Ceramic liner fracture after cementless alumina-on-alumina total hip arthroplasty

被引:94
作者
Ha, Yong-Chan
Kim, Shin-Yoon
Kim, Hee Joong
Yoo, Jeong Joon
Koo, Kyung-Hoi
机构
[1] Gyeongsang Natl Univ, Coll Med, Dept Orthopaed Surg, Chinju, South Korea
[2] Kyungpook Natl Univ, Coll Med, Dept Orthopaed Surg, Taegu, South Korea
[3] Seoul Natl Univ, Coll Med, Dept Orthopaed Surg, Seoul, South Korea
关键词
D O I
10.1097/BLO.0b013e3180303e87
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Advances in technology have reduced the risk of fracture of ceramic total hip arthroplasty implants, but concerns remain about fracture of both components. We retrospectively reviewed 133 patients (157 hips) who had cementless alumina-on-alumina total hip arthroplasties with a sandwich-type acetabular component. Six patients (seven hips) died and five patients (six hips) were interviewed by telephone (95% followup). The 122 patients (144 hips) examined had a minimum followup of 36 months (average, 45 months; range, 36-68 months). All acetabular cups and femoral stems were radiographically stable at the last followup. Five hips in five patients (3.5%) were revised because of ceramic liner fractures. Ceramic liner fractures occurred at a mean of 35 months (range, 24-48 months) postoperatively. Acetabular cups in the fracture group (n = 5) were more anteverted than those in the nonfracture group (n = 139). In three patients the fracture apparently occurred during squatting, resulting in hyperflexion and wide hip abduction. Early ceramic liner fracture was associated with impingement associated with excessive anteversion of the acetabular cup in Korean patients who habitually squat. Level of Evidence: Level IV, therapeutic study. See the Guidelines for Authors for a complete description of levels of evidence.
引用
收藏
页码:106 / 110
页数:5
相关论文
共 29 条
[1]
Computer-based motion simulation of total hip prostheses with ceramic-on-ceramic wear couple. Analysis of implant design and orientation as influence parameters [J].
Bader, R ;
Steinhauser, E ;
Gradinger, R ;
Willmann, G ;
Mittelmeier, W .
ZEITSCHRIFT FUR ORTHOPADIE UND IHRE GRENZGEBIETE, 2002, 140 (03) :310-316
[2]
Barrack Robert L, 2003, J Am Acad Orthop Surg, V11, P89
[3]
Ceramic-on-ceramic bearings in total hip arthroplasty [J].
Bierbaum, BE ;
Nairus, J ;
Kuesis, D ;
Morrison, JC ;
Ward, D .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2002, (405) :158-163
[4]
Bizot P, 2000, CLIN ORTHOP RELAT R, P134
[5]
LONG-TERM RESULTS OF UNCEMENTED ALUMINA ACETABULAR IMPLANTS [J].
BOEHLER, M ;
KNAHR, K ;
PLENK, H ;
WALTER, A ;
SALZER, M ;
SCHREIBER, V .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1994, 76B (01) :53-59
[6]
Range of motion and stability in total hip arthroplasty with 28-, 32-, 38-, and 44-mm femoral head sizes - An in vitro study [J].
Burroughs, BR ;
Hallstrom, B ;
Golladay, GJ ;
Hoeffel, D ;
Harris, WH .
JOURNAL OF ARTHROPLASTY, 2005, 20 (01) :11-19
[7]
Capello William N, 2005, Instr Course Lect, V54, P171
[8]
Catelas I, 1998, J BIOMED MATER RES, V41, P600, DOI 10.1002/(SICI)1097-4636(19980915)41:4<600::AID-JBM12>3.0.CO
[9]
2-I
[10]
Cytotoxicity and macrophage cytokine release induced by ceramic and polyethylene particles in vitro [J].
Catelas, I ;
Petit, A ;
Marchand, R ;
Zukor, DJ ;
Yahia, LH ;
Huk, OL .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1999, 81B (03) :516-521