Periotron 8000: calibration characteristics and reliability

被引:31
作者
Ciantar, M [1 ]
Caruana, DJ
机构
[1] Univ Malta, Dept Dent Surg, Msida, Malta
[2] Univ Malta, Dept Biomed Sci, Msida, Malta
关键词
Periotron; calibration; submicrolitre volumes; gingival crevicular fluid;
D O I
10.1111/j.1600-0765.1998.tb02198.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The Periotron is an instrument designed to quantify submicrolitre volumes of fluid sampled on a filter paper strip. To date 3 models have been manufactured: the Periotron 600 (1976), the Periotron 6000 (1983) and more recently the Periotron 8000 (1995). This paper investigated for the first time the calibration characteristics and reliability of the Periotron 8000. The fluids under investigation were: de-ionised water, human serum, fetal bovine serum and an ultrafiltrate of fetal bovine serum. Quantitative analysis was studied by recording a series of Periotron readings over a volume range of 0-1.0 mu l for each fluid. The average of 5 Periotron values for each particular fluid was then plotted versus the respective fluid volume. Qualitative changes in fluid composition versus Periotron Scores were also analysed. Volume conversion for Periotron scores using both Periotron MLCONVRT software and a best fit equation selected from TableCurve 2D software compared well. The results of this study revealed that: 1)differences in calibration fluid composition (e.g. protein content) are reflected in the Periotron scores 2) positioning of filter paper strip between the jaws of the Periotron should be standardised, 3) calibration of the Periotron 8000 seems to be consistent over a I-wk interval.
引用
收藏
页码:259 / 264
页数:6
相关论文
共 26 条
[1]   THE ABILITY OF GINGIVAL CREVICULAR FLUID ACUTE PHASE PROTEINS TO DISTINGUISH HEALTHY, GINGIVITIS AND PERIODONTITIS SITES [J].
ADONOGIANAKI, E ;
MOONEY, J ;
KINANE, DF .
JOURNAL OF CLINICAL PERIODONTOLOGY, 1992, 19 (02) :98-102
[2]   RELIABILITY OF STANDARDIZED NARROW STRIPS IN THE PERIOTRON [J].
ASIKAINEN, S ;
ETEMADZADEH, H ;
AINAMO, J .
JOURNAL OF PERIODONTOLOGY, 1985, 56 (11) :686-689
[3]  
BANG JS, 1975, J DENT RES, V50, P638
[4]   RELIABILITY OF VOLUME MEASUREMENTS WITH THE NEW PERIOTRON-6000 [J].
BICKEL, M ;
CIMASONI, G .
JOURNAL OF PERIODONTAL RESEARCH, 1984, 19 (03) :313-316
[5]  
BLACK GV, 1982, ILLINOIS STATE DENTA, P93
[6]   SITE-SPECIFIC ALKALINE-PHOSPHATASE LEVELS IN GINGIVAL CREVICULAR FLUID IN HEALTH AND GINGIVITIS - CROSS-SECTIONAL STUDIES [J].
CHAPPLE, ILC ;
GLENWRIGHT, HD ;
MATTHEWS, JB ;
THORPE, GHG ;
LUMLEY, PJ .
JOURNAL OF CLINICAL PERIODONTOLOGY, 1994, 21 (06) :409-414
[7]   CALIBRATION AND RELIABILITY OF THE PERIOTRON-6000 FOR INDIVIDUAL GINGIVAL CREVICULAR FLUID SAMPLES [J].
CHAPPLE, ILC ;
CROSS, IA ;
GLENWRIGHT, HD ;
MATTHEWS, JB .
JOURNAL OF PERIODONTAL RESEARCH, 1995, 30 (01) :73-79
[8]   UREA LEVELS IN GINGIVAL CREVICES OF DIABETIC AND NORMAL ADOLESCENTS [J].
CIANCIO, SG ;
GOLUB, LM ;
MOSOVICH, L ;
KATZ, C ;
KLEINBERG, I .
JOURNAL OF DENTAL RESEARCH, 1977, 56 (10) :1144-1144
[9]  
Cimasoni G., 1974, MONOGRAPHS ORAL SCI, V3
[10]   COMPARISON OF GLUCOSE AND PROTEIN-CONTENT OF GINGIVAL FLUID FROM DIABETICS AND NONDIABETICS [J].
FICARA, AJ ;
LEVIN, MP ;
GROWER, MF ;
KRAMER, GD .
JOURNAL OF PERIODONTAL RESEARCH, 1975, 10 (03) :171-175