A hip joint simulator study using new and physiologically scratched femoral heads with ultra-high molecular weight polyethylene acetabular cups

被引:51
作者
Barbour, PSM
Stone, MH
Fisher, J [1 ]
机构
[1] Univ Leeds, Sch Mech Engn, Leeds LS2 9JT, W Yorkshire, England
[2] Leeds Gen Infirm, Leeds, W Yorkshire, England
关键词
hip joint simulator; wear; UHMWPE; load conditions; scratch-damaged heads;
D O I
10.1243/0954411001535598
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This study validates a hip joint simulator configuration as compared with other machines and clinical wear rates using smooth metal and ceramic femoral heads and ultra-high molecular weight polyethylene (UIIMWPE) acetabular cups. Secondly the wear rate of UHMWPE cups is measured in the simulator with deliberately scratched cobalt-chrome heads to represent the type of mild and severe scratch damage found on retrieved heads. Finally, the scratching processes are described and the resulting scratches compared with those found in retrieved cobalt-chrome heads. For smooth cobalt-chrome and zirconia heads the wear rates were found to be statistically similar to other simulator machines and within the normal range found from clinical studies. An increased wear rate was found with cobalt-chrome heads scratched using either the diamond stylus or the bead cobalt-chrome but the greatest increase was with the diamond scratched heads which generated scratches of similar dimensions to those on retrieved heads. A greater than twofold increase in wear rate is reported for these heads when compared with smooth heads. This increased wear rate is, however, still within the limits of data from clinical wear studies.
引用
收藏
页码:569 / 576
页数:8
相关论文
共 31 条
[11]  
Isaac G H, 1987, Eng Med, V16, P167
[12]   SURFACE DAMAGE TO COBALT-CHROME FEMORAL-HEAD PROSTHESES [J].
JASTY, M ;
BRAGDON, CR ;
LEE, K ;
HANSON, A ;
HARRIS, WH .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (01) :73-77
[13]   Polyethylene wear in Scanhip(R) arthroplasty with a 22 or 32 mm head - 62 matched patients followed for 7-9 years [J].
Kesteris, U ;
Ilchmann, T ;
Wingstrand, H ;
Onnerfalt, R .
ACTA ORTHOPAEDICA SCANDINAVICA, 1996, 67 (02) :125-127
[14]   Advances in the processing, sterilization, and crosslinking of ultra-high molecular weight polyethylene for total joint arthroplasty [J].
Kurtz, SM ;
Muratoglu, OK ;
Evans, M ;
Edidin, AA .
BIOMATERIALS, 1999, 20 (18) :1659-1688
[15]   Femoral component wear in retrieved hip prostheses [J].
Kusaba, A ;
Kuroki, Y .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1997, 79B (02) :331-336
[16]  
Liao YS, 1999, J BIOMED MATER RES, V48, P465
[17]   EFFECT OF FEMORAL-HEAD SIZE ON WEAR OF THE POLYETHYLENE ACETABULAR COMPONENT [J].
LIVERMORE, J ;
ILSTRUP, D ;
MORREY, B .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1990, 72A (04) :518-528
[18]   THE WEAR BEHAVIOR OF ION-IMPLANTED TI-6AL-4V AGAINST UHMW POLYETHYLENE [J].
MCKELLOP, HA ;
ROSTLUND, TV .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1990, 24 (11) :1413-1425
[19]   Quantification of third-body damage and its effect on UHMWPE wear with different types of femoral head [J].
Minakawa, H ;
Stone, MH ;
Wroblewski, BM ;
Lancaster, JG ;
Ingham, E ;
Fisher, J .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1998, 80B (05) :894-899
[20]   Effect of sterilization on the structure and fatigue resistance of medical grade UHMWPE [J].
Pruitt, L ;
Ranganathan, R .
MATERIALS SCIENCE & ENGINEERING C-BIOMIMETIC MATERIALS SENSORS AND SYSTEMS, 1995, 3 (02) :91-93