Biomodel-guided stereotaxy

被引:17
作者
D'Urso, PS [1 ]
Hall, BI [1 ]
Atkinson, RL [1 ]
Weidmann, MJ [1 ]
Redmond, MJ [1 ]
机构
[1] Princess Alexandra Hosp, Dept Neurol Sci, Brisbane, Qld 4102, Australia
关键词
biomodeling; rapid prototyping; stereolithography; stereotaxy;
D O I
10.1097/00006123-199905000-00080
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
OBJECTIVES: To simplify the practice of stereotactic surgery by using an original method, apparatus, and solid anatomic replica for trajectory planning and to validate the method and apparatus in a laboratory and clinical trial. METHODS: The patient is marked with fiducials and scanned by using computed tomography or magnetic resonance imaging. The three-dimensional data are converted to a format acceptable to stereolithography. Stereolithography uses a laser to polymerize photosensitive resin into a solid plastic model (biomodel). Stereolithography can replicate brood vessels, soft tissue, tumor, and bone accurately (<0.8 mm). A stereotactic apparatus is referenced to fiducials replicated in the biomodel. The trajectory for the intervention is determined and saved. The apparatus is attached to the patient fiducials, and the intervention is replicated. RESULTS: Three types of apparatus (template, Brown-Roberts-Wells frame, and D'Urso frame) were tested on phantoms and patients requiring the excision/biopsy of tumors. The localization errors determined from the phantom studies were template, 0.82 mm; Brown-Roberts-Wells frame, 1.17 mm; and D'Urso frame, 0.89 mm. The surgeons reported that clinical use of the template and D'Urso frame was accurate and ergonomic. The Bt own-Roberts-Wells frame was move difficult to use and somewhat inaccurate. CONCLUSION: Biomodel-guided stereotaxy has significant advantages. It is performed quickly; it is based on simple, intuitive methodology; it enhances visualization of anatomy and trajectory planning; it enhances patient understanding; it uses inexpensive equipment; it does not require rigid head fixation; and it has greater versatility than known techniques. Disadvantages are biomodel cost and a manufacturing time of 12 to 24 hours.
引用
收藏
页码:1084 / 1093
页数:10
相关论文
共 19 条
[1]  
Anichkov A. D., 1977, US Patent, Patent No. [4,228,799, 4228799]
[2]   COMPUTED TOMOGRAPHIC GUIDANCE STEREOTAXIS IN THE MANAGEMENT OF INTRACRANIAL MASS LESIONS [J].
APUZZO, MLJ ;
SABSHIN, JK .
NEUROSURGERY, 1983, 12 (03) :277-285
[3]   INTRAOPERATIVE LOCALIZATION USING AN ARMLESS, FRAMELESS STEREOTAXIC WAND - TECHNICAL NOTE [J].
BARNETT, GH ;
KORMOS, DW ;
STEINER, CP ;
WEISENBERGER, J .
JOURNAL OF NEUROSURGERY, 1993, 78 (03) :510-514
[5]  
D'Urso P. S, 1993, US Patent, Patent No. [5 741 215, 5741215]
[6]   FRAMELESS STEREOTAXY IN CHILDREN [J].
DRAKE, JM ;
PRUDENCIO, J ;
HOLOWKA, S ;
RUTKA, JT ;
HOFFMAN, HJ ;
HUMPHREYS, RP .
PEDIATRIC NEUROSURGERY, 1994, 20 (02) :152-159
[7]  
DURSO PS, 1993, Patent No. 5752962
[8]   STEREOTAXIC VERSUS STEREOTAXIC [J].
GILDENBERG, PL .
NEUROSURGERY, 1993, 32 (06) :965-966
[9]  
GUTHRIE BL, 1992, CLIN NEUR, V38, P112
[10]   PRELIMINARY EXPERIENCE WITH BROWN-ROBERTS-WELLS (BRW) COMPUTERIZED-TOMOGRAPHY STEREOTAXIC GUIDANCE-SYSTEM [J].
HEILBRUN, MP ;
ROBERTS, TS ;
APUZZO, MLJ ;
WELLS, TH ;
SABSHIN, JK .
JOURNAL OF NEUROSURGERY, 1983, 59 (02) :217-222