Biomechanical Analysis of Pedicle Screws in Osteoporotic Bone With Bioactive Cement Augmentation Using Simulated In Vivo Multicomponent Loading

被引:68
作者
Choma, Theodore J. [1 ]
Frevert, Wesley F. [2 ]
Carson, William L. [2 ]
Waters, Nicole P. [2 ]
Pfeiffer, Ferris M. [3 ]
机构
[1] Univ Missouri, Dept Orthopaed Surg, Missouri Spine Ctr, Columbia, MO 65212 USA
[2] Univ Missouri, Comparat Orthopaed Lab, Columbia, MO 65212 USA
[3] Pfeiffer Engn LLC, Boonville, MO USA
关键词
pedicle screw; loosening; toggle-migration; pullout; osteoporosis; augmentation; bioabsorbable cement; biomechanical study; CALCIUM-PHOSPHATE CEMENT; PULLOUT STRENGTH; APATITE CEMENT; LUMBAR SPINE; FIXATION; INJECTION;
D O I
10.1097/BRS.0b013e3181d449ec
中图分类号
R74 [神经病学与精神病学];
学科分类号
100204 [神经病学];
摘要
Study Design. Biomechanical analysis of bioactive cements augmenting pedicle screw resistance to loosening in osteoporotic synthetic bone. Objective. To simulate in vivo loading-loosening of pedicle screws in osteoporotic vertebrae; and to compare biomechanical efficacy of the following bioactive cements: calcium phosphate (CP), calcium sulfate (CS), and proprietary mixture (M). Summary of Background Data. Pedicle screw instrumentation in osteoporotic spines is limited by poor bone-screw interface strength, resulting in screw loosening fixation failure. Previous in vivo studies evaluated augmented pedicle screw resistance to pure pullout, not simulating in vivo loading/failure. Methods. A pedicle screw-instrumented osteoporotic thoracic vertebra subjected to combined pullout, transverse, moment loading was simulated. Unconstrained 3-dimensional screw motion relative to vertebra was optically measured during quasi-static, and dynamic loading. Results. Augmented groups (CP, CS, M) produced (P < 8.0E-07) higher quasi-static failure initiation force (61.2,45.6, 40.3 N) than those by the nonaugmented group (21.0 N), with no significant difference in small screw displacement up to these loads. Nonaugmented screw motion after failure initiation was primarily rotation (toggle-migration) with minimal pullout until the screw tip contacted the superior endplate, followed by more prominent screw pullout. Augmented screw motion (with cement remaining intact on screw) was similar, but with eventual bone fracture anterior to the pedicle region. Dynamic loading produced similar failure initiation force and screw motion. Conclusion. We believe our test protocol produced screw loosening failure similar to that observed clinically, and that it has the ability to detect differences in failure initiation force and failure modes to compare short-term efficacy of screw augmentation techniques. All cements improved screw resistance to failure. The CP > CS > M failure initiation force (P < 0.006) was because of differences in cement distribution. Animal studies may be required to characterize the remodeling activity of bioactive cements and their longer term efficacies.
引用
收藏
页码:454 / 462
页数:9
相关论文
共 19 条
[1]
Assessment of different screw augmentation techniques and screw designs in osteoporotic spines [J].
Becker, S. ;
Chavanne, A. ;
Spitaler, R. ;
Kropik, K. ;
Aigner, N. ;
Ogon, M. ;
Redl, H. .
EUROPEAN SPINE JOURNAL, 2008, 17 (11) :1462-1469
[2]
Cook Stephen D, 2004, Spine J, V4, P402, DOI 10.1016/j.spinee.2003.11.010
[3]
Dawson J. M., 2003, SPINAL IMPLANTS ARE
[4]
Biomechanical comparison of anatomic trajectory pedicle screw versus injectable calcium sulfate graft-augmented pedicle screw for salvage in cadaveric thoracic bone [J].
Derincek, A ;
Wu, C ;
Mehbod, A ;
Transfeldt, EE .
JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2006, 19 (04) :286-291
[5]
Bone cement or bone substitute augmentation of pedicle screws improves pullout strength in posterior spinal fixation [J].
Evans, SL ;
Hunt, CM ;
Ahuja, S .
JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 2002, 13 (12) :1143-1145
[6]
EFFECTS OF BONE-MINERAL DENSITY ON PEDICLE SCREW FIXATION [J].
HALVORSON, TL ;
KELLEY, LA ;
THOMAS, KA ;
WHITECLOUD, TS ;
COOK, SD .
SPINE, 1994, 19 (21) :2415-2420
[7]
Pullout strength of pedicle screws augmented with particulate calcium phosphate: An experimental study [J].
Hashemi, Ata ;
Bednar, Drew ;
Ziada, Samir .
SPINE JOURNAL, 2009, 9 (05) :404-410
[8]
Mechanical properties of open-cell foam synthetic thoracic vertebrae [J].
Johnson, Amy E. ;
Keller, Tony S. .
JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 2008, 19 (03) :1317-1323
[9]
Konno S, 1994, Eur Spine J, V3, P299, DOI 10.1007/BF02200140
[10]
Carbonated apatite cement augmentation of pedicle screw fixation in the lumbar spine [J].
Lotz, JC ;
Hu, SS ;
Chiu, DFM ;
Yu, M ;
Colliou, O ;
Poser, RD .
SPINE, 1997, 22 (23) :2716-2723