Although dual-energy X-ray absorptiometry (DXA) is widely used in clinical research as a means of quantifying body composition, there remains at present little published information that reviews the method's underlying physical basis. Because a clear understanding of DXA physical concepts is integral to appropriate use and interpretation, we present here a three-section review that includes both relevant in vitro and in vivo experimental demonstrations. In the first section we describe the main physical principles on which DXA is based. The section that follows presents a step-by-step analysis of the DXA two-component soft tissue model. In the final section we demonstrate how knowledge of physical concepts can lead to resolution of important methodological concerns, such as the influence of hydration changes on DXA fat estimates. A thorough understanding of DXA physical concepts provides a basis for appropriate interpretation of measurement results and stimulates many new and important research questions.
引用
收藏
页码:E941 / E951
页数:11
相关论文
共 54 条
[11]   ASSESSMENT OF THE COMPOSITION OF MAJOR BODY REGIONS BY DUAL-ENERGY X-RAY ABSORPTIOMETRY (DEXA), WITH SPECIAL REFERENCE TO LIMB MUSCLE MASS [J].
FULLER, NJ ;
LASKEY, MA ;
ELIA, M .
CLINICAL PHYSIOLOGY, 1992, 12 (03) :253-266
[12]   EVALUATION OF A NEW SET OF CALIBRATION STANDARDS FOR THE MEASUREMENT OF FAT-CONTENT VIA DPA AND DXA [J].
GOODSITT, MM .
MEDICAL PHYSICS, 1992, 19 (01) :35-44
[13]   Cross-calibration of body-composition techniques against dual-energy X-ray absorptiometry in young children [J].
Goran, MI ;
Driscoll, P ;
Johnson, R ;
Nagy, TR ;
Hunter, G .
AMERICAN JOURNAL OF CLINICAL NUTRITION, 1996, 63 (03) :299-305
[14]   MEASUREMENT OF LEAN BODY-MASS AND TOTAL-BODY FAT USING DUAL PHOTON-ABSORPTIOMETRY [J].
GOTFREDSEN, A ;
JENSEN, J ;
BORG, J ;
CHRISTIANSEN, C .
METABOLISM-CLINICAL AND EXPERIMENTAL, 1986, 35 (01) :88-93
[15]   TOTAL-BODY BONE-MINERAL INVIVO BY DUAL PHOTON-ABSORPTIOMETRY .1. MEASUREMENT PROCEDURES [J].
GOTFREDSEN, A ;
BORG, J ;
CHRISTIANSEN, C ;
MAZESS, RB .
CLINICAL PHYSIOLOGY, 1984, 4 (04) :343-355
[16]  
Gurr M.I., 1991, LIPID BIOCH, V4th
[17]  
HANSEN MA, 1990, J NUCL MED, V31, P1156
[18]  
HASSAGER C, 1986, IN VIVO BODY COMPOSI, P51
[19]  
Heymsfield S. B., 1996, HUMAN BODY COMPOSITI, P129
[20]   DUAL-PHOTON ABSORPTIOMETRY - COMPARISON OF BONE-MINERAL AND SOFT-TISSUE MASS MEASUREMENTS INVIVO WITH ESTABLISHED METHODS [J].
HEYMSFIELD, SB ;
WANG, J ;
HESHKA, S ;
KEHAYIAS, JJ ;
PIERSON, RN .
AMERICAN JOURNAL OF CLINICAL NUTRITION, 1989, 49 (06) :1283-1289