Bienzyme HRP-GOx-modified gold nanoelectrodes for the sensitive amperometric detection of glucose at low overpotentials

被引:78
作者
Delvaux, M
Walcarius, A
Demoustier-Champagne, S
机构
[1] Catholic Univ Louvain, Unite Phys & Chim Hauts Polymeres, B-1348 Louvain, Belgium
[2] Univ Nancy 1, CNRS, UMR 7564, Lab Chim Phys & Microbiol Environm, F-54600 Villers Les Nancy, France
关键词
enzyme-modified gold nanoelectrodes; horseradish peroxidase; glucose oxidase; glucose amperometric detection;
D O I
10.1016/j.bios.2004.07.017
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Gold nanotubular electrode ensembles were prepared by using electroless deposition of the metal within the pores of polycarbonate track-etched membranes. Mono-enzyme (GOx) and monolayer/bilayer bienzyme (GOx/HRP) bioelectrodes were prepared by immobilizing the enzymes onto gold nanotubes surfaces modified with mercaptoethylamine. Batch amperometric responses to glucose for the different bioelectrodes were determined and compared. The response of the two geometries (monolayer and bilayer) of the bienzyme electrodes was shown to vary with regard to sensitivity at detection potentials above 0 V. On the contrary, at detection potentials below 0 V, no noticeable influence of the configuration of the bienzyme on the response intensity was observed. The mono-enzyme (650 muA mM(-1) in benzoquinone (BQ) at -0.8 V versus Ag/AgCl) and the two bienzyme bioelectrodes (+/-400 muA mM(-1) in hydroquinone (H(2)Q) at -0.2 V versus Ag/AlCl) display remark-able sensitivities compared to a classical GOx-modified gold macroelectrode (13 muA mM(-1) in BQ at -0.8 V versus Ag/AgCl). A remark-able feature of the bienzyme electrodes is the possibility to detect glucose at very low applied potentials where the noise level and interferences from other electro-oxidizable compounds are minimal. Another important characteristic of the monolayer bienzyme electrode is the possible existence of a direct electronic communication between HRP and the transducer surface. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:1587 / 1594
页数:8
相关论文
共 32 条
[21]   AMPEROMETRIC GLUCOSE SENSORS BASED ON IMMOBILIZED GLUCOSE-OXIDIZING ENZYMES AND CHEMICALLY MODIFIED ELECTRODES [J].
GORTON, L ;
BREMLE, G ;
CSOREGI, E ;
JONSSONPETTERSSON, G ;
PERSSON, B .
ANALYTICA CHIMICA ACTA, 1991, 249 (01) :43-54
[22]  
HANOT H, 2001, Patent No. 00149403
[23]   AMPEROMETRIC BIOSENSOR BASED ON DIRECT COMMUNICATION BETWEEN GLUCOSE-OXIDASE AND A CONDUCTING POLYMER INSIDE THE PORES OF A FILTRATION MEMBRANE [J].
KOOPAL, CGJ ;
DERUITER, B ;
NOLTE, RJM .
JOURNAL OF THE CHEMICAL SOCIETY-CHEMICAL COMMUNICATIONS, 1991, (23) :1691-1692
[24]   ELIMINATION OF ELECTROOXIDIZABLE INTERFERANT-PRODUCED CURRENTS IN AMPEROMETRIC BIOSENSORS [J].
MAIDAN, R ;
HELLER, A .
ANALYTICAL CHEMISTRY, 1992, 64 (23) :2889-2896
[25]   FABRICATION AND EVALUATION OF NANOELECTRODE ENSEMBLES [J].
MENON, VP ;
MARTIN, CR .
ANALYTICAL CHEMISTRY, 1995, 67 (13) :1920-1928
[26]   Enzyme electrodes based on self-assembled monolayers of thiol compounds on gold [J].
Mizutani, F ;
Sato, Y ;
Yabuki, S ;
Sawaguchi, T ;
Iijima, S .
ELECTROCHIMICA ACTA, 1999, 44 (21-22) :3833-3838
[27]   BIENZYME SENSORS BASED ON ELECTRICALLY WIRED PEROXIDASE [J].
OHARA, TJ ;
VREEKE, MS ;
BATTAGLINI, F ;
HELLER, A .
ELECTROANALYSIS, 1993, 5 (9-10) :825-831
[28]   Peroxidase-glucose oxidase-poly(amphiphilic pyrrole) bioelectrode for selectively mediated amperometric detection of glucose [J].
Popescu, IC ;
Cosnier, S ;
Labbe, P .
ELECTROANALYSIS, 1997, 9 (13) :998-1004
[29]   GLUCOSE AND ACETYLCHOLINE SENSING MULTILAYER ENZYME ELECTRODES OF CONTROLLED ENZYME LAYER THICKNESS [J].
RIKLIN, A ;
WILLNER, I .
ANALYTICAL CHEMISTRY, 1995, 67 (22) :4118-4126
[30]   ENZYME ELECTRODES WITH SUBSTRATE AND CO-ENZYME AMPLIFICATION [J].
SCHUBERT, F ;
KIRSTEIN, D ;
SCHRODER, KL ;
SCHELLER, FW .
ANALYTICA CHIMICA ACTA, 1985, 169 (MAR) :391-396