A wireless microsystem for the remote sensing of pressure, temperature, and relative humidity

被引:152
作者
DeHennis, AD [1 ]
Wise, KD [1 ]
机构
[1] Univ Michigan, Ctr Wireless Integrated Microsyst, Dept Elect Engn & Comp Sci, Ann Arbor, MI 48109 USA
基金
美国国家科学基金会;
关键词
anodic bonding; backscatter modulation; bimorph sensor; capacitive interface; electrical feedthrough; eutectic bonding; passive telemetry; relaxation oscillator; wireless sensor;
D O I
10.1109/JMEMS.2004.839650
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper presents a microsystem that utilizes inductive power and data transfer through a backscatter-modulated carrier and a transducer interface that monitors its environment through embedded capacitive transducers. Formed on a single chip, transducers for temperature, pressure, and relative humidity are realized using a silicon-on-glass process that combines anodic bonding and a silicon-gold eutectic to realize vacuum-sealed cavities with low-impedance (6 Omega) electrical feedthroughs. Temperature is sensed capacitively using a row of Si/Au bimorph beams that produce a sensitivity of 15 fF/degreesC from 20 to 100 degreesC. The absolute pressure sensors have a sensitivity of 15 fF/torr and a range from 500 to 1200 torr, while the relative humidity sensor responds with 39 fF/%RH from 20 to 95%RH. A relaxation oscillator implements low-power capacitance-to-frequency conversion on a second chip with a sensitivity of 750 Hz/pF at 10 kHz, forming a 341 muW transducer interface. The system is remotely powered by a 3-MHz carrier and has a volume of 32 mm(3), including the hybrid antenna wound around the perimeter of the system.
引用
收藏
页码:12 / 22
页数:11
相关论文
共 34 条
[1]  
AKAR O, 2000, P 14 EUR C SOL STAT, P585
[2]   A wireless implantable multichannel digital neural recording system for a micromachined sieve electrode [J].
Akin, T ;
Najafi, K ;
Bradley, RM .
IEEE JOURNAL OF SOLID-STATE CIRCUITS, 1998, 33 (01) :109-118
[3]  
AKIN T, 1994, 7TH MEDITERRANEAN ELECTROTECHNICAL CONFERENCE, VOLS 1-3, P545, DOI 10.1109/MELCON.1994.381035
[4]  
CHANG PC, 2002, P 34 ANN PREC TIM TI, P493
[5]   A miniature pressure system with a capacitive sensor and a passive telemetry link for use in implantable applications [J].
Chatzandroulis, S ;
Tsoukalas, D ;
Neukomm, PA .
JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2000, 9 (01) :18-23
[6]   Batch-processed vacuum-sealed capacitive pressure sensors [J].
Chavan, AV ;
Wise, KD .
JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2001, 10 (04) :580-588
[7]  
CHO S, 1998, Patent No. 5837562
[8]  
DeHennis A, 2003, BOSTON TRANSDUCERS'03: DIGEST OF TECHNICAL PAPERS, VOLS 1 AND 2, P1860
[9]  
DeHennis A, 2002, PROC IEEE MICR ELECT, P252, DOI 10.1109/MEMSYS.2002.984250
[10]  
DEHENNIS A, 2002, P SOL STAT SENS ACT, P165