First direct in vivo comparison of two commercially available three-dimensional quantitative coronary angiography systems

被引:33
作者
Ramcharitar, Steve [1 ]
Daeman, Joost [1 ]
Patterson, Mark [1 ]
van Guens, Robert Jan [1 ]
Boersma, Eric [1 ]
Serruys, Patrick W. [1 ]
van der Giessen, Willem J. [1 ]
机构
[1] Erasmus MC, Thoraxctr, Rotterdam, Netherlands
关键词
accuracy; in vivo validation; quantitative coronary angiography; three-dimensional reconstruction; diagnostic cardiac catheterization; percutaneous coronary intervention;
D O I
10.1002/ccd.21418
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aim: The in vivo comparison of the accuracy of two 3-dimensional quantitative coronary angiography (QCA) systems. Methods: Precision-drilled plexiglass phantoms with five different luminal diameters (0.5-1.9 mm) were percutaneously inserted into the coronary arteries of four Yorkshire pigs. Twenty-one angiographic images of these stenotic phantoms were acquired for in vivo validation testing. Quantitative assessments of the minimum, maximum, and mean luminal diameters together with the minimum luminal area were determined using two 3D QCA systems, the CardiOp-B-(R) and CAAS 5. Results: The CardiOp-B system significantly underestimated the minimum luminal diameter MLD whilst both systems significantly overestimated the maximum luminal diameter at the minimal luminal area (MLA) over the phantom's true value. The CAAS 5 system had a greater degree of accuracy/mm (mean difference = 0.01 vs. 0.03) and precision/mm (SD = 0.09 vs. 0.23) than the CardiOp-B in assessing the minimal LD. An increased precision/mm (SD = 0.01 vs. 0.29) and accuracy/mm (mean difference = 0.03 vs. 0.11) in the mean LD was observed with the CAAS 5. In comparing the MLA/mm(2) the CAAS 5 was more precise/mm(2) (SD = 0.14 vs. 0.55) and accurate/mm(2) (mean difference = 0.12 vs. 0.02) to the true phantom MLA compared to the CardiOp-B system. Conclusions: In a 21 phantom study, the CAAS 5 3D QCA system had a greater degree of accuracy and precision in both the luminal and area measurements than the CardiOp-B 3D QCA system. (c) 2008 Wiley-Liss, Inc.
引用
收藏
页码:44 / 50
页数:7
相关论文
共 15 条
[1]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[2]   Effects of point configuration on the accuracy in 3D reconstruction from biplane images [J].
Dmochowski, J ;
Hoffmann, KR ;
Singh, V ;
Xu, JH ;
Nazareth, DP .
MEDICAL PHYSICS, 2005, 32 (09) :2862-2869
[3]  
Dvir Danny, 2007, EuroIntervention, V3, P95
[4]  
Dvir Danny, 2005, Int J Cardiovasc Intervent, V7, P141
[5]   PITFALLS IN THE DETERMINATION OF ABSOLUTE DIMENSIONS USING ANGIOGRAPHIC CATHETERS AS CALIBRATION DEVICES IN QUANTITATIVE ANGIOGRAPHY [J].
FORTIN, DF ;
SPERO, LA ;
CUSMA, JT ;
SANTORO, L ;
BURGESS, R ;
BASHORE, TM .
AMERICAN JOURNAL OF CARDIOLOGY, 1991, 68 (11) :1176-1182
[6]   Utility of three-dimensional reconstruction of coronary angiography to guide percutaneous coronary intervention [J].
Gollapudi, Raghava R. ;
Valencia, Rafael ;
Lee, Steve S. ;
Wong, Garrett B. ;
Teirstein, Paul S. ;
Price, Matthew J. .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2007, 69 (04) :479-482
[7]   Clinical assessment of a new real time 3D quantitative coronary angiography system: Evaluation in stented vessel segments [J].
Gradaus, R ;
Mathies, K ;
Breithardt, G ;
Boecker, D .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2006, 68 (01) :44-49
[8]   Three-dimensional vascular angiography [J].
Green, NE ;
Chen, SYJ ;
Messenger, JC ;
Groves, BM ;
Carroll, JD .
CURRENT PROBLEMS IN CARDIOLOGY, 2004, 29 (03) :104-142
[9]   INVIVO VALIDATION OF ONLINE AND OFF-LINE GEOMETRIC CORONARY MEASUREMENTS USING INSERTION OF STENOSIS PHANTOMS IN PORCINE CORONARY-ARTERIES [J].
HAASE, J ;
DIMARIO, C ;
SLAGER, CJ ;
VANDERGIESSEN, WJ ;
DENBOER, A ;
DEFEYTER, PJ ;
REIBER, JHC ;
VERDOUW, PD ;
SERRUYS, PW .
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1992, 27 (01) :16-27
[10]   COMPARATIVE VALIDATION OF QUANTITATIVE CORONARY ANGIOGRAPHY SYSTEMS - RESULTS AND IMPLICATIONS FROM A MULTICENTER STUDY USING A STANDARDIZED APPROACH [J].
KEANE, D ;
HAASE, J ;
SLAGER, CJ ;
VANSWIJNDREGT, EM ;
LEHMANN, KG ;
OZAKI, Y ;
DIMARIO, C ;
KIRKEEIDE, R ;
SERRUYS, PW .
CIRCULATION, 1995, 91 (08) :2174-2183