duggan, DM ;
Coffey, CW ;
Fischell, DR ;
Fischell, TA .
MEDICAL PHYSICS, 1997, 24 (03) :437-445
[6]   Endovascular β-radiation to reduce restenosis after coronary balloon angioplasty -: Results of the beta energy restenosis trial (BERT) [J].
King, SB ;
Williams, DO ;
Chougule, P ;
Klein, JL ;
Waksman, R ;
Hilstead, R ;
Macdonald, J ;
Anderberg, K ;
Crocker, IR .
CIRCULATION, 1998, 97 (20) :2025-2030
[7]   FRACTION OF IONIZATION FROM ELECTRONS ARISING IN THE WALL OF AN IONIZATION-CHAMBER [J].
LEMPERT, GD ;
NATH, R ;
SCHULZ, RJ .
MEDICAL PHYSICS, 1983, 10 (01) :1-3
[8]   Characterization of a positron emitting V48 nitinol stent for intracoronary brachytherapy [J].
Li, AN ;
Eigler, NL ;
Litvack, F ;
Whiting, JS .
MEDICAL PHYSICS, 1998, 25 (01) :20-28
[9]   SENSITOMETRY OF THE RESPONSE OF A NEW RADIOCHROMIC FILM DOSIMETER TO GAMMA-RADIATION AND ELECTRON-BEAMS [J].
MCLAUGHLIN, WL ;
YUNDONG, C ;
SOARES, CG ;
MILLER, A ;
VANDYK, G ;
LEWIS, DF .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1991, 302 (01) :165-176
[10]   PHOTON ENERGY-DEPENDENCE OF THE SENSITIVITY OF RADIOCHROMIC FILM AND COMPARISON WITH SILVER-HALIDE FILM AND LIF TLDS USED FOR BRACHYTHERAPY DOSIMETRY [J].
MUENCH, PJ ;
MEIGOONI, AS ;
NATH, R ;
MCLAUGHLIN, WL .
MEDICAL PHYSICS, 1991, 18 (04) :769-775