Multichannel CdZnTe gamma ray spectrometer

被引:2
作者
Doty, FP
Lingren, CL
Apotovsky, BA
Brunsch, J
Butler, JF
Collins, T
Conwell, RL
Friesenhahn, S
Gormley, J
Pi, B
Zhao, S
Augustine, FL
Bennett, BA
Cross, E
James, RB
机构
[1] Digirad Corp, San Diego, CA 92126 USA
[2] Augustine Engn, Encinitas, CA USA
[3] Sandia Natl Labs, Livermore, CA USA
关键词
D O I
10.1016/S0168-9002(98)00952-8
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A 3 cm(3) multichannel gamma spectrometer for DOE applications is under development by Digirad Corporation. The device is based on a position sensitive detector packaged in a compact multi-chip module (MCM) with integrated readout circuitry. The modular, multichannel design will enable identification and quantitative analysis of radionuclides in extended sources, or sources containing low levels of activity. The MCM approach has the advantages that the modules are designed for imaging applications, and the sensitivity can be arbitrarily increased by increasing the number of pixels, i.e. adding modules to the instrument. For a high sensitivity probe, the outputs for each pixel can be corrected for gain and offset variations, and summed digitally. Single pixel results obtained with discrete low noise readout indicate energy resolution of 3 keV can be approached with currently available CdZnTe. The energy resolution demonstrated to date with MCMs for 511 keV gamma rays is 10 keV. (C) 1999 Published by Elsevier Science B.V. All rights reserved.
引用
收藏
页码:185 / 189
页数:5
相关论文
共 14 条
[1]  
Apotovsky B., 1997, US patent 5,677,539, Patent No. 5677539
[2]   Progress in developing focal-plane-multiplexer readout for large CdZnTe arrays for nuclear medicine applications [J].
Barber, HB ;
Marks, DG ;
Apotovsky, BA ;
Augustine, FL ;
Barrett, HH ;
Butler, JF ;
Dereniak, EL ;
Doty, FP ;
Eskin, JD ;
Hamilton, WJ ;
Matherson, KJ ;
Venzon, JE ;
Woolfenden, JM ;
Young, ET .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 380 (1-2) :262-265
[3]   Application of II-VI materials to nuclear medicine [J].
Barber, HB .
JOURNAL OF ELECTRONIC MATERIALS, 1996, 25 (08) :1232-1240
[4]   CHARGE-CARRIER MOBILITIES IN CD0.8ZN0.2TE SINGLE-CRYSTALS USED AS NUCLEAR RADIATION DETECTORS [J].
BURSHTEIN, Z ;
JAYATIRTHA, HN ;
BURGER, A ;
BUTLER, JF ;
APOTOVSKY, B ;
DOTY, FP .
APPLIED PHYSICS LETTERS, 1993, 63 (01) :102-104
[5]   GAMMA-RAY AND X-RAY-DETECTORS MANUFACTURED FROM CD1-XZNX TE GROWN BY A HIGH-PRESSURE BRIDGMAN METHOD [J].
BUTLER, JF ;
DOTY, FP ;
APOTOVSKY, B ;
LAJZEROWICZ, J ;
VERGER, L .
MATERIALS SCIENCE AND ENGINEERING B-SOLID STATE MATERIALS FOR ADVANCED TECHNOLOGY, 1993, 16 (1-3) :291-295
[6]  
*DOE, 1995, EM30 DOE
[7]   PROPERTIES OF CDZNTE CRYSTALS GROWN BY A HIGH-PRESSURE BRIDGMAN METHOD [J].
DOTY, FP ;
BUTLER, JF ;
SCHETZINA, JF ;
BOWERS, KA .
JOURNAL OF VACUUM SCIENCE & TECHNOLOGY B, 1992, 10 (04) :1418-1422
[8]   PIXELLATED CDZNTE DETECTOR ARRAYS [J].
DOTY, FP ;
BARBER, HB ;
AUGUSTINE, FL ;
BUTLER, JF ;
APOTOVSKY, BA ;
YOUNG, ET ;
HAMILTON, W .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 353 (1-3) :356-360
[9]  
DOTY FP, 1993, P SOC PHOTO-OPT INS, V1945, P145, DOI 10.1117/12.158756
[10]  
DOTY FP, 1994, EMIS DATAREVIEWS SER, V10, P540