Comprehensive two-dimensional gas chromatography with isotope dilution time-of-flight mass spectrometry for the measurement of dioxins and polychlorinated biphenyls in foodstuffs - Comparison with other methods

被引:59
作者
Focant, JF
Eppe, G
Scippo, ML
Massart, AC
Pirard, C
Maghuin-Rogister, G
De Pauw, E
机构
[1] Univ Liege, Dept Chem, CART Mass Spect Lab, B-4000 Liege, Belgium
[2] Univ Liege, CART Lab Anal Foodstuffs Anim Origin, B-4000 Liege, Belgium
关键词
comprehensive two-dimensional gas chromatography (GC x GC); time-of-flight mass spectrometry (TOF-MS); quadrupole ion storage mass spectrometry (QIST-MS); tandem-in-time mass spectrometry (MS/MS); high-resolution mass spectrometry (HRMS); dioxin-responsive chemically activated; luciferase gene expression (DR-CALUX); polychlorinated dibenzo-p-dioxins (PCDDs); polychlorinated dibenzofurans (PCDFs); polychlorinated biphenyls (PCBs);
D O I
10.1016/j.chroma.2005.05.090
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A comprehensive two-dimensional gas chromatography time-of-flight mass spectrometry (GC x GC-TOF-MS) experimental setup was tested for the measurement of seven 2,3,7,8-substituted polychlorinated dibenzo-p-dioxins (PCDDs), ten 2,3,7,8-substituted polychlorinated dibenzofurans (PCDFs), four non-ortho-polychlorinated biphenyls (PCBs), eight mono-ortho-PCBs, and six indicator PCBs (Aroclor 1260) in foodstuff samples. A 40 m RTX-500 (0.18 mm I.D., 0.10 mu m df) was used as the first dimension (D-1) and a 1.5 nn BPX-50 (0.10 mm I.D., 0.10 mu m df) as the second dimension (2 D). The GC x GC chromatographic separation was completed in 45 min. Quantification was performed using C-13-label isotope dilution (11)). Isotope ratios of the selected quantification ions were checked against theoretical values prior to peak assignment and quantification. The dynamic working range spanned three orders of magnitude. The lowest detectable amount of 2,3,7,8-TCDD was 0.2 pg. Fish, pork, and milk samples were considered. On a congener basis, the GC x GC-ID-TOF-MS method was compared to the reference GC-ID high resolution mass spectrometry (HRMS) method and to the alternative GC-ID tandem-in-time quadrupole ion storage mass spectrometry (QIST-MS/MS). PCB levels ranged from low picogram (pg) to low nanogram (ng) per gram of sample and data compared very well between the different methods. For all matrices, PCDD/Fs were at a low pg level (0.05-3 pg) on a fresh weight basis. Although congener profiles were accurately described, RSDs of GC x GC-ID-TOF-MS and GC-QIST-MS/MS were much higher than for GC-ID-HRMS, especially for low level pork and milk. On a toxic equivalent (TEQ) basis, all methods, including the dioxin-responsive chemically activated luciferase gene expression (DR-CALUX) assay, produced similar responses. A cost comparison is also presented. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:45 / 60
页数:16
相关论文
共 37 条
[1]  
Aarts J., 1993, ORGANOHALOGEN COMPOU, V13, P361
[2]  
AARTS JMMJG, 1995, EUR J PHARM-ENVIRON, V293, P463
[3]  
[Anonymous], OFFIC J EUR COMMUN
[4]   Bioanalytical screening methods for dioxins and dioxin-like compounds - a review of bioassay/biomarker technology [J].
Behnisch, PA ;
Hosoe, K ;
Sakai, S .
ENVIRONMENT INTERNATIONAL, 2001, 27 (05) :413-439
[5]   REVIEW OF THE DIOXIN PROBLEM - MASS-SPECTROMETRIC ANALYSES OF TETRACHLORODIOXINS IN ENVIRONMENTAL-SAMPLES [J].
CAIRNS, T ;
FISHBEIN, L ;
MITCHUM, RK .
BIOMEDICAL MASS SPECTROMETRY, 1980, 7 (11-1) :484-492
[6]   Comprehensive two-dimensional gas chromatography:: a powerful and versatile analytical tool [J].
Dallüge, J ;
Beens, J ;
Brinkman, UAT .
JOURNAL OF CHROMATOGRAPHY A, 2003, 1000 (1-2) :69-108
[7]   GC x GC [J].
Dimandja, JMD .
ANALYTICAL CHEMISTRY, 2004, 76 (09) :167A-174A
[8]   PTV-LV-GC/MS/MS as screening and complementary method to HRMS for the monitoring of dioxin levels in food and feed [J].
Eppe, G ;
Focant, JF ;
Pirard, C ;
De Pauw, E .
TALANTA, 2004, 63 (05) :1135-1146
[9]   Performances and limitations of the HRMS method for dioxins, furans and dioxin-like PCBs analysis in animal feedingstuffs Part II: Does it comply with the European pro-active approach? [J].
Eppe, G ;
Maghuin-Rogister, G ;
De Pauw, E .
ANALYTICA CHIMICA ACTA, 2004, 519 (02) :243-253
[10]   Recent advances in mass spectrometric measurement of dioxins [J].
Focant, JF ;
Pirard, C ;
Eppe, G ;
De Pauw, E .
JOURNAL OF CHROMATOGRAPHY A, 2005, 1067 (1-2) :265-275