Nodule detection in digital chest radiography:: Effect of nodule location

被引:30
作者
Håkansson, M
Båth, M
Börjesson, S
Kheddache, S
Flinck, A
Ullman, G
Månsson, LG
机构
[1] Sahlgrens Univ Hosp, Dept Med Phys & Biomed Engn, SE-41345 Gothenburg, Sweden
[2] Sahlgrens Univ Hosp, Dept Radiol, SE-41345 Gothenburg, Sweden
[3] Linkoping Univ, Dept Radiat Phys, SE-58185 Linkoping, Sweden
关键词
D O I
10.1093/rpd/nch524
中图分类号
X [环境科学、安全科学];
学科分类号
083001 [环境科学];
摘要
Most detection studies in chest radiography treat the entire chest image as a single background or divided Into the two regions parenchyma and mediastinum. However, the different parts of the lung show great variations in attenuation and structure, leading to different amounts of quantum noise and scattered radiation as well as different complexity. Detailed data on the difference in detectability in the different regions are of importance. The purpose of this study was to quantify the difference In detectability between different regions of a chest image. The chest X ray was divided into six different regions. where each region was considered to be uniform in terms of detectability. Thirty clinical chest images were collected and divided Into the different regions. Simulated designer nodules with a full-width-at-fifth-maximum of 10 mm but with varying contrast were added to the images. An equal number of images lacking pathology were included and a receiver operating characteristic (ROC) study was conducted with five observers. Results show that the image contrast needed to obtain a constant value of A, (area under an ROC curve) differs by more than a factor of four between different regions.
引用
收藏
页码:92 / 96
页数:5
相关论文
共 10 条
[1]
Nodule detection in digital chest radiography:: Part of image background acting as pure noise [J].
Båth, M ;
Håkansson, M ;
Börjesson, S ;
Kheddache, S ;
Grahn, A ;
Bochud, FO ;
Verdun, FR ;
Månsson, LG .
RADIATION PROTECTION DOSIMETRY, 2005, 114 (1-3) :102-108
[2]
Nodule detection in digital chest radiography:: Introduction to the radius chest trial [J].
Båth, M ;
Håkansson, M ;
Börjesson, S ;
Kheddache, S ;
Grahn, A ;
Ruschin, M ;
Tingberg, A ;
Mattsson, S ;
Månsson, LG .
RADIATION PROTECTION DOSIMETRY, 2005, 114 (1-3) :85-91
[3]
Nodule detection in digital chest radiography:: Effect of anatomical noise [J].
Båth, M ;
Håkansson, M ;
Börjesson, S ;
Hoeschen, C ;
Tischenko, O ;
Kheddache, S ;
Vikgren, J ;
Månsson, LG .
RADIATION PROTECTION DOSIMETRY, 2005, 114 (1-3) :109-113
[4]
A software tool for increased efficiency in observer performance studies in radiology [J].
Börjesson, S ;
Håkansson, M ;
Båth, M ;
Kheddache, S ;
Svensson, S ;
Tingberg, A ;
Grahn, A ;
Ruschin, M ;
Hemdal, B ;
Mattsson, S ;
Månsson, LG .
RADIATION PROTECTION DOSIMETRY, 2005, 114 (1-3) :45-52
[5]
BURGER C, 1997, HLTH CARE INFORMATIO, V2, P13
[6]
Nodule detection in digital chest radiography:: Summary of the radius chest trial [J].
Håkansson, M ;
Båth, M ;
Börjesson, S ;
Kheddache, S ;
Grahn, A ;
Ruschin, M ;
Tingberg, A ;
Mattsson, S ;
Månsson, LG .
RADIATION PROTECTION DOSIMETRY, 2005, 114 (1-3) :114-120
[7]
Nodule detection in digital chest radiography:: Effect of system noise [J].
Håkansson, M ;
Båth, M ;
Börjesson, S ;
Kheddache, S ;
Johnsson, ÅA ;
Månsson, LG .
RADIATION PROTECTION DOSIMETRY, 2005, 114 (1-3) :97-101
[8]
INFLUENCE OF STRUCTURED NOISE ON DETECTION OF RADIOLOGIC ABNORMALITIES [J].
REVESZ, G ;
KUNDEL, HL ;
GRABER, MA .
INVESTIGATIVE RADIOLOGY, 1974, 9 (06) :479-486
[9]
Distributions of scatter-to-primary and signal-to-noise ratios per pixel in digital chest imaging [J].
Ullman, G ;
Sandborg, M ;
Dance, DR ;
Hunt, R ;
Carlsson, GA .
RADIATION PROTECTION DOSIMETRY, 2005, 114 (1-3) :355-358
[10]
COMPUTERIZED 3-DIMENSIONAL SEGMENTED HUMAN ANATOMY [J].
ZUBAL, IG ;
HARRELL, CR ;
SMITH, EO ;
RATTNER, Z ;
GINDI, G ;
HOFFER, PB .
MEDICAL PHYSICS, 1994, 21 (02) :299-302