Fracture of a Titanium Sleeve-encased Third-generation Ceramic Liner in a Modern THA

被引:10
作者
Chotai, Pranit N. [1 ]
Su, Edwin P. [1 ,2 ]
机构
[1] Hosp Special Surg, New York, NY 10021 USA
[2] Weill Cornell Med Coll, New York, NY USA
关键词
TOTAL HIP-ARTHROPLASTY; SANDWICH CUP;
D O I
10.3928/01477447-20110826-30
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
100224 [整形外科学];
摘要
Due to their excellent tribology, ceramics are increasingly used for total hip arthroplasty (THA) in young patients. Fracture rates for contemporary ceramics range from 0% to 0.004%. Recently, ceramic liners are encased in a titanium sleeve to further decrease the chances of fracture. We encountered 1 case of a metal-encased acetabular liner fracture in a ceramic-on-ceramic articulation in a series of 764 hips. Our literature review revealed no reports of metal-encased ceramic liner fracture. A 60-year-old woman presented 27 months after a bilateral ceramic-on-ceramic THA. She reported mechanical grinding and clicking from the left hip on extension. There was no history of trauma or fall. Examination revealed a nonantalgic gait and audible-palpable crepitations on the left hip. Range of motion of the left hip was intact with no subluxation. Radiographs revealed fractured ceramic insert and an excessively anteverted socket on the left side. Intraoperative findings revealed gross impingement in the form of indentation of the metal femoral neck against the elevated metal rim encasing the liner. Revision THA was performed using an uncemented polyethylene liner while retaining the well-fixed cup and stem. The Harris Hip Score at 4.5-year follow-up was 100, with no evidence of osteolysis or polyethylene wear.
引用
收藏
页码:E682 / E684
页数:3
相关论文
共 14 条
[1]
Third-Generation Ceramic-on-Ceramic Bearing Surfaces in Revision Total Hip Arthroplasty [J].
Chang, Jun-Dong ;
Kamdar, Rutuj ;
Yoo, Je-Hyun ;
Hur, Mina ;
Lee, Sang-Soo .
JOURNAL OF ARTHROPLASTY, 2009, 24 (08) :1231-1235
[2]
A titanium-encased alumina ceramic bearing for total hip arthroplasty - 3-to 5-year results [J].
D'Antonio, JA ;
Capello, WN ;
Manley, MT ;
Naughton, M ;
Sutton, K .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2005, (441) :151-158
[3]
Fracture of the ABC® ceramic liner -: A report of three cases [J].
Diwanji, Sanket R. ;
Seon, Jong Kenn ;
Song, Eun. Kyoo ;
Yoon, Taek Rim .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2007, (464) :242-246
[4]
Patterns of failure of ceramic liner in total hip replacement: report of two cases [J].
Faraj, Adnan .
EUROPEAN JOURNAL OF ORTHOPAEDIC SURGERY AND TRAUMATOLOGY, 2009, 19 (01) :43-46
[5]
Garino JP, 2000, CLIN ORTHOP RELAT R, P41
[6]
Ceramic-on-Ceramic Total Hip Arthroplasty [J].
Greene, Joseph W. ;
Malkani, Arthur L. ;
Kolisek, Frank R. ;
Jessup, Nenette M. ;
Baker, Dale L. .
JOURNAL OF ARTHROPLASTY, 2009, 24 (06) :15-18
[7]
Ceramic liner fracture after cementless alumina-on-alumina total hip arthroplasty [J].
Ha, Yong-Chan ;
Kim, Shin-Yoon ;
Kim, Hee Joong ;
Yoo, Jeong Joon ;
Koo, Kyung-Hoi .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2007, (458) :106-110
[8]
Ceramic acetabular liner fracture in total hip arthroplasty with a ceramic sandwich cup [J].
Hasegawa, M ;
Sudo, A ;
Hirata, H ;
Uchida, A .
JOURNAL OF ARTHROPLASTY, 2003, 18 (05) :658-661
[9]
Heros R., 1998, SEMIN ARTHROPLASTY, V9, P114
[10]
Incidence of Ceramic Liner Malseating in TridentA® Acetabular Shell [J].
Miller, Anna N. ;
Su, Edwin P. ;
Bostrom, Mathias P. G. ;
Nestor, Bryan J. ;
Padgett, Douglas E. .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2009, 467 (06) :1552-1556