Effect of the forward-projected model-based iterative reconstruction solution algorithm on image quality and radiation dose in pediatric cardiac computed tomography

被引:39
作者
Nishiyama, Yukako [1 ]
Tada, Keiji [1 ]
Nishiyama, Yuichi [2 ]
Mori, Hiroshi [2 ]
Maruyama, Mitsunari [2 ]
Katsube, Takashi [2 ]
Yamamoto, Nobuko [2 ]
Kanayama, Hidekazu [1 ]
Yamamoto, Yasushi [1 ]
Kitagaki, Hajime [2 ]
机构
[1] Shimane Univ Hosp, Dept Radiol, 89-1 Enya Cho, Izumo, Shimane 6938501, Japan
[2] Shimane Univ, Fac Med, Dept Radiol, Matsue, Shimane, Japan
关键词
Cardiac; Children; Computed tomography; Congenital heart disease; Image quality; Iterative reconstruction; Phantom study; Radiation dose; CT; REDUCTION; ANGIOGRAPHY; EXPOSURE; PHANTOM; NOISE;
D O I
10.1007/s00247-016-3676-x
中图分类号
R72 [儿科学];
学科分类号
100202 [儿科学];
摘要
Some iterative reconstruction algorithms are useful for reducing the radiation dose in pediatric cardiac CT. A new iterative reconstruction algorithm (forward-projected model-based iterative reconstruction solution) has been developed, but its usefulness for radiation dose reduction in pediatric cardiac CT is unknown. To investigate the effect of the new algorithm on CT image quality and on radiation dose in pediatric cardiac CT. We obtained phantom data at six dose levels, as well as pediatric cardiac CT data, and reconstructed CT images using filtered back projection, adaptive iterative dose reduction 3-D (AIDR 3-D) and the new algorithm. We evaluated phantom image quality using physical assessment. Four radiologists performed visual evaluation of cardiac CT image quality. In the phantom study, the new algorithm effectively suppressed noise in the low-dose range and moderately generated modulation transfer function, yielding a higher signal-to-noise ratio compared with filtered back projection or AIDR 3-D. When clinical cardiac CT was performed, images obtained by the new method had less perceived image noise and better tissue contrast at similar resolution compared with AIDR 3-D images. The new algorithm reduced image noise at moderate resolution in low-dose CT scans and improved the perceived quality of cardiac CT images to some extent. This new algorithm might be superior to AIDR 3-D for radiation dose reduction in pediatric cardiac CT.
引用
收藏
页码:1663 / 1670
页数:8
相关论文
共 27 条
[1]
Iterative reconstruction methods in X-ray CT [J].
Beister, Marcel ;
Kolditz, Daniel ;
Kalender, Willi A. .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2012, 28 (02) :94-108
[2]
Simulated 50 % radiation dose reduction in coronary CT angiography using adaptive iterative dose reduction in three-dimensions (AIDR3D) [J].
Chen, Marcus Y. ;
Steigner, Michael L. ;
Leung, Steve W. ;
Kumamaru, Kanako K. ;
Schultz, Kurt ;
Mather, Richard T. ;
Arai, Andrew E. ;
Rybicki, Frank J. .
INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2013, 29 (05) :1167-1175
[3]
Radiation dose exposure in multislice computed tomography of the coronaries in comparison with conventional coronary angiography [J].
Dill, Thorsten ;
Deetjen, Anja ;
Ekinci, Okan ;
Moellmann, Susanne ;
Conradi, Guido ;
Kluge, Alexander ;
Weber, Christoph ;
Weber, Michael ;
Nef, Holger ;
Hamm, Christian W. .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 2008, 124 (03) :307-311
[4]
Dose reduction with adaptive statistical iterative reconstruction for paediatric CT: phantom study and clinical experience on chest and abdomen CT [J].
Gay, F. ;
Pavia, Y. ;
Pierrat, N. ;
Lasalle, S. ;
Neuenschwander, S. ;
Brisse, H. J. .
EUROPEAN RADIOLOGY, 2014, 24 (01) :102-111
[5]
Estimated Radiation Dose Associated With Cardiac CT Angiography [J].
Hausleiter, Joerg ;
Meyer, Tanja ;
Hermann, Franziska ;
Hadamitzky, Martin ;
Krebs, Markus ;
Gerber, Thomas C. ;
McCollough, Cynthia ;
Martinoff, Stefan ;
Kastrati, Adnan ;
Schoemig, Albert ;
Achenbach, Stephan .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2009, 301 (05) :500-507
[6]
Japan Network for Research and Information on Medical Exposure, 2015, DIAGN REF LEV BAS LA
[7]
Diagnostic Application of Multidetector-Row Computed Tomographic Coronary Angiography to Assess Coronary Abnormalities in Pediatric Patients: Comparison With Invasive Coronary Angiography [J].
Juan, Chien-Chang ;
Hwang, Betau ;
Lee, Pi-Chang ;
Meng, Chun-Chang Laura .
PEDIATRICS AND NEONATOLOGY, 2011, 52 (04) :208-213
[8]
Radiation Dose Reduction With Sinogram Affirmed Iterative Reconstruction Technique for Abdominal Computed Tomography [J].
Kalra, Mannudeep K. ;
Woisetschlager, Mischa ;
Dahlstrom, Nils ;
Singh, Sarabjeet ;
Lindblom, Maria ;
Choy, Garry ;
Quick, Petter ;
Schmidt, Bernhard ;
Sedlmair, Martin ;
Blake, Michael A. ;
Persson, Anders .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2012, 36 (03) :339-346
[9]
Dose reduction in pediatric abdominal CT: use of iterative reconstruction techniques across different CT platforms [J].
Khawaja, Ranish Deedar Ali ;
Singh, Sarabjeet ;
Otrakji, Alexi ;
Padole, Atul ;
Lim, Ruth ;
Nimkin, Katherine ;
Westra, Sjirk ;
Kalra, Mannudeep K. ;
Gee, Michael S. .
PEDIATRIC RADIOLOGY, 2015, 45 (07) :1046-1055
[10]
THE NOISE POWER SPECTRUM OF CT IMAGES [J].
KIJEWSKI, MF ;
JUDY, PF .
PHYSICS IN MEDICINE AND BIOLOGY, 1987, 32 (05) :565-575