Clinical experience with the Novare Enclose II manual proximal anastomotic device during off-pump coronary artery surgery

被引:17
作者
Akpinar, B
Gulden, M
Sagbas, E
Sanisoglu, H
Ergenoglu, MU
Turkoglu, C
机构
[1] Florence Nightingale Hosp, Dept Cardiac Surg, TR-80220 Istanbul, Turkey
[2] Florence Nightingale Hosp, Dept Cardiol, TR-80220 Istanbul, Turkey
关键词
anastomosis; device; off-pump; surgery;
D O I
10.1016/j.ejcts.2005.02.041
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: The aim of this study was to assess the feasibility and safety of a new proximal anastomotic device during off-pump coronary surgery. Methods: The Novare Enclose II is a manual proximal anastomotic device that enables the surgeon to perform proximal anastomoses without side-biting clam ps in a fully pressurized aorta. The device was used in 30 off-pump coronary artery bypass graft (CABG) patients (Group A) for performing 25 vein, 10 free right internal thoracic artery and 14 radial artery anastomoses to the aorta. The number of proximal anastomoses varied between 1 and 3 per patient. Intraoperative transcranial Doppler measurements were performed to detect cerebral microemboli during the use of the device and were compared with transcranial Doppler measurements of 30 off-pump patients where a partial side clamp was used for proximal anastomoses (Group B). The aorta was evaluated using transesophageal echocardiography (TEE) and epiaortic echocardiography (EAE) in both groups before the procedure and patients with a,calcified aorta were excluded. Results: In Group A, 49 proximal anostomoses were performed as planned with the device using conventional hand-sewn methods. Twenty-four of these were arterial conduits. In Group B, 32 proximal anastomoses were performed using 25 veins and 7 arterial grafts. There were no procedure related to adverse events or complications in both groups. The median number of microemboli was 15 (range 5-48) in Group A and 68 (range 35-290) in Group B (P < 0.05). Conclusion: The Enclose II device can be a valuable tool. to perform venous or arterial proximal anastomoses during off-pump CABG. Transcranial Doppler results suggest that the device seems to be less traumatizing than side clamping and may reduce clamp-associated complications during off-pump coronary artery surgery. (c) 2005 Elsevier B.V. All rights reserved.
引用
收藏
页码:1070 / 1073
页数:4
相关论文
共 10 条
[1]   The role of microembolisation in cerebral injury as defined by functional magnetic resonance imaging [J].
Abu-Omar, Y ;
Cifelli, A ;
Matthews, PM ;
Taggart, DP .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2004, 26 (03) :586-590
[2]  
BARZILAI B, 1989, CIRCULATION, V80, P275
[3]  
BLAUTH CI, 1992, J THORAC CARDIOV SUR, V103, P1104
[4]   Clinical experience with devices for facilitated anastomoses in coronary artery bypass surgery [J].
Carrel, TP ;
Eckstein, FS ;
Englberger, L ;
Berdat, PA ;
Schmidli, J .
ANNALS OF THORACIC SURGERY, 2004, 77 (03) :1110-1120
[5]   Increased incidence of acute ascending aortic dissection with off-pump aortocoronary bypass surgery? [J].
Chavanon, O ;
Carrier, M ;
Cartier, R ;
Hébert, Y ;
Pellerin, M ;
Pagé, P ;
Perrault, LP .
ANNALS OF THORACIC SURGERY, 2001, 71 (01) :117-121
[6]   Minimizing aortic manipulation during OPCAB using the symmetry aortic connector system for proximal vein graft anastomoses [J].
Eckstein, FS ;
Bonilla, LF ;
Englberger, L ;
Stauffer, E ;
Berg, TA ;
Schmidli, J ;
Carrel, TP .
ANNALS OF THORACIC SURGERY, 2001, 72 (03) :S995-S998
[7]   Benefit of bilateral over single internal mammary artery grafts for multiple coronary artery bypass grafting [J].
Endo, M ;
Nishida, H ;
Tomizawa, Y ;
Kasanuki, H .
CIRCULATION, 2001, 104 (18) :2164-2170
[8]   Rupture of a polypropylene suture after aortic operation: A scanning electronic microscopical assessment of potential mechanisms [J].
Huber, C ;
Eckstein, FS ;
Halbeisen, M ;
Carrel, TP .
ANNALS OF THORACIC SURGERY, 2003, 75 (04) :1318-1321
[9]   Superiority of bilateral internal thoracic artery grafting - It's been a long time comin' [J].
Lytle, BW ;
Loop, FD .
CIRCULATION, 2001, 104 (18) :2152-2154
[10]   Anastomotic devices for coronary surgery [J].
Wolf, RK .
MINIMALLY INVASIVE THERAPY & ALLIED TECHNOLOGIES, 2004, 13 (01) :4-10