A hip joint simulator study using new and physiologically scratched femoral heads with ultra-high molecular weight polyethylene acetabular cups

被引:51
作者
Barbour, PSM
Stone, MH
Fisher, J [1 ]
机构
[1] Univ Leeds, Sch Mech Engn, Leeds LS2 9JT, W Yorkshire, England
[2] Leeds Gen Infirm, Leeds, W Yorkshire, England
关键词
hip joint simulator; wear; UHMWPE; load conditions; scratch-damaged heads;
D O I
10.1243/0954411001535598
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study validates a hip joint simulator configuration as compared with other machines and clinical wear rates using smooth metal and ceramic femoral heads and ultra-high molecular weight polyethylene (UIIMWPE) acetabular cups. Secondly the wear rate of UHMWPE cups is measured in the simulator with deliberately scratched cobalt-chrome heads to represent the type of mild and severe scratch damage found on retrieved heads. Finally, the scratching processes are described and the resulting scratches compared with those found in retrieved cobalt-chrome heads. For smooth cobalt-chrome and zirconia heads the wear rates were found to be statistically similar to other simulator machines and within the normal range found from clinical studies. An increased wear rate was found with cobalt-chrome heads scratched using either the diamond stylus or the bead cobalt-chrome but the greatest increase was with the diamond scratched heads which generated scratches of similar dimensions to those on retrieved heads. A greater than twofold increase in wear rate is reported for these heads when compared with smooth heads. This increased wear rate is, however, still within the limits of data from clinical wear studies.
引用
收藏
页码:569 / 576
页数:8
相关论文
共 31 条
[1]   CORROSION AND WEAR OF 304-STAINLESS STEEL USING A SCRATCH TEST [J].
ADLER, TA ;
WALTERS, RP .
CORROSION SCIENCE, 1992, 33 (12) :1855-1876
[2]  
[Anonymous], CLIN ORTHOPAEDICS
[3]   LABORATORY WEAR TESTS AND CLINICAL OBSERVATIONS OF THE PENETRATION OF FEMORAL HEADS INTO ACETABULAR CUPS IN TOTAL REPLACEMENT HIP JOINTS .3. THE MEASUREMENT OF INTERNAL VOLUME CHANGES IN EXPLANTED CHARNLEY SOCKETS AFTER 2-16 YEARS INVIVO AND THE DETERMINATION OF WEAR FACTORS [J].
ATKINSON, JR ;
DOWSON, D ;
ISAAC, JH ;
WROBLEWSKI, BM .
WEAR, 1985, 104 (03) :225-244
[4]   A hip joint simulator study using simplified loading and motion cycles generating physiological wear paths and rates [J].
Barbour, PSM ;
Stone, MH ;
Fisher, J .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1999, 213 (H6) :455-467
[5]  
Bragdon C R, 1996, Proc Inst Mech Eng H, V210, P157, DOI 10.1243/PIME_PROC_1996_210_408_02
[6]   Influence of packaging conditions on the properties of gamma-irradiated UHMWPE following accelerated ageing and shelf ageing [J].
Buchanan, FJ ;
Sim, B ;
Downes, S .
BIOMATERIALS, 1999, 20 (09) :823-837
[7]  
Cooper J. R., 1993, Clinical Materials, V14, P295, DOI 10.1016/0267-6605(93)90016-Z
[8]   PROPERTY CHANGES OF UHMW POLYETHYLENE HIP CUP ENDOPROSTHESES DURING IMPLANTATION [J].
EYERER, P ;
KE, YC .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1984, 18 (09) :1137-1151
[9]  
Hailey J L, 1996, Proc Inst Mech Eng H, V210, P3, DOI 10.1243/PIME_PROC_1996_210_385_02
[10]  
Hall R M, 1996, Proc Inst Mech Eng H, V210, P197, DOI 10.1243/PIME_PROC_1996_210_413_02