Ensuring both normality and homocedasticity of chromatographic data-ratios for internal-standard least-squares calibration

被引:7
作者
Cuadros-Rodriguez, L [1 ]
Gonzalez-Casado, A [1 ]
Garcia-Campana, AM [1 ]
Vilchez, JL [1 ]
机构
[1] Univ Granada, Fac Sci, Dept Analyt Chem, E-18071 Granada, Spain
关键词
chromatographic data; internal-standard calibration; least-squares regression; ratio-data homocedasticity; ratio-data normality;
D O I
10.1007/BF02467494
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Because the quotient of two normal variables does not fit a normal distribution and tends to be heterocedastic, the use of internal-standard least-squares regression on data obtained from such ratios gives a calibration function which can result in inaccurate determination of the concentration of analyte in a sample. In this work we have studied simulated situations in which area-ratios are fitted to a normal distribution. It is shown that the least-squares model is inappropriate when the ordinate data-ratios are not normally distributed, and some recommendations are proposed to ensure both normality and homocedasticity of the quotient of analyte/internal-standard signals. Finally, GC-MS separation and quantification of nine fatty acids by use of an internal standard are used for experimental verification of the conclusions reached.
引用
收藏
页码:550 / 556
页数:7
相关论文
共 26 条
[1]  
ANDRES AM, 1994, BIOESTADISTICA CIENC, P461
[2]   An internal standard method in alpha spectrometric determination of uranium and thorium radioisotopes using instrumental neutron activation analysis [J].
Byrne, AR ;
Benedik, L .
ANALYTICAL CHEMISTRY, 1997, 69 (06) :996-999
[3]   ESTIMATION OF THE LINEAR WORKING RANGE FOR A GAS-CHROMATOGRAPH MASS-SPECTROMETER [J].
CROCKETT, PW .
ANALYTICAL CHEMISTRY, 1986, 58 (04) :699-705
[4]   INTERNAL STANDARDIZATION TECHNIQUE FOR CAPILLARY ZONE ELECTROPHORESIS [J].
DOSE, EV ;
GUIOCHON, GA .
ANALYTICAL CHEMISTRY, 1991, 63 (11) :1154-1158
[5]   INTERNAL STANDARDS IN DIFFERENTIAL PULSE ANODIC-STRIPPING VOLTAMMETRY [J].
FERNANDO, AR ;
KRATOCHVIL, B .
CANADIAN JOURNAL OF CHEMISTRY-REVUE CANADIENNE DE CHIMIE, 1991, 69 (04) :755-758
[6]   The frequency distribution of the quotient of two normal variates [J].
Geary, RC .
JOURNAL OF THE ROYAL STATISTICAL SOCIETY, 1930, 93 :442-446
[7]  
GRAHAM RC, 1993, DATA ANAL CHEM SCI
[8]  
HEFELFINGER P, 1981, J CHROMATOGR, V218, P73
[9]   ON RATIO OF 2 CORRELATED NORMAL RANDOM VARIABLES [J].
HINKLEY, DV .
BIOMETRIKA, 1969, 56 (03) :635-&
[10]   EFFECT OF THE AMOUNT OF INTERNAL STANDARD ON THE PRECISION OF AN ANALYTICAL METHOD [J].
JEMAL, M ;
BERGUM, J .